• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OSTEORAPTOR CRVD 2.3 SA UB COBRD BLUE; STAPLE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OSTEORAPTOR CRVD 2.3 SA UB COBRD BLUE; STAPLE, FIXATION, BONE Back to Search Results
Model Number 72203290
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that during surgery the osteoraptor gave of lots of metal debris when drilling anchor holes, all the metal was removed from the joint.The procedure was successfully completed using a back-up device.A delay of 5 minutes was reported.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10: the reported 2.3 osteoraptor curved sa, intended for use in treatment, has not been returned for evaluation.Without the reported product a visual or functional evaluation cannot be performed and the customers complaint cannot be confirmed.From the information provided, the complaint was filed for shedding during drilling.This device has no drilling action associated with it.An exact root cause cannot be determined with confidence without the return of the devices in question.Per communication, all of the metal debris was removed from the joint using a shaver, a backup device was used to complete the procedure without a significant delay.No additional bone holes were needed.Since no patient harm is being alleged and no further harm is anticipated, no further clinical assessment is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSTEORAPTOR CRVD 2.3 SA UB COBRD BLUE
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key9999638
Report Number1219602-2020-00693
Device Sequence Number3071089
Product Code JDR
UDI-Device Identifier00885554022254
UDI-Public00885554022254
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2023
Device Model Number72203290
Device Catalogue Number72203290
Device Lot Number50742590
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/02/2020
Supplement Date Received by Manufacturer05/29/2020
Initial Report FDA Received Date04/27/2020
Supplement Report FDA Received Date06/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-