Date Initiated by Firm | August 10, 2021 |
Create Date | September 24, 2021 |
Recall Status1 |
Terminated 3 on September 16, 2024 |
Recall Number | Z-2560-2021 |
Recall Event ID |
88531 |
510(K)Number | K922637 K931377 |
Product Classification |
Set, administration, intravascular - Product Code FPA
|
Product | Outlook Pump Blood Administration Set with SAFELINE Injection Site-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids
Ref: NF3120 |
Code Information |
Lot Code:
0061769045
DI: 04046964187322
|
Recalling Firm/ Manufacturer |
B. Braun Medical, Inc. 901 Marcon Blvd Allentown PA 18109-9512
|
For Additional Information Contact | SAME 610-596-0500 |
Manufacturer Reason for Recall | Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections. |
FDA Determined Cause 2 | Under Investigation by firm |
Action | BBraun issued Urgent Medical Device Correction letter to direct end customers and BBMI distributors, dated 8/10/21. Letter states reason for recall, health risk and action to take:
Review the Device Recall Notification in its entirety and ensure that all users in your organization of the impacted product, and other concerned persons, are informed about this voluntary product recall. If you are a distributor, please forward this recall notification to your customers. The recall is to be extended to the customer level.
2. Copies of Attachment 02 should be distributed to all personnel and staff who utilize Outlook Pump Sets within their care area to ensure they identify impacted units prior to use. DO NOT use any Impacted units identified by the Instructions. Discard any Identified products to ptevent Inadvertent use.
3. B. Braun does not recommend return of this material at this time as it may lead to shortages of supply which may adversely impact patients. Not all pump sets within the identified lot are impacted. If your facility has both impacted and non-impacted lots of material, you may wish to direct impacted lots to care areas where less critical medications are administered or where risks of exposure to hazardous (e.g. cytotoxic) medications may be decreased.
4. Complete the attached "Product Correction Acknowledgement," form and return it to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to PA_QualityAssurance.BBMUS_Service@bbraunusa.com within two (2) weeks of receipt.
5. To receive credit for any units which you have had to discard via this process, please contact customer support by calling 1-800-227-2862. Please be sure to identify your local distributor when contacting customer support so we can ensure that proper credit is provided |
Quantity in Commerce | 336 units |
Distribution | Worldwide distribution - US Nationwide distribution and the country of Canada. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = FPA
|