|
We have reviewed the production records of the excor blood pump, s/n (b)(4).This pump was produced according to our specification.The excor blood pump, s/n (b)(4) was in use by the patient from (b)(6) 2019 until (b)(6) 2020 (237 days).The blood pump is designed with a triple layer membrane separating the air chamber from blood chamber for safety reasons.The entire membrane consists of an air-side layer, a middle layer and a blood-side layer.In case of disruption in one of the three layers, there are two more layers that will maintain the integrity of the air and blood chambers.Because the affected blood pump will not be send back to berlin heart, only the pictures provided by the clinic can be used for analysis.The clinic reported that they cut the blood pump open and found blood remains in between the blood-side layer and the middle layer of the triple layer membrane.Blood can only get between the blood-side layer and the middle layer of the triple layer membrane when the blood-side layer has a defect.The cause for the defect in the blood-side layer cannot be determined as the blood pump was not send back for analysis.According to the site report, the blood pump functioned as intended, supporting the patient until the blood pump was exchanged on (b)(6) 2020 without any negative effects on the patient.
|