Catalog Number 110137
Device Problem Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/28/2018
Event Type  malfunction  
Event Description
It was reported that, during an ukr surgery, the handpiece jammed when burring the posterior aspect of the femur.It threw the jam error on the screen: after pressing dismiss, the system went back to the connected screen where the handpiece re-calibrated just fine, then went to home, but the drill did not budge an inch.Only the handpiece was swapped out and the rest of the case transferred without any problem.Surgery was not significantly delayed.The patient was not harmed.The results of investigation have pointed out that the snaplock assembly was broken, which makes it a reportable event.
 
Manufacturer Narrative
The navio handpiece, used in treatment, was jammed and would not home during a procedure.A device history record review found no conditions which could contribute to the reported event and the device met all manufacturing specifications during release for distribution.A complaint history review found similar reports and this issue will continue to be monitored.The navio handpiece was returned for further evaluation and the initial visual/functional inspection was performed, which confirmed the reported event.The snap lock nut was broken.A broken snap lock nut would prevent the snap lock from moving along the lead screw, causing the jam and homing issue.The root cause of this issue was found to be mechanical component failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, NAVIO (BBT HANDPIECE), ROHS COMPLIANT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth, mn
MDR Report Key9996100
MDR Text Key188859481
Report Number3010266064-2020-00674
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number110137
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-