|
(b)(4) date sent to the fda: 06/01/2020.A manufacturing record evaluation was performed for the finished device lot kmj392, and no non-conformances were identified.Additional h3 investigation summary: it was received for analysis a needle-suture piece of product code osm8738, lot kmj392.During the visual inspection of the needle-suture, the swage and attachment area were noted to be as expected.However, marks the end of the suture pieces cut that appears to be by the use of a surgical instrument could be observed.Additionally, the functional test was performed on a suture piece and the tensile strength force was above the minimum requirement.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Due to condition of the opened sample, the assignable cause of performance breakage suture suggests an improper handling.
|