| Model Number |
DF56 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Siemens healthcare diagnostics headquarters support center (hsc) concluded their investigation of the discordant, falsely depressed lipase (lipl) result on a dimension exl system.Siemens hsc has reviewed the information provided by the customer and the dimension exl instrument data.The cause of the event is attributable to operator error in making the 1/3 dilution that generated the erroneous result.There is no indication of a assay or instrument malfunction.The customer is operational.The device is performing within specifications.No further evaluation is required.
|
| |
|
Event or Problem Description
|
|
A discordant, falsely depressed lipase (lipl) result was obtained on a diluted patient sample on a dimension exl system.The result was reported to the physician(s).The same sample was reprocessed on an alternate dimension exl and a higher result was obtained, considered correct, and reported.There are no known reports of patient intervention or adverse health consequences due to the discordant falsely depressed lipase result.
|
| |
|