BLUE BELT TECHNOLOGIES NAVIO TKA PLANE VISUALIZATION TOOL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110005
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/05/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that plate probes need to be replaced due to fit issues.No patient involved.
 
Additional Manufacturer Narrative
Additional info: d10, d11, g4, g7, g9, h2, h3, h6, h10.Results of investigation: the device was being used.During treatment and was returned for evaluation.The functional inspection found that the plate probe could not be locked into the collet and was slightly tight when removing it.The dhr was reviewed for (b)(4) /lot c12834-2-1 and c13069-1-2 and found that they were related to an non conformance which was for fit issues between the handpiece and visualization tool.There were no other issues that would potentially lead to a future performance issue.A complaint history review was performed and found 36 reports of the issue.This issue will continue to be monitored.A relationship between the device and reported event has been established.Factors that could have contributed to the reported event is that the plate probe had an incorrect notch curvature.Therefore, this issue prevented the plate probe from fully locking in and it was not machined with sufficient space to allow proper locking with the handpiece.The reported event can be confirmed.Therefore, the root cause of the reported event was due to supplier/raw material fault.There has been a corrective action opened to address this issue and it is being further investigated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO TKA PLANE VISUALIZATION TOOL
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN
MDR Report Key9999141
Report Number3010266064-2020-01353
Device Sequence Number11371900
Product Code OLO
UDI-Device Identifier00885556628522
UDI-Public00885556628522
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K191223
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPFSR110005
Device Catalogue NumberROB00004
Device Lot NumberC13069-1-2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2019
Initial Date Received by Manufacturer 08/05/2019
Supplement Date Received by Manufacturer08/05/2019
Initial Report FDA Received Date04/25/2020
Supplement Report FDA Received Date08/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NPFS02020 NAVIO SURGICAL SYSTEM.
-
-