Date Initiated by FirmSeptember 11, 2026
Date PostedNovember 23, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0532-2027
Recall Event ID 99877
510(K)NumberK130572 
Product Classification Implant, endosseous, root-form - Product Code DZE
ProductSIMPLY INTERACTIVE SBM IMPLANT ASSEMBLY 3.2mmD X 3.0mmD PLATFORM REF: 653211U UDI-DI code: 10841307106821 Dental implant used to replace, support or enhance the tooth structure.
Code Information Part Number: 653211U UDI-DI code: 10841307106821 Lot Number: 15035029
FEI Number 3001617766
Recalling Firm/
Manufacturer
Implant Direct Sybron Manufacturing LLC
3050 E Hillcrest Dr
Westlake Village CA 91362-3171
For Additional Information ContactShani Aroch
818-444-3300
Manufacturer Reason
for Recall
Dental implant were manufacture out of specification and could potentially fracture due to excessive bending force or high insertion torque during placement.
FDA Determined
Cause 2
Under Investigation by firm
ActionOn 09/11/2026, an "URGENT: MEDICAL DEVICE RECALL" Letter was emailed to customers informing them that a specific lot (Lot Number 15035029) of Part/Item Number 653211U Simply Interactive SBM Implant Assembly 3.2mmD x 3.0mmD Platform were manufactured out of specification and may result with immediate fracture due to excessive bending forces, or high insertion torque during placement. Customers are instructed to: For affected device already implanted - It is recommended to continue routine clinical follow-up of patients implanted with the products from the affected lot. There is no need for medical intervention, as the worst-case fatigue testing results do not possess any additional risk to the product. Patients should be managed according to standard clinical practice if signs or symptoms suggestive of implant failure, mobility, prosthetic loosening, or fracture are observed. For products not implanted - check inventory immediately for any affected units from the affected Lot in your possession from the table above to determine if you have any of the affected products in your inventory. Please quarantine any units from the identified Lot in a manner that ensures the product will not be used. If you have the affected product listed above, contact the Implant Direct Complaints Team at 805-778-5924 to receive an RMA number. The RMA will allow for a quick return and replacement or credit. For questions, complaints, or adverse events via phone at +1-805-778-5924, Monday through Friday, 7:00 AM to 3:30 PM, PST, or via email at customer.claims@implantdirect.com
Quantity in Commerce57 units
DistributionU.S. Nationwide distribution in the states of AR, CA, CO, FL, ID, KS, KY, MI, MN, MO, MS, NC, OH, PA, TX, UT and VA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = DZE
-
-