| Model Number |
00884450263181 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); Non specific EKG/ECG Changes (1817); Cardiac Tamponade (2226); Decreased Respiratory Rate (2485); Low Cardiac Output (2501); Blood Loss (2597)
|
| Date of Event |
03/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a difficult transseptal puncture to complete an ablation procedure, a pericardial effusion occurred.The physician had acquired access through the fossa ovalis and into the left atrium with a coronary sinus catheter attempting to effectively map the left atrium for this patient.After a few seconds of use, the catheter would not deflect as intended.The catheter was replaced, and the procedure was continued.After mapping the left atrium, the patient became hypotensive.Ultrasound confirmed a pericardial effusion.A pericardiocentesis was performed to stabilize the patient and the procedure was discontinued.The physician alleged the effusion occurred during the transseptal puncture procedure.The physician attempted several times to acquire access with the needle which resulted in the effusion.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.Should the suspect medical device be returned at a later date, the investigation will be re-opened.
|
| |
|
Search Alerts/Recalls
|