| Model Number |
N/A |
| Medical Device Problem Codes |
Difficult to Advance (2920); Material Deformation (2976)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
03/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
|
| |
|
Event or Problem Description
|
|
The user facility reported that the r2p destination sl guiding sheath was used for a severely curved iliac.The senri balloon catheter could not go through the guiding sheath.Both the balloon catheter and the guiding sheath were withdrawn to be checked, and a slight kink was observed in the sheath.The device was not used and was replaced with another guiding sheath to continue the procedure.The procedure was successfully completed.There was no health damage to the patient.The blood loss was less than 250cc's.There was no patient injury, medical/surgical intervention required.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.With no return of the actual device, the exact cause of the reported event cannot be definitively determined based on the available information.
|
| |
|