| Model Number |
MMT-523LNAP |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the keypad on the insulin pump was not going anywhere and not moving.Customer stated that the time was advancing.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with no button response at esc, act, up and down due to unlocked lcd keypad connector during visual inspection.Unable to perform operating currents, self test and displacement test or verify battery out limit due to no button response.No button error alarm during testing.
|
| |
|