| Model Number |
86-4181 |
| Medical Device Problem Codes |
Unintended Movement (3026); Migration (4003)
|
| Health Effect - Clinical Codes |
Osteolysis (2377); No Code Available (3191)
|
| Date of Event |
03/31/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the tibial tray fell into a bit of varus and has lysis around it.Femur did not look loose while tibial tray looked loose.Poly wear was suspected.Surgical findings tibial was grossly loose.Cementless femoral component wore through the plastic poly on posterior medial portion and poly.Femur was not grossly loose but did have a large distal medial defect necessitating removal.Patella appeared to be well fixed.Cement manufacturer competitor.Doi: (b)(6) 1996; dor: (b)(6) 2020; right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|