| Model Number |
500-55150 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation of the returned device confirmed that the rf wires had shorted inside of the msd hub.A void was observed in the loctite connecting the strain relief to the white portion of the msd luer lock, which likely contributed to fluid ingress into the hub.It was reported that the msd was replaced, and no adverse patient outcome has been reported.The device short was not reported by the user.No definitive root cause could be identified.Review of the device history record confirmed that the product was manufactured per standard processes and met all acceptance criteria.There have been no other reported complaints from msds of the same lot.No patient harm was reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the microsonic device (msd) catheter would not run any longer than one minute during a dvt case.No fluid ingress or bent pins was found by the customer during troubleshooting.The control unit (cu) and connector interface cable (cic) was exchanged during troubleshooting, but the issue did not resolve.The msd was removed and replaced with a new one.No issues were encountered after exchanging the msd.The patient was reported to be doing fine.On 13 april 2020 a short was noted in the hub of the msd during investigation of the returned device.This was not reported by the customer.
|
| |
|