ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 25MECJ-502
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Further information regarding this event has been requested.Investigation results will be provided in a subsequent submission.
 
Event or Problem Description
On (b)(6) 2020, a 25mm masters series valve was selected for implant.After placing the valve the posterior leaflet was interfering with the leaflet opening and closing and the valve was removed.A 29mm master series valve was selected for implanted and while rotating the valve the leaflet cracked and the valve was removed from the patient and replaced with another 29mm masters series valve.There was a clinically significant delay in procedure due to the removal of the first 2 valves.The patient was reported to be stable condition.Manufacturer report number: 2648612-2020-00038.
 
Additional Manufacturer Narrative
Additional information: d10, h3, h6; an event of the leaflet mobility being interfered with by the native posterior leaflet was reported.Morphological and hydrodynamic examination indicated the valve met abbott specifications.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Hydrodynamic testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.Information from the field indicated that the leaflet movement was inhibited due to interference by the posterior leaflet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALVE EXPANDED CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key9977148
Report Number2648612-2020-00037
Device Sequence Number11282076
Product Code LWQ
UDI-Device Identifier05414734006347
UDI-Public05414734006347
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/27/2024
Device Model Number25MECJ-502
Device Catalogue Number25MECJ-502
Device Lot Number6926356
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/24/2020
Supplement Date Received by Manufacturer05/20/2020
Initial Report FDA Received Date04/20/2020
Supplement Report FDA Received Date05/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
29MECJ-502, 18290440.; 29MECJ-502, 18290440.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
-
-