MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522NAS
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 04/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer reported via call that the customer had high blood glucose level of 535 mg/dl.Customer had blood glucose levels of 478, 409, 346, 435 and 443 mg/dl.Customer was treated with manual injection.Customer was using auto mode.Customer did not allege insulin pump for under delivery.Customer stated that there was no air bubbles and leak.Customer did high pressure test and it was passed.Customer was using insulin pump system within 48 hours of reported high blood glucose level event.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose/flush drive support disk.Unable to perform prime/a33 test, excessive no delivery test and occlusion test due to motor error alarm.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9974875
Report Number2032227-2020-105719
Device Sequence Number14151740
Product Code OYC
UDI-Device Identifier00613994402097
UDI-Public(01)00613994402097
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-522NAS
Device Catalogue NumberMMT-522NAS
Device Lot NumberA5522NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2020
Initial Date Received by Manufacturer 04/14/2020
Supplement Date Received by Manufacturer06/25/2020
Initial Report FDA Received Date04/17/2020
Supplement Report FDA Received Date06/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRN-MMT-326A, UNOMED MMT-397
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
Patient Weight130
-
-