ARTHREX, INC. UNIVERSAL GLENOID - BASEPLATE SMALL; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number UNIVERSAL GLENOID - BASEPLATE SMALL
Health Effect - Clinical Codes Fall (1848); Device Embedded In Tissue or Plaque (3165)
Date of Event 03/20/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that on (b)(6) 2019 a revers shoulder arthroplasty was performed.Everything went well and it was finished successfully.The surgeon was satisfied after the case.This was also showed on the x-rays.Approximately 6 months post-op the patient reported subsequently failing while putting on trousers.He felt an issue.This was when the glenoid side gave way.It was an elective revers and not a fracture revers.X-rays were done on (b)(6) 2020.A revision surgery is planned to be on (b)(6) 2020.Update 02-apr-2020: further information was received: the revision surgery went ahead on (b)(6) 2020 afternoon.It was revised with the new modular glenoid system, successfully using an iliac crest graft.Everything went very well.The failed devices were retrieved, however both peripheral screws were broken and the tips remain in the patient.Baseplate, central screw & remaining portion of peripheral screws x 2 will be returned assembled.Also from the humeral side the cup and liner, which needed to be removed for access, will be returned.There was no issue with the humeral side, however they will be returned regardless.
 
Event or Problem Description
Follow-up update on 29-apr-2020 : further information was provided that the patient was elderly and the bone quality was not in top condition but also not too bad.The damage to the implants occurred (as the patient claims) when putting on trousers, this necessitated the revision, they did not break on removal, they were already broken from whatever trauma they suffered.It was further reported that it is very unlikely that just putting on trousers caused both screws to break.
 
Additional Manufacturer Narrative
Complaint not confirmed, no abnormality was observed on the device that may have contributed to the event.Additional information provided has been added to section b.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL GLENOID - BASEPLATE SMALL
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key9985350
Report Number1220246-2020-01797
Device Sequence Number3675998
Product Code PHX
UDI-Device Identifier00888867057340
UDI-Public00888867057340
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K142863
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberUNIVERSAL GLENOID - BASEPLATE SMALL
Device Catalogue NumberAR-9120-01
Device Lot Number18.01824
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/26/2020
Supplement Date Received by Manufacturer03/26/2020
Initial Report FDA Received Date04/22/2020
Supplement Report FDA Received Date05/07/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-