Class 2 Device Recall System 6 Aseptic Housing |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
January 07, 2015 |
Date Posted |
February 09, 2015 |
Recall Status1 |
Terminated 3 on March 09, 2016 |
Recall Number |
Z-1066-2015 |
Recall Event ID |
70298 |
Product Classification |
Instrument, surgical, orthopedic, ac-powered motor and accessory/attachment - Product Code HWE
|
Product |
System 6 Aseptic Housing.
The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker Non-sterile Battery. Accessory for powered orthopedic surgical instrument.
|
Code Information |
Part No. 6126-120-000, Lots 13205, 13209, 13210, 13212 |
Recalling Firm/ Manufacturer |
Stryker Instruments Div. of Stryker Corporation 4100 E Milham Ave Portage MI 49002-9704
|
For Additional Information Contact |
Kara Spath 269-289-4518
|
Manufacturer Reason for Recall |
Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Stryker sent notice to customers via Fed ex overnight on 1/16/2015, regarding its voluntary recall of Stryker System 6 Aseptic Housings because the laser welder may have failed to produce a continuous bond line for the four specific lots: 13205, 13209, 13210, 13212. Customers are instructed to discontinue use, complete the business reply form, and return all recalled product. Customers may contact Kara Spath at 269-389-4518, or kara.spath@stryker.com. |
Quantity in Commerce |
399 |
Distribution |
Worldwide Distribution-US (nationwide) including Puerto Rico and the states of AR, CA, CO, FL, GA, HI, MD, NC, NY, PA, SC, TX, and VA, and the countries of Australia, Canada, Switzerland, Hong Kong, South Africa, Chile, Korea, Poland, India, Colombia, The Netherlands, and France. |