| Class 2 Device Recall Brilliance CT series, iCT, iCT SP, Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex, | |
Date Initiated by Firm | April 01, 2014 |
Date Posted | June 04, 2014 |
Recall Status1 |
Terminated 3 on June 26, 2018 |
Recall Number | Z-1722-2014 |
Recall Event ID |
68245 |
510(K)Number | K010817 K012009 K033326 K033357 K060937 |
Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
Product | Philips Medical Systems Computed Tomography X-ray systems (Brilliance CT series, iCT, iCT SP, Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex, and the MX8000 Dual v. EXP scanners) |
Code Information |
Brilliance CT 6: M/N: 728256; S/N: 3001, 3002, 3005, 3007, 3014, 3016, 3019-3021, 3025, 3026, 3028, 3032-3036, 3045, 3046, 3049, 3051, 3052, 3057, 3058, 3060, 3062, 3064-3066, 3068, 3071, 3073, 3074, 3076-3081, 3083, 3084, 3090, 3093-3095, 3097, 3101, 3102, 3104, 3106-3108, 3112-3115, 3118-3121, 3124, 3126, 3128, 3129, 3131, 3132, 3134, 3136, 3137, 3139-3141, 3143-3145, 3147, 3148, 3151-3158, 3162, 3165-3169, 3171, 3173-3177, 3180, 3181, 3184-3190, 3192, 3193, 3196, 3200, 3201, 3203, 3204, 3207, 3208, 3210-3218, 3220, 3221, 3225, 3226, 3228, 3229, 3232, 3235, 3237, 3240, 3241, 3243-3245, 3248, 3249, 3253-3257, 3260-3264, 3266-3272, 3275-3282, 3284, 3285, 3287, 3290, 3292, 3296, 3299, 3301, 3303-3305, 3307-3311, 3313, 3315, 3317-3323, 3325-3328, 3332-3334, 3336, 3338, 3339, 3341, 3343-3345, 3347-3349, 3351, 3352, 3354, 3357-3364, 3366-3368, 3370-3374, 3377-3384, 3386, 3388-3391, 3394, 3396-3400, 3402-3404, 3408, 3409, 3412, 3413, 3415-3418, 3423-3426, 3428-3432, 3434-3437, 3439, 3443, 3444, 3447, 3449, 3451-3453, 3455, 3457, 3459, 3460, 3461, 3463, 3465, 3467-3475, 3477-3483, 3485, 3486, 3488, 3489, 3491-3506, 3508-3511, 3513, 3515, 3517-3519, 3522-3524, 3526, 3527, 3532-3535, 3537-3549, 3551-3553, 3556-3559, 3562, 3564, 3566, 3568, 3570, 3571, 3573-3575, 3577, 3578, 3580, 3582-3588, 3590-3595, 3597-3608, 3611-3616, 3618-3622, 3625-3632, 3634-3637, 3639-3646, 3648, 3649, 3651-3661, 3663-3665, 3667-3670, 3672, 3674-3677, 3679-3693, 3695-3698, 3703-3712, 3714, 3715, 3717-3720, 3722, 6039, 6040, 30003, 30005, 30007-30009, 30011-30018, 30021, 30022, 30024, 30026-30031, 30033-30040, 30042, 30043, 30045-30048, 30050, 30052-30054, 30056-30062, 30064-30066, 30068, 30069, 30072, 30073, 30075, 30077-30080, 30082-30090, 30092, 30094, 30100, 30101, 30104-30108, 30111, 30113, 30117-30119, 30121-30131 & 300043. Brilliance CT 10: M/N: 728251; S/N: 3008, 3098, 4001, 4003, 4005, 4006, 4009-4012, 4014, 4016, 4020, 4021, 4023, 4026, 4027, 4029-4031, 4035, 4040, 4041, 4044, 4047-4050, 4052-4054, 4058, 4059, 4061, 4063-4067, 30099, 40006-40012, 40014-40018, 40021, 40023, 40024, 40028, 40030 & 40032-40035. Brilliance CT 16: M/N: 728246; S/N: 3011, 3013, 3018, 3027, 3031, 3038, 3041, 3043, 3070, 3072, 3082, 3085, 3117, 3127, 3149, 3160, 3205, 3216, 3227, 3293-3295, 3302, 3306, 3324, 3336, 3376, 3393, 3466, 3490, 3514, 3521, 3563, 3565, 3624, 3647, 3673, 3721, 4004, 4017, 4018, 4019, 4024, 4025, 4036, 4043, 4055, 5001, 5004-5008, 5010-5021, 5023-5026, 5028-5032, 5034-5038, 5040-5042, 5044-5046, 5049-5055, 5058-5062, 5064, 5066-5070, 5072-5076, 5078, 5081-5086, 5088-5095, 5097-5099, 5101-5105, 5107, 5108, 5110-5118, 5120-5128, 5130-5141, 5143, 5145-5152, 5155, 5156, 5158, 5160-5164, 5166-5170, 5173-5181, 5183-5202, 5204, 5207-5209, 5211-5219, 5221-5226, 5228-5240, 5242, 5243, 5248, 5249, 5252-5254, 5258, 5264-5267, 5269-5275, 5278, 5279, 5281-5283, 5285-5297, 5301-5309, 5311-5313, 5315-5320, 5322, 5323, 5325-5332, 5334, 5336, 5338-5344, 5346-5351, 5353, 5355, 5357, 5359-5362, 5365-5369, 5371, 5372, 5375-5380, 5382, 5384, 5386-5389, 5394, 5396-5400, 5402-5410, 5414-5429, 5432-5435, 5437-5443, 5451, 5452, 5456, 5460, 5468, 5469, 5485, 5489, 5515, 5522, 5531, 5535, 5537, 5543, 5552, 5555, 5557, 5567, 5581-5584, 5586-5588, 5600, 5607, 5621, 5624, 5630, 5642, 5644, 5645, 5651, 5654, 5658, 5663, 5664, 5667, 5675, 5679, 5684, 5685, 5697, 5707, 5709, 5716, 5724, 5725, 5735, 5738, 5743-5745, 5748-5750, 5752, 5754, 5756, 5772, 5774, 5776, 5777, 5780, 5782, 5790, 5793, 5794, 5796, 5801, 5802, 5806, 5807, 5812-5815, 5823-5827, 5830-5834, 5840-5849, 5851, 5852, 5854, 5855, 5857, 5858, 5860, 5861, 5866, 5871-5873, 5880, 5881, 5882, 5884-5886, 5892, 5899, 5901, 5904, 5915, 5921-5923, 5939-5942, 5947, 5948, 5961-5963, 5966-5968, 5976, 5980, 6011, 6012, 6014, 6015, 6028, 6033, 6034, 6036-6038, 6045, 6050, 6068-6071, 6082, 6090, 6095, 6104, 6105, 6108, 6112, 6115, 6126, 6132, 6133, 6161-6165, 6173, 6180, 6221, 6240, 6241-6243, 6245-6257, 6259-6262, 6264-6273, 6275-6277, 6279-6283, 6285, 6287-6304, 6306-6317, 6319-6325, 6327-6364, 6366-6370, 6373-6380, 6382-6387, 6389, 6392-6395, 6397-6416, 6418-6421, 6423-6444, 6447-6458, 6460-6493, 6497-6500, 6502-6532, 6534-6563, 6565-6577, 6580-6590, 6592-6616, 6618-6636, 6638-6640, 6643-6645, 6647, 6648, 6650-6655, 6657, 6658, 6660-6671, 6673, 6675-6718, 6720-6726, 6729, 6733, 6738, 6740-6743, 6745, 6751, 6753-6755, 6757- 6760, 6762-6768, 30019, 30041, 30061, 30069, 30081, 30091, 30093, 30097, 30109, 32287, 40002, 40003, 40020, 40022, 40027, 50002, 50003, 50005, 50007-50013, 50015, 50016, 50018, 50021-50023, 50025-50036, 50038, 50040-50042, 50044-50059, 50061-50070, 50073-50088, 50091-50099, 50101-50107, 50109-50113, 50115, 50116, 50118-50122, 50124-50129, 50131, 50133-50138, 50141-50166, 50168, 50170, 50172, 50175-50182, 50184-50195, 50197-50200, 50202, 50204, 50206-50208, 50210 & 270001. Brilliance CT 16 Power: M/N: 728240; S/N: 3004, 3046, 3100, 3109, 3236, 5003, 5043, 5047, 5071, 5345, 5356, 5374, 5390, 5395, 5413, 5430, 5455, 5467, 6007, 6009-6011, 6013, 6024, 6026-6028, 6031-6038, 6041-6043, 6045-6049, 6051-6060, 6062-6067, 6069-6072, 6075, 6077-6090, 6093, 6095, 6096, 6099-6113, 6115-6117, 6119, 6121-6126, 6129-6137, 6139, 6141-6151, 6153-6156, 6159, 6161-6163, 50039, 50073, 60001-60006, 60008-60014, 60017, 60022, 60024, 60025 & 60027-60029. Brilliance CT 40: M/N: 728235; S/N: 4012, 9031, 9032, 9034, 9036, 9039, 9046, 9047, 9050, 9053, 9055, 9057, 9060, 9061, 9076, 9085, 9088, 9094, 9095, 9097, 9103, 9109, 9111, 9113-9116, 9119, 9122, 9124, 9125, 9128-9134, 9136, 9141, 9144-9147, 9149, 9150, 9153, 9154, 9157, 9158, 9161, 9164, 9165, 9167-9170, 9172, 9174, 9176, 9178-9183 9185, 9186, 9188, 9189, 9191-9194, 9198-9200, 9203,9204, 9207,9209-9212, 9219, 9222, 9225-9233, 9237-9241, 9244, 9245, 9247-9251, 9253, 9254, 9256-9258, 9260-9273, 9275, 9276, 9278-9282, 9284-9286, 29001, 29004, 29006-29008, 29010-29013, 29015, 29019-29024, 29028-29031, 29033, 29036, 29038-29041, 29043, 29046-29048, 29050, 29051, 29055, 29058-29061, 29063-29067, 29069, 29071-29073, 29075, 29077, 29078, 29080, 29082-29087, 29089-29091, 29093-29096, 29098-29106, 29108, 29109, 29113, 29114, 29116-29123, 29125, 29127-29132, 29134-29137, 29139,29140-29145, 29147-29150, 29152, 29153, 90034, 90035, 90040, 90043, 90044, 90049, 90051-90054, 90056, 90059, 90061, 90064, 90065, 90067, 90069, 90071, 90075, 90076, 90083, 90084, 90087, 90090, 90093, 90095, 90101, 90103, 90108-90110, 90121, 90122, 90125, 90126, 90130, 90138, 90142, 90144, 90146, 90148, 90149, 90155, 90157-90159, 90164-90166, 90169, 90171-90174, 90178, 90179, 90181, 90183, 90184, 90186, 90189, 90190, 90192-90194, 90196-90198, 90200, 90202-90204, 90206-90215, 90219, 90220, 90222, 90224, 90226, 90228, 90229, 90231-90237 & 92034. Brilliance CT 64: M/N: 728231; S/N: 4001, 4003, 4016, 4020, 5006, 9029, 9030, 9033-9035, 9037, 9038, 9040, 9041, 9043-9045, 9048, 9049, 9051, 9052, 9054, 9056, 9058, 9059, 9062, 9064-9075, 9077, 9080, 9083, 9086, 9087, 9089, 9092-9094, 9098, 9099, 9104-9108, 9112, 9117, 9118, 9120, 9121, 9126, 9127, 9135, 9137, 9138, 9140-9143, 9148, 9151, 9152, 9155, 9156, 9159, 9160, 9162, 9163, 9166, 9171, 9173, 9175, 9177, 9187, 9190, 9195-9197, 9201, 9202, 9205, 9206, 9208, 9213-9218, 9235, 9236, 9243, 9255, 9259, 9501-9505, 9507-9512, 9514-9531, 9533-9536, 9538-9543, 9545-9552, 9554-9566, 9568, 9569, 9570, 9572-9585, 9587-9596, 9598-9608, 9610, 9612-9620, 9622-9655, 9657-9659, 9662-9668, 9670-9674, 9676-9680, 9683-9695, 9697-9703, 9705-9714, 9716, 9718-9720, 9722-9726, 9728-9732, 9734-9742, 9745-9761, 9763, 9765-9769, 9772-9782, 9784-9795, 9797, 9800, 9801, 9803-9813, 9815-9822, 9824, 9826, 9829, 9830, 9832-9835, 9837-9847, 9849-9853, 9858-9871, 9873-9879, 9881-9885, 9888-9896, 9898, 9899, 9901-9906, 9909-9911, 9913-9927, 9929-9934, 9936-9938, 9940-9942, 9944-9948, 9950-9956, 9958-9961, 9963-9971, 9974-10009, 10011-10028, 10031-10037, 10039-10042, 10044-10054, 10056-10073, 10075, 10077-10081, 10083-10090, 10092, 10094-10108, 10110-10132, 10134-10137, 10139-10146, 10148-10158, 10160-10162, 10164-10168, 10170-10174, 10176-10198, 10200-10235, 10237-10239, 10241-10245, 10247, 10248, 10250-10283, 10286-10288, 10291-10305, 10307-10319, 10321-10327, 10330-10339, 10341-10353, 10355, 10357-10381, 10383-10386, 10388-10402, 10404-10422, 10424-10432, 10435-10492, 10494-10500, 10502-10523, 10525-10528, 10530-10535, 10537-10543, 10545-10619, 10621-10671, 10673, 10675-10681, 10683-10689, 10691-10708, 10710, 10712-10715, 10717, 10718, 10720-10745, 10747-10762, 10765, 10767-10786, 10789, 10790, 10792, 10793, 10795-10804, 29002, 29003, 29005, 29014, 29016-29018, 29025, 29026, 29032, 29034, 29035, 29037, 29044, 29045, 29049, 29053, 29054, 29056, 29057, 29062, 29066, 29068, 29070, 29092 ,29101, 29107, 29113, 29115, 29124, 29138, 29151, 32129, 64015, 64016-64020, 64022-64039, 64041-64080, 64082-64094, 64096-64104, 64106-64112, 64115-64118, 64121-64131, 64133-64140, 64142, 64143-64154, 64157, 64158, 64160, 64162-64169, 64171-64175, 65006, 90029-90031, 90036, 90037, 90041, 90042, 90046-90048, 90057, 90058, 90060, 90062, 90063, 90066, 90070, 90072, 90074, 90077-90082, 90084-90089, 90091, 90092, 90094, 90096-90100, 90105-90107, 90117-90120, 90123, 90124, 90129, 90131, 90132, 90135-90137, 90139-90141, 90143, 90147, 90150-90154, 90156, 90160, 90161, 90163, 90167-90170, 90175-90177, 90180, 90182, 90187, 90195, 90199, 90205-90208, 90213, 90218, 90221-90223, 91003, 95003, 95008, 95009, 95011-95024, 95029-95052, 95054-95069, 95071-95076, 95078-95085, 95088-95095, 95097-95104, 95106-95114, 95116-95119, 95121-95123, 95125-95129, 95132-95143, 95147, 95149, 95151, 95153, 95154, 95157, 95161, 95162, 95165-95170, 95177, 95178, 95181-95186, 95189-95191, 95194, 95195, 95197, 95199, 95201-95204, 95207-95210, 95214, 95216-95219, 95221-95223, 95227, 95228, 95231-95234, 95236-95239, 95242, 95243, 95245-95247, 95249-95254, 95257-95259, 95261, 95262, 95264-95266, 95268-95271, 95273-95287, 95289-95294, 95296-95298, 95300, 95301, 95303-95305, 95308-95310, 95312-95335, 95338, 95342, 95344-95346, 95348-95351, 95353-95366, 95369, 95370, 95374-95376, 95378-95400, 95402-95404, 95406, 95408-95411, 95413-95436, 95438-95440, 95442-95458, 95460-95492, 95494-95529, 95531-95584, 95586, 95588-95592, 95595-95641, 95643-95662, 95664-95700, 95702-95727, 95729-95761, 95764-95770, 95772-95812, 95814-95832, 95835-95845, 95848-95858, 95860-95862, 95864-95876, 95878, 95879, 95884, 95920, 96494, 100062, 110058, 110391 & 999999. Brilliance CT Big Bore Oncology: M/N: 728243; S/N: 7006-7009, 7013-7051, 7053-7062, 7064-7120, 7122-7142, 7144-7166, 7168, 7169, 7171, 7173-7175, 7177-7179, 7181, 7183-7186, 7188, 7189, 7191, 7192, 7194-7199, 7201-7214, 7216-7243, 7245-7247, 7249, 7250, 7253-7263, 7265-7267, 7269, 7270, 7272-7289, 7291-7295, 7297-7300, 7302, 7304, 7306-7316, 7319, 7320, 7322-7374, 7376-7379, 7381,7382, 7387-7389, 7392, 7393, 7395, 7397-7401, 7403-7406, 7409,7410,7411-7422, 7424-7426, 7428, 7429, 7431-7433, 7435-7438, 7440-7451, 7454-7457, 7459, 7461-7471, 7473-7475, 7477-7485, 7487-7509, 7511-7513, 7515-7524, 7526-7528, 7530-7532, 7534, 7535, 7537, 7539-7542, 7545, 7547-7556 ,7558-7561, 7563-7569, 7573, 7574,7577-7582, 7584, 7585, 7588-7593, 7596-7599, 7601-7603, 7605, 7606, 7608, 7609, 7611-7613, 7615-7618, 7620, 7622-7624, 7626-7630, 7634, 7635, 7637-7650, 7652, 7655-7659, 7661, 7662, 7665-7667, 7671, 7672, 7674, 7675, 7677-7681, 7684, 7685, 7687, 7689-7691, 7693, 7695, 7696, 7698-7707, 7709, 7715, 7716, 7718, 7720, 7724-7727, 7729-7731, 7735-7737, 7739-7741, 7743-7745, 7749-7755, 7757, 7759, 7761-7763, 7765, 7769, 7770, 7774-7777, 7779, 7782-7788, 7791, 7793-7797, 7799-7803, 7806-7808, 7810-7812, 7814, 7815, 7817, 7818, 7820, 7822-7824, 7827-7829, 7831-7838, 7840-7842, 7846-7852, 7855-7858, 7860, 7861, 7864-7868, 7872, 7873, 7875-7877, 7881, 7884, 7886, 7891-7893, 7895, 7896, 7898, 7900-7902, 7904, 7905, 7907-7911, 7914, 7916, 7918, 7920, 7921, 7923-7925, 7927-7929, 7931-7933, 7935-7942, 7944-7946, 7948-7950, 7955, 7956, 7963, 7968, 7970, 7972, 7973, 7976, 7984, 7987, 7990, 7991 & 7993. Brilliance CT Big Bore Radiology: M/N: 728244; S/N: 7052, 7063, 7121, 7155, 7172, 7176, 7180, 7182, 7190, 7200, 7244, 7252, 7268, 7271, 7301, 7318, 7375, 7380, 7383, 7384, 7386, 7390, 7396, 7402, 7407, 7408, 7427, 7430, 7439, 7452, 7453, 7458, 7486, 7510, 7514, 7525, 7529, 7533, 7538, 7543, 7546, 7570, 7572, 7575, 7583, 7586, 7587, 7594, 7595, 7604, 7610, 7614, 7619, 7621, 7625, 7631, 7632, 7633, 7636, 7651, 7653, 7654, 7660, 7668-7670, 7676, 7683, 7686, 7688, 7692, 7694, 7697, 7708, 7710-7714, 7717, 7719, 7721-7723, 7732-7734, 7738, 7742, 7746, 7747, 7756, 7760, 7764, 7766, 7767, 7771-7773, 7778, 7780, 7781, 7789, 7790, 7798, 7804, 7809, 7813, 7819, 7825, 7826, 7830, 7839, 7844, 7845, 7853, 7859, 7862, 7869, 7870, 7874, 7878, 7879, 7880, 7882, 7887, 7888, 7894, 7897, 7899, 7903, 7906, 7912, 7913, 7915, 7917, 7919, 7922, 7926, 7930, 7934, 7943, 7952,7958, 7959, 7964, 7967, 7986, 7988 & 7992. Brilliance iCT: M/N: 728306; S/N: 85000, 100016, 100019-100025, 100028, 100031, 100032, 100037-100040, 100042-100074, 100077-100079, 100081-100084, 100086-100106, 100108-100135, 100137-100142, 100144-100155, 100157-100182, 100184-100236, 100238, 100240, 100242-100254, 100256-100268, 100270-100278, 100280-100283, 100285-100297, 100299-100325, 100327, 100329-100332, 100334, 100336, 100338-100343, 100345-100350, 100352-100358, 100360-100365, 100368-100380, 100382-100389, 100391, 100393-100402, 100404-100412, 100414-100418, 100423-100426, 100429, 100430, 100433-100435, 100438-100442, 100449-100458, 100460-100462, 100464-100470, 100473, 100476, 100477, 100480-100482, 100484-100486, 100488, 100490, 100493, 100494, 100497-100499, 100503, 100505-100508, 100511, 200005, 200083, 200098 & 987654. Brilliance iCT SP: M/N: 728311; S/N: 100039, 100062, 200003, 200006-200016, 200018-200027, 200029-200058, 200060-200065, 200067, 200068-200071, 200073-200079, 200084-200097, 200099-200117, 200119, 200120, 200123-200131, 200133-200137. Ingenuity CT: M/N: 728326; S/N: 6372, 6395, 6569, 10607, 300005, 300011, 300012, 300014, 300016-300022, 300024, 300025, 300027-300042, 300044-300066, 300068, 300069, 300071-300075, 300077-300082, 300084-300114, 300116-300133, 300135-300140, 300142-300145, 300147-300151, 300156-300163, 300164 & 310058. Ingenuity Core: M/N: 728321; S/N: 6563, 7700, 10614, 10621, 10633, 50171, 50204, 95732, 100507, 310001, 310004, 310006, 310008, 310009, 310010, 310012-310015, 310017, 310018, 310020-310022, 310024-310026, 310028-310035, 310037-310057, 310059-310143, 310146-310148, 310156, 310157, 310159, 333001-333003, 333005-333012, 333016 & 3100016. Ingenuity Core128: M/N: 728323; S/N: 100315, 320002-320011, 320013-320015, 320017-320112, 320115, 320119, 320123, 320126, 320131-320133, 320135, 320136, 320140, 336001-336003, 336006-336008 & 336014. MX8000 Dual v. EXP: M/N: 728130; S/N: 2110, 8010, 8016, 8018, 8034, 8035, 8039-8041, 8043, 8048, 8053, 8055, 8056, 8058, 8061, 8065, 8067, 8068, 8070, 8073-8077, 8079, 8081, 8082, 8084-8086, 8092, 8095, 8097, 8098, 8104, 8105-8108, 8111, 8112, 8114-8120, 8123-8125,8127-8131, 8134-8138, 8140, 8142-8147, 8149-8151, 8153-8156, 8159, 8160, 8163-8165, 8167-8169, 8171-8175, 8177, 8178, 8181, 8184-8186, 8188, 8190-8192, 8194-8196, 8198, 8199, 8201-8208, 8211-8213, 8215-8218, 8220, 8221, 8223, 8225-8227, 8232-8240, 8242, 8243, 8245, 8247-8250, 8253, 8256-8258, 8260-8289, 8292-8296, 8298-8306, 8308, 8310-8315, 8317, 8318, 8320-8324, 8326, 8328-8335, 8337-8342, 8344-8347, 8349, 8351-8355, 8357-8359, 8361-8373, 8375-8396, 8398, 8400-8407, 8409-8413, 8442, 9528, 80002-80005, 80007, 80009-80019, 80021 & 80022. Ingenuity Flex: M/N728317; S/N: 260002, 260003, 340002, 340003, 345002, 345003 & 345004. |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Rd Cleveland OH 44143-2131
|
For Additional Information Contact | Sr. Manager, Post-Market Surveillance 440-483-3032 |
Manufacturer Reason for Recall | A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged. |
FDA Determined Cause 2 | Component design/selection |
Action | Philips Healthcare sent an Urgent Medical Device Correction letter dated April 1, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed If they experience a horizontal, free-floating couch motion, to contact their Field Service Engineer immediately. Customers were instructed to retain a copy of this Field Safety Notice with the equipment Instructions for
USER Use (IFU). Customers with questions or needing further information were instructed to contact their local Philips representataive or local Philips Healthcare office. For North America, contact the Customer Care Solutions Center at 1-800-722-9377, Option 5.
For questions regarding this recall call 440-483-3032. |
Quantity in Commerce | 6, 839 units |
Distribution | Worldwide Distribution: US nationwide including PR and DC, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Canada, Cambodia, Cape Verde, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dubai, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Greenland, Guadeloupe, Herzegovina, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, LAO People's Dem, Rep., Latvia, Lebanon, Libya, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malawi, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nahar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam & Yemen. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAK 510(K)s with Product Code = JAK 510(K)s with Product Code = JAK
|
|
|
|