Model Number U128
Device Problems Failure to Capture (1081); Interrogation Problem (4017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2020
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that during this right ventricular automatic threshold (rvat) testing, this patient experienced a period of six to ten seconds of asystole.The caller reported a cancelled message on the electrograms and telemetry was intermittent.A boston scientific technical services consultant discussed that telemetry interference could have caused the test to register cancelled but the device kept going with the test until telemetry was lost or regained.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE X4 CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key9981879
MDR Text Key188369136
Report Number2124215-2020-08521
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/06/2021
Device Model NumberU128
Device Catalogue NumberU128
Device Lot Number745059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age61 YR
-
-