• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number TCM1510X3
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Muscle Weakness/Atrophy (1967); Pain (1994); Hernia (2240); Injury (2348); Abdominal Distention (2601); No Code Available (3191); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced recurrence, pain, diastasis, separation of the rectus muscles, bulge, laxity in upper-abdomen, and dense adhesions.Post-operative patient treatment included revision surgery, hernia repair with mesh, resection of hernia, and lysis of adhesions.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced recurrence, pain, diastasis, separation of the rectus muscles, bulge, laxity in upper-abdomen, and dense adhesions.Post-operative patient treatment included revision surgery, hernia repair with mesh, resection of hernia, ct-scan, medication, and lysis of adhesions.Relevant tests/laboratory data: on (b)(6) 2018: op note- ct scan showed significant diastasis and separation of the rectus muscles, approximately 7 x 7cm in size, with internal bulge consistent with recurrent ventral hernia versus severe laxity.
 
Additional Manufacturer Narrative
Additional information: b5, b7, d8, e1 (facility name, street 1, city, region, postal code), g1(manufacturer name, first name, last name, street 1, city, region, postal code, email, phone), h6 (patient codes).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARIETEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
60 middletown ave.
8200 coral sea st ne
mounds view,mn, CT 55112
7635265677
MDR Report Key9996180
Report Number9615742-2020-00988
Device Sequence Number19922345
Product Code FTL
UDI-Device Identifier10884521178861
UDI-Public10884521178861
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K090858
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2017
Device Model NumberTCM1510X3
Device Catalogue NumberTCM1510X3
Device Lot NumberSMI00655
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2020
Supplement Date Received by Manufacturer02/22/2023
02/14/2024
Initial Report FDA Received Date04/24/2020
Supplement Report FDA Received Date02/27/2023
02/27/2024
Date Device Manufactured10/09/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK-PERMACOL (LOT #: UNKNOWN).; UNK-PERMACOL (LOT #: UNKNOWN)
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight62 KG
Patient EthnicityHispanic
-
-