| Model Number |
0102L18 |
| Health Effect - Clinical Code |
Pain (1994)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the foleys were too painful during insertion.It is unknown if the patient was able to void.
|
| |
|
Event or Problem Description
|
|
It was reported that the foleys were too painful during insertion.It is unknown if the patient was able to void.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The reported event could not be confirmed.A potential failure mode could be ¿outside diameter too large¿ with a potential root cause of ¿incorrect catheter buildup and/or finish dipping or incorrect catheter size selected¿.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 15cc balloon: use 20ml sterile water 20cc balloon: use 25ml sterile water 30cc balloon: use 35ml sterile water 40cc balloon: use 45ml sterile water 75cc balloon: use 80ml sterile water.Do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.".
|
| |
|