Model Number E053
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/03/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product has been received for analysis.Upon completion of the analysis of the complaint device, if there is any further relevant information from that review, a supplemental report will be filed.
 
Event or Problem Description
It was reported that this device was explanted as it declared a 1003 code indicative to voltage too low for remaining capacity.No additional adverse patient effects were reported.Dc.
 
Additional Manufacturer Narrative
The returned pulse generator was analyzed and a review of the device memory confirmed that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device's battery.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.Malfunction of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.Boston scientific has issued an advisory communication regarding an older subset of cognis/teligen devices that is more susceptible to this anomaly.Specifically, the performance of a low voltage capacitor may be compromised over time, causing an increased current drain that can lead to premature battery depletion.This particular device was not included in the advisory population.Patient code 3191 was used to capture the patient undergoing surgical intervention to explant the device.
 
Event or Problem Description
It was reported that this device was explanted as it declared a 1003 code indicative to voltage too low for remaining capacity.Device was returned and analyzed.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUNCTUA ICD
Common Device Name
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key9974454
Report Number2124215-2020-07647
Device Sequence Number16669468
Product Code LWP
UDI-Device Identifier00802526480706
UDI-Public00802526480706
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P960040/S235
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/05/2014
Device Model NumberE053
Device Catalogue NumberE053
Device Lot Number100816
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2020
Initial Date Received by Manufacturer 03/10/2020
Supplement Date Received by Manufacturer04/23/2020
Initial Report FDA Received Date04/17/2020
Supplement Report FDA Received Date05/26/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age52 YR
-
-