| Model Number |
MMT-523RNAP |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had button error alarm.Customer was unable to clear the alarm.Customer stated that time was advance.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to corroded battery tube spring contact and corroded battery tube.Unable to perform all functional testing including the operating currents, a21 error test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test or verify button keypad anomaly due to blank display.
|
| |
|