• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ICY Intravascular Heat Exchange Catheter Kit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ICY Intravascular Heat Exchange Catheter Kitsee related information
Date Initiated by FirmAugust 06, 2021
Create DateOctober 06, 2021
Recall Status1 Terminated 3 on October 18, 2024
Recall NumberZ-0046-2022
Recall Event ID 88671
510(K)NumberK150046 
Product Classification System, hypothermia, intravenous, cooling - Product Code NCX
ProductZOLL, REF 8700-0787-40 (IC-3893AE), ICY Intravascular Heat Exchange Catheter Kit with central venous infusion capabilities (3 lumens), STERILE EO, Rx Only, UDI: (01)00849111075220
Code Information All Lots
FEI Number 3003793491
Recalling Firm/
Manufacturer
ZOLL Circulation, Inc.
2000 Ringwood Ave
San Jose CA 95131-1728
For Additional Information ContactSam Nanavati
408-419-2950
Manufacturer Reason
for Recall
Labeling update concerning potential leaks from the catheter or the start-up kit (SUK) tubing
FDA Determined
Cause 2
Labeling design
ActionBetween the dates of August 27, 2021 and September 3, 2021, ZOLL, issued an "Urgent Medical Device Recall (Correction)" Notification to all affected US consignees via UPS. In addition to informing consignees about the labeling correction, Zoll asked consignees to take the following actions: 1. If you have any questions, please contact Doug Lam (Director of Quality and Compliance) or Sam Nanavati (VP of Quality and Regulatory Affairs) at IVTM_Recall@zoll.com. 2 Provide this notice (letter) to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. 3 Maintain a copy of the recall notification and attachment with the IFU for reference. ZOLL will provide a method to obtain the updated IFU once it is released. 4 If you experience any adverse reactions or quality problems with the use of our products (catheters), you may report adverse reactions or quality problems to the FDA s MedWatch Adverse Event Reporting program either online (https://www.accessdata.fda.gov/scripts/medwatch/index.cfm?action=consumer.reporting1), by regular mail or by fax (1-800-332-0178). 5 Sign and return the pre-stamped post card (enclosed) acknowledging that you have read and understood the content and requirements on this recall notice (letter) within 5 working days from receipt of this notice. 6. If you do not believe that you are the right person to implement the above-mentioned actions, please forward this letter to the right person in your organization. Actions Underway by ZOLL: 1. ZOLL will submit a revised IFU to FDA with heavy focus on providing clear instructions to the users on how to investigate/mitigate a potential catheter leak whenever a depleted saline bag is noticed by the users. 2. ZOLL will provide all affected users a method to obtain the revised IFU once it is reviewed and cleared by the FDA. 3. ZOLL is retraining all Sales staff to the revised IFU. 4. If we could be of any further assi
Quantity in Commerce333 units
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the countries Australia, CANADA, CHINA, GERMANY, HUNGARY, PORTUGAL, SINGAPORE, TAIWAN, THAILAND.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = NCX
-
-