| Model Number |
C40+ |
| Medical Device Problem Codes |
Circuit Failure (1089); Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user suddenly experienced that the sound was "bad and loud".
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusions: device investigation confirmed that the stimulator electronics failed.The failure of the electronic circuitry was probably caused by humidity ingress at detected micro-leaks at the housing braze joint.In addition, a short circuit between two implant channels was present prior to complete loss of function, for undetermined reasons.Other damages found during investigation are most likely related to explantation surgery.The problems described in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user suddenly experienced that the sound was "bad and loud".The user was re implanted.
|
| |
|