WRIGHTS LANE SYNTHES USA PRODUCTS LLC MATRIXMANDIBLE SHORT CUT PLATE CUTTER; 03.503.057 Back to Search Results
Model Number 03.503.057
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter company employee.A product investigation was conducted.Visual inspection: the matrixmandible short cut plate cutter (p/n: 03.503.057, lot number: l259491) was received at us cq.Upon visual inspection, the cutting head jaws are deformed.Dimensional inspection: no dimensional inspection can be performed due to post manufacturing damage.Document/specification review: relevant drawings were reviewed.No design issues or discrepancies were identified.Investigation conclusion: this complaint is confirmed the cutting head jaws are deformed.No definitive root cause could be determined based on the provided information.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.A device history record (dhr) review was conducted: part: 03.503.057.Lot: l259491.Manufacturing site: (b)(4).Release to warehouse date: 10.Feb.2017.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance's were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, during routine incoming inspection of loaner set it was observed that the matrixmandile short cut plate cutter was damaged.There was no patient involvement.During manufacturer's investigation of the returned device it was identified that the cutting head jaws are deformed.This product condition was reassessed and determined to be reportable on april 3, 2020.This report is for one (1) matrixmandible short cut plate cutter.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MATRIXMANDIBLE SHORT CUT PLATE CUTTER
Common Device Name
03.503.057
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9988199
Report Number2939274-2020-02002
Device Sequence Number14377689
Product Code HTZ
UDI-Device Identifier10887587011783
UDI-Public(01)10887587011783
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.503.057
Device Catalogue Number03.503.057
Device Lot NumberL259491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/03/2020
Initial Report FDA Received Date04/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-