| | Class 1 Device Recall FreeStyle Libre 3 App |  |
| Date Initiated by Firm | October 15, 2026 |
| Date Posted | November 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0536-2027 |
| Recall Event ID |
100151 |
| 510(K)Number | K212132 K213996 |
| Product Classification |
FreeStyle Libre 3 App - Product Code QLG
|
| Product | FreeStyle Libre 3 App
Software versions: Versions 3.2.0 through 3.6.8
Model/Catalog Number: 72077-01
UDI-DI code: 00357599817008
Model/Catalog Number: 72075-01
UDI-DI code: 00357599816001
The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated. |
| Code Information |
Model/Catalog Number: 72077-01
UDI-DI code: 00357599817008
Serial/Lot Numbers: N/A
Software Versions: Versions 3.2.0 through 3.6.8
Model/Catalog Number: 72075-01
UDI-DI code: 00357599816001
Serial/Lot Numbers: N/A
Software Versions: Versions 3.2.0 through 3.6.7
|
| FEI Number |
3002803444
|
Recalling Firm/ Manufacturer |
Abbott Diabetes Care, Inc.
1360 S Loop Rd
Alameda
CA
94502-7000
|
| For Additional Information Contact | Lindsy Delco 224-802-2475 |
Manufacturer Reason for Recall | Due to two issues:
1. Security vulnerability issue.
2. Extended Signal Loss - users may experience extended signal loss, meaning their sensor loses connection with the app and does not reconnect. |
FDA Determined Cause 2 | Software change control |
| Action | On 16 Oct 2026, Abbott initiated "URGENT MEDICAL DEVICE CORRECTION" Notification to customers of Security vulnerability communications and were paused 17 Oct 2026 due to the extended signal loss condition. Signal loss notifications commenced 20 Oct 2026. Updated app versions (FreeStyle Libre 3 v3.6.10 and Libre by Abbott v1.4.2) were released 22 Oct 2026. In-app notifications, phone warnings, and email communications began 23 Oct 2026, followed by notifications to non-updated users on 26 Oct 2026 and notifications to auto-updated users on 28 Oct 2026.
Reason for Notification Customers are being notified to update their apps to resolve two issues: (1) an extended signal loss condition identified 17 Oct 2026 affecting FreeStyle Libre 3 (v3.6.8) and Libre by Abbott (v1.4.1) app users, and (2) a security vulnerability affecting FreeStyle Libre 3, Libre by Abbott, and Lingo app users. The Lingo app is not affected by the signal loss issue.
Instructions to Customers:
Immediately update to the latest app version:
FreeStyle Libre 3: Update via Google Play or App Store. Confirm the Software Full Version starts with 3.6.9.
Libre by Abbott: Update via Google Play or App Store. Confirm the Software Full Version starts with 1.4.2.
Lingo: Update via Google Play or App Store. Confirm the Software Version starts with 1.12.3.
To update: open the Google Play Store or App Store tap your Profile icon tap Manage Apps (Android) or App Updates (iOS) tap Update next to the relevant app. If "Open" appears instead of "Update," the latest version is already installed.
If symptoms don't match your sensor reading, confirm with a fingerstick blood glucose test.
For questions, call Abbott Customer Service at 1-833-795-4273 (Mon-Sun, 8 a.m.-8 p.m. EDT) or chat 24/7 at freestyle.abbott/us-en/support/contact-us.html. |
| Quantity in Commerce | 1,836,506 (US 1,382,849; OUS 453,657) app downloads total |
| Distribution | Worldwide distribution - US Nationwide and the countries of Andorra, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Italy, Lichtenstein, Luxembourg, Netherlands, Norway, San Marino, Sweden, Qatar, Saudi Arabia, Australia, New Zealand, and Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = QLG
|
|
|
|