COOK IRELAND LTD EVOLUTION COLONIC CONTROLLED-RELEASE STENT - UNCOVERED; MQR STENT, COLONIC METALLIC EXPANDABLE Back to Search Results
Model Number G48038
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #: k163468.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
Device failed to deploy "as per complaint form": tight proximal sigmoid stricture wire position good the stent was positioned across the stricture on depressing the trigger to deploy a snapping sound was heard in the gun and stent would not deploy so it was removed and another was used a section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
Device failed to deploy "as per complaint form": tight proximal sigmoid stricture wire position good the stent was positioned across the stricture on depressing the trigger to deploy a snapping sound was heard in the gun and stent would not deploy so it was removed and another was used a section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Pma/510(k) #: k163468.Device evaluation: the evo-25-30-10-c device of lot number c1489172 involved in this complaint device was not returned for evaluation.With the information provided document based investigation was conducted.It was indicated that complaint device was to be returned but has not been returned to date.If it is returned in the future then the file will be updated accordingly.As per descriptions of events "a snapping sound was heard in the gun and stent would not deploy so it was removed and another was used" which indicates that either the flexor inside the handle, the shuttle to sheath tube joint or the shuttle cap may have broken.However, as the device has not been returned, it is not possible to determine which part is affected.Documents review including ifu review: prior to distribution all evo-25-30-10-c devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for evo-25-30-10-c device of c1489172 lot number did not reveal any discrepancies that could have contributed to this issue.There is no evidence to suggest that this issue affects the entire lot #c1489172; upon review of complaints this failure mode has not occurred previously with this lot #c1489172.The instructions for use ifu0052-10 which accompanies this device instructs the user to "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use.Root cause review: a definitive root cause could not be determined from the available information.The break could possibly be a result of compression from tortuous anatomy on the introducer causing a build-up of pressure and subsequently a break in an internal component.However, as the device was not returned for evaluation; the cause of this complaint could not be conclusively determined.Summary: according to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Customer complaint is confirmed based on customer testimony.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION COLONIC CONTROLLED-RELEASE STENT - UNCOVERED
Common Device Name
MQR STENT, COLONIC METALLIC EXPANDABLE
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key9991212
Report Number3001845648-2020-00253
Device Sequence Number4483163
Product Code MQR
UDI-Device Identifier10827002480381
UDI-Public(01)10827002480381(17)200529(10)C1489172
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2020
Device Model NumberG48038
Device Catalogue NumberEVO-25-30-10-C
Device Lot NumberC1489172
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Hospital
Initial Date Received by Manufacturer 03/27/2020
Supplement Date Received by Manufacturer03/27/2020
Initial Report FDA Received Date04/23/2020
Supplement Report FDA Received Date06/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-