Title: intraoperative electrical burn caused by stainless tube tree with noncontact electrosurgical ground -a case report.Authors: sung sik park, jung a lim, and jin-seok yeo.Citation: anesth pain med 2014; 9: 274-276.The authors described an accidental intraoperative electrical burn due to improper use of noncontact grounding pad.A (b)(6)-year-old female patient with rectal cancer and rectovaginal fistula presented for an elective abdominoperineal resection, hysterectomy, vagina posterior wall resection.During the procedure, a third degree burn on the third finger of left hand was found.The patient¿s left third finger was contacted to stainless tube tree on the head side of operation table during prone position and scorched stain was found on the tube tree.A noncontact electrosurgical ground (megadyne 2000; ethicon) was set prior to surgery beneath the patient.The ground was slipped in plastic sheath and covered linen sheet without contact of the patient.The burn site was managed with regular sterile dressings and did not require additional surgical procedure.After this event, we changed tube tree to non-conducting material and insulated other metallic objects on the operation table.The patient¿s burn was completely healed in approximately 2 months.The noncontact grounding pad induced current to tube tree and lower resistance of stainless steel made alternate pathway.The impedance of contact area with body and stainless tube tree had high current density and heat built up resulting in a burn.In conclusion, the authors found an accidental intraoperative electrical burn due to improper use of noncontact grounding pad.Largeness of the pad and the flow of electric current without direct contact to metallic object caused electrical burn in this case.Staffs in operating room should be educated not to place metallic object on the pad and remain vigilant for iatrogenic injuries caused by metallic object on the noncontact grounding pad.".
|