• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO SINGLE TRIGGER WIRE COLLET; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO SINGLE TRIGGER WIRE COLLET; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 8203036000
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/31/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 event was reported for this quarter.Product return status: 1 device was received.Event confirmation status: 1 reported event was confirmed.Evaluation results: 1 device was found to be affected by a detached and/or missing component.Additional information: 1 device was not labeled for single-use.1 device was not reprocessed or reused.
 
Event or Problem Description
This report summarizes 1 malfunction event in which the device had a component detach.1 event had no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE TRIGGER WIRE COLLET
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key9997318
Report Number0001811755-2020-01091
Device Sequence Number4442523
Product Code HWE
UDI-Device Identifier07613327304497
UDI-Public07613327304497
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8203036000
Device Catalogue Number8203036000
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/31/2020
Initial Report FDA Received Date04/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-