| Model Number |
3721 |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
| Date of Event |
03/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The charger was included in a field correction.Date of event is estimated.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 1627487-2020-03556.It was reported the patient experienced pocket heating while charging the ipg.As a result, the ipg was explanted.The date of explant is unknown.The system was also explanted due to ineffective stimulation as documented in related manufacturer reference numbers 1627487-2020-03665 and 1627487-2020-03666.
|
| |
|
Additional Manufacturer Narrative
|
|
Added the checkmark that was inadvertently left off the initial report.During processing of this complaint, patient weight and complete event information were not requested due to patient not consenting to further contact.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|