• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR; DH CPK NG TUBES Back to Search Results
Model Number 40-7431
Medical Device Problem Codes Material Rupture (1546); Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of 22 apr 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported that the device was "placed on (b)(6) 2020, ruptured on (b)(6) 2020.Tube noted to be clogged on (b)(6) 2020, attempted clog zapper with no improvement, kub [kidney, ureter, bladder x-ray] showed kink, pulled back tube 2 cm, clog zapper again, then flushed, (b)(6) 2020 pink coloring of ett [endotracheal tube] secretions which matched meds prev.Given via tube." "kub findings: likely small kink in the region the proximal gastric portion of the enteric tube." "(b)(6) 2020 unchanged feeding tube opening at distal duodenum with kink in gastric portion of tubing." "(b)(6) 2020 continued focal kinking in the proximal gastric portion of the enteric tube." patient was not reported to be injured.Additional information received on 02-apr-2020 indicated "a kink at the 25cm marking-also the tube displays a break above the site of the kink- it is in two pieces-the distal portion was retrieved via endoscopy on (b)(6) 2020." no adverse effects were reported.Customer reported that "tube in for greater than 30 day and that staff may have used force in an attempt to de-clog.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key9990407
Report Number9611594-2020-00070
Device Sequence Number3896062
Product Code KNT
UDI-Device Identifier10680651460322
UDI-Public10680651460322
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 04/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number40-7431
Device Catalogue Number104603200
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2020
Initial Report FDA Received Date04/23/2020
Patient Sequence Number1
Patient Age8 YR
Patient Weight23
-
-