MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723NAS
Medical Device Problem Codes Failure to Cycle (1142); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump alarmed compromised force sensor system.The customer also reported that the insulin was squirting during manual prime process.The customer stated that the drive support cap was sticking out.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed displacement test.However device alarmed motor error during basic occlusion test due to loose and protruded drive support disk.Unable to perform occlusion test, prime/a33 test, excessive no delivery test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9975955
Report Number2032227-2020-105889
Device Sequence Number20264546
Product Code OYC
UDI-Device Identifier00643169077737
UDI-Public(01)00643169077737
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723NAS
Device Catalogue NumberMMT-723NAS
Device Lot NumberA6723NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2020
Initial Date Received by Manufacturer 04/15/2020
Supplement Date Received by Manufacturer05/11/2020
Initial Report FDA Received Date04/19/2020
Supplement Report FDA Received Date05/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight104
-
-