MAUDE Adverse Event Report: BARD ACCESS SYSTEMS 4FR DL POWERMIDLINE MAX KIT; MIDLINE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS 4FR DL POWERMIDLINE MAX KIT; MIDLINE CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was unconfirmed because the problem could not be reproduced.The product returned for evaluation was one 4fr d/l powermidline midline catheter.Usage residues were observed throughout the sample.Microscopic inspection of the sample did not reveal any evidence of damage or prior leakage.An attempt to infuse water through the sample using a 12ml syringe revealed both lumens to be patent to infusion and aspiration with no observed leaks.No leaks were observed during sustained (greater than 15 seconds) hydraulic pressurization of the sample.No deficiencies were discovered during evaluation of the returned sample.Consequently this complaint is unconfirmed at this time.A lot history review (lhr) of reds1303 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that the midline catheter was ¿leaking at site¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4FR DL POWERMIDLINE MAX KIT
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key9985735
Report Number3006260740-2020-01477
Device Sequence Number2092808
Product Code PND
UDI-Device Identifier00801741129353
UDI-Public(01)00801741129353
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K162900
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2020
Device Model NumberN/A
Device Catalogue NumberP4254108D
Device Lot NumberREDS1303
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2020
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2020
Initial Report FDA Received Date04/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-