| Model Number |
APB-2-4-3D-ES |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding an axium prime coil that released prematurely during a procedure.The patient was being treated for an unruptured saccular aneurysm of the left anterior communicating artery.The aneurysm max diameter was 3.2mm with a neck diameter of 2mm.The patient's blood flow and vessel tortuosity were normal.It was reported that the echelon-10 microcatheter was in place and the first coil was placed successfully.The second coil was selected for embolization but was not shaped ideally.The microcatheter and axium coil were withdrawn together in order to reshape the microcatheter.When the axium spring coil was pushed out, it was observed that it had already been released and could not be used.The microcatheter was then reshaped and placed and the remaining coils were used normally and the procedure was completed successfully.The patient did not sustain any harm or injury during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Received the patient's weight.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|