Model Number W04200
Medical Device Problem Codes Failure to Cut (2587); Activation Problem (4042)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 07/2021).
 
Event or Problem Description
It was reported that during an activation attempt, the venous catheter allegedly failed to activate.Therefore, both the venous and arterial catheters were removed from the patient without incident and a new set of catheters was used.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a manufacturing review was performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an activation attempt, the venous catheter allegedly failed to cut the tissue.Therefore, both the venous and arterial catheters were removed from the patient without incident and a new set of catheters were used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVELINQ 4 FR
Common Device Name
ENDOVASCULAR ARTERIOVENOUS FISTULA DEVICE
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9992769
Report Number3011392541-2020-00001
Device Sequence Number11911488
Product Code PQK
UDI-Device Identifier00801741182754
UDI-Public(01)00801741182754
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K182796
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberW04200
Device Catalogue NumberW04200
Device Lot NumberS0075
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/02/2020
Supplement Date Received by Manufacturer06/24/2020
Initial Report FDA Received Date04/23/2020
Supplement Report FDA Received Date07/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-