| Class 2 Device Recall Powerheart G3 | |
Date Initiated by Firm | October 13, 2008 |
Date Posted | February 25, 2009 |
Recall Status1 |
Terminated 3 on December 08, 2011 |
Recall Number | Z-0916-2009 |
Recall Event ID |
50064 |
510(K)Number | K031987 |
Product Classification |
Automated External Defibrillator - Product Code MKJ
|
Product | Cardiac Science Powerheart Automated External Defibrillator G3 9300A |
Code Information |
Serial numbers: 361248, 361266, 361299, 361313, 361317, 362365, 362744, 362823, 362874, 364236, 364916, 365018, 365029, 365036, 365207, 365291, 370763, 370926, 372286, 372838, 372885, 373523, 377725, 378408, 379285, 381083, 387110, 389251, 390171, 4000002, 4000004, 4000006, 4000007, 4000011, 4000013, 4000014, 4000016, 4000018, 4000021, 4000024, 4000031, 4000044, 4000045, 4000047, 4000048, 4000051, 4000054, 4000057, 4000063, 4000064, 4000065, 4000072, 4000075, 4000079, 4000082, 4000085, 4000086, 4000090, 4000097, 4000099, 4000138, 4000156, 4000164, 4000179, 4000187, 4000249, 4000250, 4000255, 4000256, 4000264, 4000271, 4000287, 4000288, 4000291, 4000295, 4000305, 4000306, 4000314, 4000322, 4000323, 4000324, 4000332, 4000333, 4000334, 4000335, 4000337, 4000342, 4000535, 4000706, 4000909, 4000912, 4001112, 4001561, 4003528, 4003745, 4003773, 4003779, 4009269, 4011367, 4013318, 4015632, 4018464, 4020662, 4020703, 4020809, 4020905, 4021392, 4021401, 4021417, 4021503, 4021507, 4021742, 4021743, 4021798, 4021812, 4021831, 4021873, 4021879, 4021892, 4021896, 4021899, 4021935, 4021946, 4021993, 4021994, 4022011, 4022016, 4022019, 4023689, 4023690, 4023694, 4023696, 4023719, 4023721, 4023723, 4023724, 4023726, 4023731, 4023739, 4023743, 4023744, 4023754, 4025817, 4025822, 4025823, 4025824, 4025825, 4025826, 4025827, 4025828, 4025835, 4025849, 4025850, 4025851, 4025853, 4025859, 4025865, 4025866, 4025867, 4025868, 4025869, 4025870, 4025871, 4025872, 4025873, 4025874, 4025875, 4025878, 4025880, 4025881, 4025884, 4025885, 4025886, 4025888, 4025890, 4025891, 4025892, 4025893, 4025894, 4025896, 4025897, 4025899, 4025900, 4025901, 4025902, 4025903, 4025904, 4025905, 4025906, 4025907, 4025908, 4025909, 4025911, 4025912, 4025913, 4025916, 4025917, 4025918, 4025919, 4025920, 4025921, 4025922, 4025923, 4025924, 4025925, 4025926, 4025928, 4025929, 4025930, 4025931, 4025932, 4025933, 4025938, 4025940, 4025946, 4025947, 4025948, 4025949, 4025950, 4025951, 4025955, 4025956, 4025958, 4025962, 4025963, 4025965, 4025966, 4025967, 4025971, 4025977, 4025979, 4025982, 4025983, 4025986, 4025988, 4025990, 4025991, 4025992, 4025993, 4025994, 4025995, 4025996, 4026000, 4026002, 4026005, 4026006, 4026007, 4026008, 4026009, 4026010, 4026011, 4026012, 4026013, 4026014, 4026015, 4026016, 4026017, 4026018, 4026019, 4026020, 4026021, 4026022, 4026023, 4026024, 4026025, 4026026, 4026027, 4026028, 4026029, 4026030, 4026031, 4026032, 4026033, 4026034, 4026035, 4026036, 4026037, 4026038, 4026039, 4026040, 4026041, 4026042, 4026043, 4026044, 4026045, 4026046, 4026047, 4026048, 4026049, 4026050, 4026051, 4026052, 4026053, 4026054, 4026055, 4026056, 4026057, 4026058, 4026059, 4026060, 4026061, 4026062, 4026063, 4026064, 4026065, 4026066, 4026067, 4026068, 4026069, 4026070, 4026071, 4026072, 4026073, 4026074, 4026075, 4026076, 4026077, 4026078, 4026079, 4026080, 4026081, 4026082, 4026083, 4026084, 4026085, 4026086, 4026087, 4026088, 4026089, 4026090, 4026091, 4026092, 4026093, 4026094, 4026095, 4026096, 4026097, 4026098, 4026099, 4026100, 4026101, 4026102, 4026103, 4026104, 4026105, 4026106, 4026107, 4026108, 4026109, 4026110, 4026111, 4026112, 4026113, 4026114, 4026115, 4026116, 4026117, 4026118, 4026119, 4026120, 4026121, 4026122, 4026123, 4026124, 4026125, 4026126, 4026127, 4026128, 4026129, 4026130, 4026131, 4026132, 4026133, 4026134, 4026135, 4026136, 4026137, 4026138, 4026139, 4026140, 4026141, 4026142, 4026143, 4026144, 4026145, 4026146, 4026147, 4026148, 4026149, 4026150, 4026151, 4026152, 4026153, 4026154, 4026155, 4026156, 4026157, 4026158, 4026159, 4026160, 4026161, 4026162, 4026163, 4026164, 4026166, 4026167, 4026168, 4026169, 4026170, 4026171, 4026172, 4026173, 4026174, 4026175, 4026176, 4026177, 4026178, 4026179, 4026180, 4026181, 4026182, 4026183, 4026184, 4026185, 4026186, 4026187, 4026188, 4026189, 4026190, 4026191, 4026192, 4026193, 4026194, 4026195, 4026196, 4026197, 4026198, 4026199, 4026200, 4026201, 4026202, 4026203, 4026204, 4026205, 4026206, 4026207, 4026208, 4026209, 4026210, 4026211, 4026212, 4026213, 4026214, 4026215, 4026216, 4026217, 4026218, 4026219, 4026220, 4026221, 4026222, 4026223, 4026224, 4026225, 4026226, 4026227, 4026228, 4026229, 4026230, 4026231, 4026232, 4026233, 4026234, 4026235, 4026236, 4026237, 4026238, 4026239, 4026240, 4026241, 4026242, 4026243, 4026244, 4026245, 4026246, 4026247, 4026248, 4026249, 4026250, 4026251, 4026252, 4026253, 4026254, 4026255, 4026256, 4026257, 4026258, 4026259, 4026260, 4026261, 4026262, 4026263, 4026264, 4026265, 4026266, 4026267, 4026268, 4026269, 4026270, 4026271, 4026272, 4026273, 4026274, 4026275, 4026276, 4026277, 4026278, 4026279, 4026280, 4026281, 4026282, 4026283, 4026284, 4026285, 4026286, 4026287, 4026288, 4026289, 4026290, 4026291, 4026292, 4026293, 4026294, 4026295, 4026296, 4026297, 4026298, 4026299, 4026300, 4026301, 4026302, 4026303, 4026304, 4026305, 4026306, 4026307, 4026308, 4027909, 4027910, 4027911, 4027912, 4027913, 4027914, 4027915, 4027918, 4027919, 4027920, 4027921, 4027922, 4027923, 4027924, 4027925, 4027926, 4027927, 4027928, 4027929, 4027930, 4027931, 4027932, 4027933, 4028234, 4028235, 4028236, 4028237, 4028238, 4028239, 4028240, 4028241, 4028242, 4028243, 4028244, 4028245, 4028246, 4028247, 4028248, 4028249, 4028250, 4028251, 4028252, 4028253, 4028254, 4028255, 4028256, 4028257, 4028258, 4028259, 4028260, 4028261, 4028262, 4028263, 4028264, 4028265, 4028266, 4028268, 4028269, 4028270, 4028271, 4028272, 4028273, 4028274, 4028275, 4028276, 4028277, 4028278, 4028279, 4028280, 4028281, 4028282, 4028283, 4028284, 4028285, 4028286, 4028287, 4028288, 4028289, 4028290, 4028291, 4028292, 4028293, 4028294, 4028295, 4028296, 4028297, 4028298, 4028299, 4028300, 4028301, 4028302, 4028303, 4028304, 4028305, 4028306, 4028307, 4028308, 4028309, 4028310, 4028311, 4028312, 4028313, 4028314, 4028315, 4028316, 4028317, 4028318, 4028319, 4028320, 4028321, 4028322, 4028323, 4028324, 4028325, 4028326, 4028327, 4028328, 4028329, 4028330, 4028331, 4028332, 4028333, 4028334, 4028335, 4028336, 4028337, 4028338, 4028339, 4028340, 4028341, 4028342, 4028343, 4028344, 4028345, 4028346, 4028347, 4028348, 4028349, 4028350, 4028351, 4028352, 4028353, 4028354, 4028355, 4028356, 4028357, 4028358, 4028359, 4028360, 4028361, 4028362, 4028363, 4028364, 4028365, 4028366, 4028367, 4028368, 4028369, 4028370, 4028372, 4028373, 4028374, 4028375, 4028376, 4028377, 4028378, 4028379, 4028380, 4028381, 4028382, 4028383, 4028384, 4028385, 4028386, 4028387, 4028388, 4028389, 4028390, 4028391, 4028392, 4028393, 4028394, 4028395, 4028396, 4028397, 4028398, 4028399, 4028400, 4028401, 4028402, 4028403, 4028404, 4028405, 4028406, 4028407, 4028408, 4028409, 4028410, 4028411, 4028412, 4028413, 4028414, 4028415, 4028416, 4028417, 4028418, 4028419, 4028420, 4028421, 4028422, 4028423, 4028424, 4028425, 4028426, 4028427, 4028428, 4028429, 4028430, 4028431, 4028432, 4028433, 4028434, 4028435, 4028436, 4028437, 4028438, 4028439, 4028440, 4028441, 4028442, 4028443, 4028444, 4028445, 4028446, 4028447, 4028448, 4028449, 4028450, 4028451, 4028452, 4028453, 4028454, 4028455, 4028456, 4028457, 4028458, 4028459, 4028460, 4028461, 4028462, 4028463, 4028464, 4028465, 4028466, 4028467, 4028468, 4028469, 4028470, 4028471, 4028472, 4028473, 4028474, 4028475, 4028476, 4028477, 4028478, 4028479, 4028480, 4028481, 4028482, 4028483, 4028484, 4028485, 4028486, 4028487, 4028488, 4028489, 4028490, 4028491, 4028492, 4028493, 4028494, 4028495, 4028496, 4028497, 4028498, 4028499, 4028500, 4028501, 4028502, 4028503, 4028504, 4028505, 4028506, 4028507, 4028508, 4028509, 4028510, 4028511, 4028512, 4028513, 4028514, 4028515, 4028516, 4028517, 4028518, 4028519, 4028520, 4028521, 4028522, 4028523, 4028524, 4028525, 4028526, 4028527, 4028528, 4028529, 4028530, 4028531, 4028532, 4028533, 4028534, 4028535, 4028536, 4028537, 4028538, 4028539, 4028540, 4028541, 4028542, 4028543, 4028544, 4028545, 4028546, 4028547, 4028548, 4028549, 4028550, 4028551, 4028552, 4028553, 4028554, 4028555, 4028556, 4028557, 4028558, 4028559, 4028560, 4028561, 4028562, 4028563, 4028564, 4028565, 4028566, 4028567, 4028568, 4028569, 4028570, 4028571, 4028572, 4028573, 4028574, 4028575, 4028576, 4028577, 4028578, 4028579, 4028580, 4028581, 4028582, 4028583, 4028584, 4028585, 4028586, 4028587, 4028588, 4028589, 4028590, 4028591, 4028592, 4028593, 4028594, 4028595, 4028596, 4028597, 4028598, 4028599, 4028600, 4028601, 4028602, 4028603, 4028604, 4028605, 4028606, 4028607, 4028608, 4028609, 4028610, 4028611, 4028612, 4028613, 4028614, 4028615, 4028616, 4028617, 4028618, 4028619, 4028620, 4028621, 4028622, 4028623, 4028624, 4028625, 4028626, 4028627, 4028628, 4028629, 4028630, 4028631, 4028632, 4028633, 4028634, 4028635, 4028636, 4028637, 4028638, 4028639, 4028640, 4028641, 4028642, 4028643, 4028644, 4028645, 4028646, 4028647, 4028648, 4028649, 4028650, 4028651, 4028652, 4028653, 4028654, 4028655, 4028656, 4028657, 4028658, 4028659, 4028660, 4028661, 4028662, 4028663, 4028664, 4028665, 4028666, 4028667, 4028668, 4028669, 4028670, 4028671, 4028672, 4028673, 4028674, 4028675, 4028676, 4028677, 4028678, 4028679, 4028680, 4028681, 4028682, 4028683, 4028684, 4028685, 4028686, 4028687, 4028688, 4028689, 4028690, 4028691, 4028692, 4028693, 4028694, 4028695, 4028696, 4028697, 4028698, 4028699, 4028700, 4028701, 4028702, 4028703, 4028704, 4028705, 4028706, 4028707, 4028708, 4028709, 4028710, 4028711, 4028712, 4028713, 4028714, 4028715, 4028716, 4028717, 4028718, 4028719, 4028720, 4028721, 4028722, 4028723, 4028724, 4028725, 4028726, 4028727, 4028728, 4028729, 4028730, 4028731, 4028732, 4028733, 4030909, 4030910, 4030911, 4030912, 4030913, 4030914, 4030915, 4030916, 4030917, 4030918, 4030919, 4030920, 4030921, 4030922, 4030923, 4030924, 4030925, 4030926, 4030927, 4030928, 4030929, 4030930, 4030931, 4030932, 4030933, 4030934, 4030935, 4030936, 4030937, 4030938, 4030939, 4030940, 4030941, 4030942, 4030943, 4030944, 4030945, 4030946, 4030947, 4030948, 4030949, 4030950, 4030951, 4030952, 4030953, 4030954, 4030955, 4030956, 4030957, 4030958, 4030959, 4030960, 4030961, 4030962, 4030963, 4030964, 4030965, 4030966, 4030967, 4030968, 4030969, 4030970, 4030971, 4030972, 4030973, 4030974, 4030975, 4030976, 4030977, 4030978, 4030979, 4030980, 4030981, 4030982, 4030983, 4030984, 4030985, 4030986, 4030987, 4030988, 4030989, 4030990, 4030991, 4030992, 4030993, 4030994, 4030995, 4030996, 4030997, 4030998, 4030999, 4031000, 4031001, 4031002, 4031003, 4031004, 4031005, 4031006, 4031007, 4031008, 4031009, 4031010, 4031011, 4031012, 4031013, 4031014, 4031015, 4031016, 4031017, 4031018, 4031019, 4031020, 4031021, 4031022, 4031023, 4031024, 4031025, 4031026, 4031027, 4031028, 4031029, 4031030, 4031031, 4031032, 4031033, 4031034, 4031035, 4031036, 4031037, 4031038, 4031039, 4031040, 4031041, 4031042, 4031043, 4031044, 4031045, 4031046, 4031047, 4031048, 4031049, 4031050, 4031051, 4031052, 4031053, 4031054, 4031055, 4031056, 4031057, 4031058, 4031059, 4031060, 4031061, 4031062, 4031063, 4031064, 4031065, 4031066, 4031067, 4031068, 4031069, 4031070, 4031071, 4031072, 4031073, 4031074, 4031075, 4031076, 4031077, 4031078, 4031079, 4031080, 4031081, 4031082, 4031083, 4031084, 4031085, 4031086, 4031087, 4031088, 4031089, 4031090, 4031091, 4031092, 4031093, 4031094, 4031095, 4031096, 4031097, 4031098, 4031099, 4031100, 4031101, 4031102, 4031103, 4031104, 4031105, 4031106, 4031107, 4031108, 4031109, 4031110, 4031111, 4031112, 4031113, 4031114, 4031115, 4031116, 4031117, 4031118, 4031119, 4031120, 4031121, 4031122, 4031123, 4031124, 4031125, 4031126, 4031127, 4031128, 4031129, 4031130, 4031131, 4031132, 4031133, 4031134, 4031135, 4031136, 4031137, 4031138, 4031139, 4031140, 4031141, 4031142, 4031143, 4031144, 4031145, 4031146, 4031147, 4031148, 4031149, 4031150, 4031151, 4031152, 4031153, 4031154, 4031155, 4031156, 4031157, 4031158, 4031159, 4031160, 4031161, 4031162, 4031163, 4031164, 4031165, 4031166, 4031167, 4031168, 4031169, 4031170, 4031171, 4031172, 4031173, 4031174, 4031175, 4031176, 4031177, 4031178, 4031179, 4031180, 4031181, 4031182, 4031183, 4031184, 4031185, 4031186, 4031187, 4031188, 4031189, 4031190, 4031191, 4031192, 4031193, 4031194, 4031195, 4031196, 4031197, 4031198, 4031199, 4031200, 4031201, 4031202, 4031203, 4031204, 4031205, 4031206, 4031207, 4031208, 4031209, 4031210, 4031211, 4031212, 4031213, 4031214, 4031215, 4031216, 4031217, 4031218, 4031219, 4031220, 4031221, 4031222, 4031223, 4031224, 4031225, 4031226, 4031227, 4031228, 4031229, 4031230, 4031231, 4031232, 4031233, 4031234, 4031235, 4031236, 4031237, 4031238, 4031239, 4031240, 4031241, 4031242, 4031243, 4031244, 4031245, 4031246, 4031247, 4031248, 4031249, 4031250, 4031251, 4031252, 4031253, 4031254, 4031255, 4031256, 4031257, 4031258, 4031259, 4031260, 4031261, 4031262, 4031263, 4031264, 4031265, 4031266, 4031267, 4031268, 4031269, 4031270, 4031271, 4031272, 4031273, 4031274, 4031275, 4031276, 4031277, 4031278, 4031279, 4031280, 4031281, 4031282, 4031283, 4031284, 4031285, 4031286, 4031287, 4031288, 4031289, 4031290, 4031291, 4031292, 4031293, 4031294, 4031295, 4031296, 4031297, 4031298, 4031299, 4031300, 4031301, 4031302, 4031303, 4031304, 4031305, 4031306, 4031307, 4031308, 4031309, 4031310, 4031311, 4031312, 4031313, 4031314, 4031315, 4031316, 4031317, 4031318, 4031319, 4031320, 4031321, 4031322, 4031323, 4031324, 4031325, 4031326, 4031327, 4031328, 4031329, 4031330, 4031331, 4031332, 4031333, 4031334, 4031335, 4031336, 4031337, 4031338, 4031339, 4031340, 4031341, 4031342, 4031343, 4031344, 4031345, 4031346, 4031347, 4031348, 4031349, 4031350, 4031351, 4031352, 4031353, 4031354, 4031355, 4031356, 4031357, 4031358, 4031359, 4031360, 4031361, 4031362, 4031363, 4031364, 4031365, 4031366, 4031367, 4031368, 4031369, 4031370, 4031371, 4031372, 4031373, 4031374, 4031375, 4031376, 4031377, 4031378, 4031379, 4031380, 4031381, 4031382, 4031383, 4031384, 4031385, 4031386, 4031387, 4031388, 4031389, 4031390, 4031391, 4031392, 4031393, 4031394, 4031395, 4031396, 4031397, 4031398, 4031399, 4031400, 4031401, 4031402, 4031403, 4031404, 4031405, 4031406, 4031407, 4031408, 4032517, 4032518, 4032519, 4032520, 4032521, 4032522, 4032523, 4032524, 4032525, 4032526, 4032527, 4032528, 4032529, 4032530, 4032531, 4032532, 4032533, 4032534, 4032535, 4032536, 4032537, 4032538, 4032539, 4032540, 4032541, 4032542, 4032543, 4032544, 4032545, 4032546, 4032547, 4032548, 4032549, 4032550, 4032551, 4032552, 4032553, 4032554, 4032555, 4032556, 4032557, 4032558, 4032559, 4032560, 4032561, 4032562, 4032563, 4032564, 4032565, 4032566, 4032567, 4032568, 4032569, 4032570, 4032571, 4032572, 4032573, 4032574, 4032575, 4032576, 4032577, 4032578, 4032579, 4032580, 4032581, 4032582, 4032583, 4032584, 4032585, 4032586, 4032587, 4032588, 4032589, 4032590, 4032591, 4032592, 4032593, 4032594, 4032595, 4032596, 4032597, 4032598, 4032599, 4032600, 4032601, 4032602, 4032603, 4032604, 4032605, 4032606, 4032607, 4032608, 4032609, 4032610, 4032611, 4032612, 4032613, 4032614, 4032615, 4032616, 4032617, 4032618, 4032619, 4032620, 4032621, 4032622, 4032623, 4032624, 4032625, 4032626, 4032627, 4032628, 4032629, 4032630, 4032631, 4032632, 4032633, 4032634, 4032635, 4032636, 4032637, 4032638, 4032639, 4032640, 4032641, 4032642, 4032643, 4032644, 4032645, 4032646, 4032647, 4032648, 4032649, 4032650, 4032651, 4032652, 4032653, 4032654, 4032655, 4032656, 4032657, 4032658, 4032659, 4032660, 4032661, 4032662, 4032663, 4032664, 4032665, 4032666, 4032667, 4032668, 4032669, 4032670, 4032671, 4032672, 4032673, 4032674, 4032675, 4032677, 4032678, 4032679, 4032680, 4032681, 4032682, 4032683, 4032684, 4032685, 4032686, 4032687, 4032688, 4032689, 4032690, 4032691, 4032692, 4032693, 4032694, 4032695, 4032696, 4032697, 4032698, 4032699, 4032700, 4032701, 4032702, 4032703, 4032704, 4032705, 4032706, 4032707, 4032708, 4032709, 4032710, 4032711, 4032712, 4032713, 4032714, 4032715, 4032716, 4032717, 4032718, 4032719, 4032720, 4032721, 4032722, 4032723, 4032724, 4032725, 4032726, 4032727, 4032728, 4032729, 4032730, 4032731, 4032732, 4032733, 4032734, 4032735, 4032736, 4032737, 4032738, 4032739, 4032740, 4032741, 4032742, 4032743, 4032744, 4032745, 4032746, 4032747, 4032748, 4032749, 4032750, 4032751, 4032752, 4032753, 4032754, 4032755, 4032756, 4032757, 4032758, 4032759, 4032760, 4032761, 4032762, 4032763, 4032764, 4032765, 4032766, 4032767, 4032768, 4032769, 4032770, 4032771, 4032772, 4032773, 4032774, 4032775, 4032776, 4032777, 4032778, 4032779, 4032780, 4032781, 4032782, 4032783, 4032784, 4032785, 4032786, 4032787, 4032788, 4032789, 4032790, 4032791, 4032792, 4032793, 4032794, 4032795, 4032796, 4032797, 4032798, 4032799, 4032800, 4032801, 4032802, 4032803, 4032804, 4032805, 4032806, 4032807, 4032808, 4032809, 4032810, 4032811, 4032812, 4032813, 4032814, 4032815, 4032816, 4032817, 4032818, 4032819, 4032820, 4032821, 4032822, 4032823, 4032824, 4032825, 4032826, 4032827, 4032828, 4032829, 4032830, 4032831, 4032832, 4032833, 4032834, 4032835, 4032836, 4032837, 4032838, 4032839, 4032840, 4032841, 4032842, 4032843, 4032844, 4032845, 4032846, 4032847, 4032848, 4032849, 4032850, 4032851, 4032852, 4032853, 4032854, 4032855, 4032856, 4032857, 4032858, 4032859, 4032860, 4032861, 4032862, 4032863, 4032864, 4032865, 4032866, 4032867, 4032868, 4032869, 4032870, 4032871, 4032872, 4032873, 4032874, 4032875, 4032876, 4032877, 4032878, 4032879, 4032880, 4032881, 4032882, 4032883, 4032884, 4032885, 4032886, 4032887, 4032888, 4032889, 4032890, 4032891, 4032892, 4032893, 4032894, 4032895, 4032896, 4032897, 4032898, 4032899, 4032900, 4032901, 4032902, 4032903, 4032904, 4032905, 4032906, 4032907, 4032908, 4032909, 4032910, 4032911, 4032912, 4032913, 4032914, 4032915, 4032916, 4032917, 4032918, 4032919, 4032920, 4032921, 4032922, 4032923, 4032924, 4032925, 4032926, 4032927, 4032928, 4032929, 4032930, 4032931, 4032932, 4032933, 4032934, 4032935, 4032936, 4032937, 4032938, 4032939, 4032940, 4032941, 4032942, 4032943, 4032944, 4032945, 4032946, 4032947, 4032948, 4032949, 4032950, 4032951, 4032952, 4032953, 4032954, 4032955, 4032956, 4032957, 4032958, 4032959, 4032960, 4032961, 4032962, 4032963, 4032964, 4032965, 4032966, 4032967, 4032968, 4032969, 4032970, 4032971, 4032972, 4032973, 4032974, 4032975, 4032976, 4032977, 4032978, 4032979, 4032980, 4032981, 4032982, 4032983, 4032984, 4032985, 4032986, 4032987, 4032988, 4032989, 4032990, 4032991, 4032992, 4032993, 4032994, 4032995, 4032996, 4032997, 4032998, 4032999, 4033000, 4033001, 4033002, 4033003, 4033004, 4033005, 4033006, 4033007, 4033008, 4033009, 4033010, 4033011, 4033012, 4033013, 4033014, 4033015, 4033016, 4034788, 4034789, 4034790, 4034791, 4034792, 4034793, 4034794, 4034795, 4034796, 4034797, 4034798, 4034799, 4034800, 4034801, 4034802, 4034803, 4034804, 4034805, 4034806, 4034807, 4034808, 4034809, 4034810, 4034811, 4034812, 4034813, 4034814, 4034815, 4034816, 4034817, 4034818, 4034819, 4034820, 4034821, 4034822, 4034823, 4034824, 4034825, 4034826, 4034827, 4034828, 4034829, 4034830, 4034831, 4034832, 4034833, 4034834, 4034835, 4034836, 4034837, 4034838, 4034839, 4034840, 4034841, 4034842, 4034843, 4034844, 4034845, 4034846, 4034847, 4034848, 4034849, 4034850, 4034851, 4034852, 4034853, 4034854, 4034855, 4034856, 4034857, 4034858, 4034859, 4034860, 4034861, 4034862, 4034863, 4034864, 4034865, 4034866, 4034867, 4034868, 4034869, 4034870, 4034871, 4034872, 4034873, 4034874, 4034875, 4034876, 4034877, 4034878, 4034879, 4034880, 4034881, 4034882, 4034883, 4034884, 4034885, 4034886, 4034887, 4034888, 4034889, 4034890, 4034891, 4034892, 4034893, 4034894, 4034895, 4034896, 4034897, 4034898, 4034899, 4034900, 4034901, 4034902, 4034903, 4034904, 4034905, 4034906, 4034907, 4034908, 4034909, 4034910, 4034911, 4034912, 4034913, 4034914, 4034915, 4034916, 4034917, 4034918, 4034919, 4034920, 4034921, 4034922, 4034923, 4034924, 4034925, 4034926, 4034927, 4034928, 4034929, 4034930, 4034931, 4034932, 4034933, 4034934, 4034936, 4034937, 4034938, 4034939, 4034940, 4034941, 4034942, 4034943, 4034944, 4034945, 4034946, 4034947, 4034948, 4034949, 4034950, 4034951, 4034952, 4034953, 4034954, 4034955, 4034956, 4034957, 4034958, 4034959, 4034960, 4034961, 4034962, 4034963, 4034964, 4034965, 4034966, 4034967, 4034968, 4034969, 4034970, 4034971, 4034972, 4034973, 4034974, 4034975, 4034976, 4034977, 4034978, 4034979, 4034980, 4034981, 4034982, 4034983, 4034984, 4034985, 4034986, 4034987, 4034988, 4034989, 4034990, 4034991, 4034992, 4034993, 4034994, 4034995, 4034996, 4034997, 4034998, 4034999, 4035000, 4035001, 4035002, 4035003, 4035004, 4035005, 4035006, 4035007, 4035008, 4035009, 4035010, 4035011, 4035012, 4035013, 4035014, 4035015, 4035016, 4035017, 4035018, 4035019, 4035020, 4035021, 4035022, 4035023, 4035024, 4035025, 4035026, 4035027, 4035028, 4035029, 4035030, 4035031, 4035032, 4035033, 4035034, 4035035, 4035036, 4035037, 4035038, 4035039, 4035040, 4035041, 4035042, 4035043, 4035044, 4035045, 4035046, 4035047, 4035048, 4035049, 4035050, 4035051, 4035052, 4035053, 4035054, 4035055, 4035056, 4035057, 4035058, 4035059, 4035060, 4035061, 4035062, 4035063, 4035064, 4035065, 4035066, 4035067, 4035068, 4035069, 4035070, 4035071, 4035072, 4035073, 4035074, 4035075, 4035076, 4035077, 4035078, 4035079, 4035080, 4035081, 4035082, 4035083, 4035084, 4035085, 4035086, 4035087, 4035088, 4035089, 4035090, 4035091, 4035092, 4035093, 4035094, 4035095, 4035096, 4035097, 4035098, 4035099, 4035100, 4035101, 4035102, 4035103, 4035104, 4035105, 4035106, 4035107, 4035108, 4035109, 4035110, 4035111, 4035112, 4035113, 4035114, 4035115, 4035116, 4035117, 4035118, 4035119, 4035122, 4035123, 4035124, 4035125, 4035126, 4035127, 4035128, 4035129, 4035130, 4035131, 4035132, 4035133, 4035134, 4035135, 4035136, 4035137, 4035138, 4035139, 4035140, 4035141, 4035142, 4035143, 4035144, 4035145, 4035146, 4035147, 4035148, 4035149, 4035150, 4035151, 4035152, 4035153, 4035154, 4035155, 4035157, 4035158, 4035159, 4035160, 4035161, 4035162, 4035163, 4035164, 4035165, 4035166, 4035167, 4035168, 4035169, 4035170, 4035171, 4035172, 4035173, 4035174, 4035175, 4035176, 4035177, 4035178, 4035179, 4035180, 4035181, 4035182, 4035183, 4035184, 4035185, 4035186, 4035187, 4035188, 4035189, 4035190, 4035191, 4035192, 4035193, 4035194, 4035195, 4035196, 4035197, 4035198, 4035199, 4035200, 4035201, 4035202, 4035203, 4035204, 4035205, 4035206, 4035207, 4035208, 4035209, 4035210, 4035211, 4035212, 4035213, 4035214, 4035215, 4035216, 4035217, 4035218, 4035219, 4035220, 4035221, 4035222, 4035223, 4035224, 4035225, 4035226, 4035227, 4035228, 4035229, 4035230, 4035231, 4035232, 4035233, 4035234, 4035235, 4035236, 4035237, 4035238, 4035239, 4035240, 4035241, 4035242, 4035243, 4035244, 4035245, 4035246, 4035247, 4035248, 4035249, 4035250, 4035251, 4035252, 4035253, 4035254, 4035255, 4035256, 4035257, 4035258, 4035259, 4035260, 4035261, 4035262, 4035263, 4035264, 4035265, 4035266, 4035267, 4035268, 4035269, 4035270, 4035271, 4035272, 4035273, 4035274, 4035275, 4035276, 4035277, 4035278, 4035279, 4035280, 4035281, 4035282, 4035283, 4035284, 4035285, 4035286, 4035287, 4035288, 4035289, 4035290, 4035291, 4035292, 4035293, 4035294, 4035295, 4035296, 4035297, 4035298, 4035299, 4035300, 4035301, 4035302, 4035303, 4035304, 4035305, 4035306, 4035307, 4035308, 4035309, 4035310, 4035311, 4035312, 4035313, 4035314, 4035315, 4035316, 4035317, 4035318, 4035319, 4035320, 4035321, 4035322, 4035323, 4035324, 4035325, 4035326, 4035327, 4035328, 4035329, 4035330, 4035331, 4035333, 4035334, 4035335, 4035336, 4035337, 4035484, 4035485, 4035486, 4035487, 4035488, 4035489, 4035490, 4035491, 4035492, 4035493, 4035494, 4035495, 4035496, 4035497, 4035498, 4035499, 4035500, 4035501, 4035502, 4035503, 4035504, 4035505, 4035506, 4035507, 4035508, 4035509, 4035510, 4035511, 4035512, 4035513, 4035514, 4035515, 4035516, 4035517, 4035518, 4035519, 4035520, 4035521, 4035522, 4035523, 4035524, 4035525, 4035526, 4035527, 4035528, 4035529, 4035530, 4035531, 4035532, 4035533, 4035534, 4035535, 4035536, 4035537, 4035538, 4035539, 4035540, 4035541, 4035542, 4035543, 4035544, 4035545, 4035546, 4035547, 4035548, 4035549, 4035550, 4035551, 4035552, 4035553, 4035554, 4035555, 4035556, 4035557, 4035558, 4035559, 4035560, 4035561, 4035562, 4035563, 4035564, 4035565, 4035566, 4035567, 4035568, 4035569, 4035570, 4035571, 4035572, 4035573, 4035574, 4035575, 4035576, 4035577, 4035578, 4035579, 4035580, 4035581, 4035582, 4035583, 4035584, 4035585, 4035586, 4035587, 4035588, 4035589, 4035590, 4035591, 4035592, 4035593, 4035594, 4035595, 4035596, 4035597, 4035598, 4035599, 4035600, 4035601, 4035602, 4035603, 4035604, 4035605, 4035606, 4035607, 4035608, 4035609, 4035610, 4035611, 4035612, 4035613, 4035614, 4035615, 4035616, 4035617, 4035618, 4035619, 4035620, 4035621, 4035622, 4035623, 4035624, 4035625, 4035626, 4035627, 4035628, 4035629, 4035630, 4035631, 4035632, 4035633, 4035634, 4035635, 4035636, 4035637, 4035638, 4035639, 4035640, 4035641, 4035642, 4035643, 4035644, 4035645, 4035646, 4035647, 4035648, 4035649, 4035650, 4035651, 4035652, 4035653, 4035654, 4035655, 4035656, 4035657, 4035658, 4035659, 4035660, 4035661, 4035662, 4035663, 4035664, 4035665, 4035666, 4035667, 4035668, 4035669, 4035670, 4035671, 4035672, 4035673, 4035674, 4035675, 4035676, 4035677, 4035678, 4035679, 4035680, 4035681, 4035682, 4035683, 4035684, 4035685, 4035686, 4035687, 4035688, 4035689, 4035690, 4035691, 4035692, 4035693, 4035694, 4035695, 4035696, 4035697, 4035698, 4035699, 4035700, 4035701, 4035702, 4035703, 4035704, 4035705, 4035706, 4035707, 4035708, 4035709, 4035710, 4035711, 4035712, 4035713, 4035714, 4035715, 4035716, 4035717, 4035718, 4035719, 4035720, 4035721, 4035722, 4035723, 4035724, 4035725, 4035726, 4035727, 4035729, 4035730, 4035731, 4035732, 4035733, 4035734, 4035735, 4035736, 4035737, 4035738, 4035739, 4035740, 4035741, 4035742, 4035743, 4035744, 4035745, 4035746, 4035747, 4035748, 4035749, 4035750, 4035751, 4035752, 4035753, 4035754, 4035755, 4035756, 4035757, 4035758, 4035759, 4035760, 4035761, 4035762, 4035763, 4035764, 4035765, 4035766, 4035767, 4035768, 4035769, 4035770, 4035771, 4035772, 4035773, 4035774, 4035775, 4035776, 4035777, 4035778, 4035779, 4035780, 4035781, 4035782, 4035783, 4035784, 4035785, 4035786, 4035787, 4035788, 4035789, 4035790, 4035791, 4035792, 4035793, 4035794, 4035795, 4035796, 4035797, 4035798, 4035799, 4035800, 4035801, 4035802, 4035803, 4035804, 4035805, 4035806, 4035807, 4035808, 4035809, 4035810, 4035811, 4035812, 4035813, 4035814, 4035815, 4035816, 4035817, 4035818, 4035819, 4035820, 4035821, 4035822, 4035823, 4035824, 4035825, 4035826, 4035827, 4035828, 4035829, 4035830, 4035831, 4035832, 4035833, 4035834, 4035835, 4035836, 4035837, 4035838, 4035839, 4035840, 4035841, 4035842, 4035843, 4035844, 4035845, 4035846, 4035847, 4035848, 4035849, 4035850, 4035851, 4035852, 4035853, 4035854, 4035855, 4035856, 4035857, 4035858, 4035859, 4035860, 4035861, 4035862, 4035863, 4035864, 4035865, 4035866, 4035867, 4035868, 4035869, 4035870, 4035871, 4035872, 4035873, 4035874, 4035875, 4035876, 4035877, 4035878, 4035879, 4035880, 4035881, 4035882, 4035883, 4035884, 4035885, 4035886, 4035887, 4035888, 4035889, 4035890, 4035891, 4035892, 4035893, 4035894, 4035895, 4035896, 4035897, 4035898, 4035899, 4035900, 4035901, 4035902, 4035903, 4035904, 4035905, 4035906, 4035907, 4035908, 4035909, 4035910, 4035911, 4035912, 4035913, 4035914, 4035915, 4035916, 4035917, 4035918, 4035919, 4035920, 4035921, 4035922, 4035923, 4035924, 4035925, 4035926, 4035927, 4035928, 4035929, 4035930, 4035931, 4035932, 4035933, 4035934, 4035935, 4035936, 4035937, 4035938, 4035939, 4035940, 4035941, 4035942, 4035943, 4035944, 4035945, 4035946, 4035947, 4035948, 4035949, 4035950, 4035951, 4035952, 4035953, 4035954, 4035955, 4035956, 4035957, 4035958, 4035959, 4035960, 4035961, 4035962, 4035963, 4035964, 4035965, 4035966, 4035967, 4035968, 4035969, 4035970, 4035971, 4035972, 4035973, 4035974, 4035975, 4035976, 4035977, 4035978, 4035979, 4035980, 4035981, 4035982, 4035983, 4037448, 4037449, 4037450, 4037451, 4037452, 4037453, 4037454, 4037455, 4037456, 4037457, 4037458, 4037459, 4037460, 4037461, 4037462, 4037463, 4037464, 4037465, 4037466, 4037467, 4037468, 4037469, 4037470, 4037471, 4037472, 4037473, 4037474, 4037475, 4037476, 4037477, 4037478, 4037479, 4037480, 4037481, 4037482, 4037483, 4037484, 4037485, 4037486, 4037487, 4037488, 4037489, 4037490, 4037491, 4037492, 4037493, 4037494, 4037495, 4037496, 4037497, 4037498, 4037499, 4037500, 4037501, 4037502, 4037503, 4037504, 4037505, 4037506, 4037507, 4037508, 4037509, 4037510, 4037511, 4037512, 4037513, 4037514, 4037515, 4037516, 4037517, 4037518, 4037519, 4037520, 4037521, 4037522, 4037523, 4037524, 4037525, 4037526, 4037527, 4037528, 4037529, 4037530, 4037531, 4037532, 4037533, 4037534, 4037535, 4037536, 4037537, 4037538, 4037539, 4037540, 4037541, 4037542, 4037543, 4037544, 4037545, 4037546, 4037547, 4037548, 4037549, 4037550, 4037551, 4037552, 4037553, 4037554, 4037555, 4037556, 4037557, 4037558, 4037559, 4037560, 4037561, 4037562, 4037563, 4037564, 4037565, 4037566, 4037567, 4037568, 4037569, 4037570, 4037571, 4037572, 4037573, 4037574, 4037575, 4037576, 4037577, 4037578, 4037579, 4037580, 4037581, 4037582, 4037583, 4037584, 4037585, 4037586, 4037587, 4037588, 4037589, 4037590, 4037591, 4037592, 4037593, 4037594, 4037595, 4037596, 4037597, 4037598, 4037599, 4037600, 4037601, 4037602, 4037603, 4037604, 4037605, 4037606, 4037607, 4037608, 4037609, 4037610, 4037611, 4037612, 4037613, 4037614, 4037615, 4037616, 4037617, 4037618, 4037619, 4037620, 4037621, 4037622, 4037623, 4037624, 4037625, 4037626, 4037627, 4037628, 4037629, 4037630, 4037631, 4037632, 4037633, 4037634, 4037635, 4037636, 4037637, 4037638, 4037639, 4037640, 4037641, 4037642, 4037643, 4037644, 4037645, 4037646, 4037647, 4037648, 4037649, 4037650, 4037651, 4037652, 4037653, 4037654, 4037655, 4037656, 4037657, 4037658, 4037659, 4037660, 4037661, 4037662, 4037663, 4037664, 4037665, 4037666, 4037667, 4037668, 4037669, 4037670, 4037671, 4037672, 4037673, 4037674, 4037675, 4037676, 4037677, 4037678, 4037679, 4037680, 4037681, 4037682, 4037683, 4037684, 4037685, 4037686, 4037687, 4037688, 4037689, 4037690, 4037691, 4037692, 4037693, 4037694, 4037695, 4037696, 4037697, 4037698, 4037699, 4037700, 4037701, 4037702, 4037703, 4037704, 4037705, 4037706, 4037707, 4037708, 4037709, 4037710, 4037711, 4037712, 4037713, 4037714, 4037715, 4037716, 4037717, 4037718, 4037719, 4037720, 4037721, 4037722, 4037723, 4037724, 4037725, 4037726, 4037727, 4037728, 4037729, 4037730, 4037731, 4037732, 4037733, 4037734, 4037735, 4037736, 4037737, 4037738, 4037739, 4037740, 4037741, 4037742, 4037743, 4037744, 4037745, 4037746, 4037747, 4037748, 4037749, 4037750, 4037751, 4037752, 4037753, 4037754, 4037755, 4037756, 4037757, 4037758, 4037759, 4037760, 4037761, 4037762, 4037763, 4037764, 4037765, 4037766, 4037767, 4037768, 4037769, 4037770, 4037771, 4037772, 4037773, 4037774, 4037775, 4037776, 4037777, 4037778, 4037779, 4037780, 4037781, 4037782, 4037783, 4037784, 4037785, 4037786, 4037787, 4037788, 4037789, 4037790, 4037791, 4037792, 4037793, 4037795, 4037796, 4037797, 4037798, 4037799, 4037800, 4037801, 4037802, 4037803, 4037804, 4037805, 4037806, 4037807, 4037808, 4037809, 4037810, 4037811, 4037812, 4037813, 4037814, 4037815, 4037816, 4037817, 4037818, 4037819, 4037820, 4037821, 4037822, 4037823, 4037824, 4037825, 4037826, 4037827, 4037828, 4037829, 4037830, 4037831, 4037832, 4037833, 4037834, 4037835, 4037836, 4037837, 4037838, 4037839, 4037840, 4037841, 4037842, 4037843, 4037844, 4037845, 4037846, 4037847, 4037848, 4037849, 4037850, 4037851, 4037852, 4037853, 4037854, 4037855, 4037856, 4037857, 4037858, 4037859, 4037860, 4037861, 4037862, 4037863, 4037864, 4037865, 4037866, 4037867, 4037868, 4037869, 4037870, 4037871, 4037872, 4037873, 4037874, 4037875, 4037876, 4037877, 4037878, 4037879, 4037880, 4037881, 4037882, 4037883, 4037884, 4037885, 4037886, 4037887, 4037888, 4037889, 4037890, 4037891, 4037892, 4037893, 4037894.4037895, 4037896, 4037897, 4037898, 4037899, 4037900, 4037901, 4037902, 4037903, 4037904, 4037905, 4037906, 4037907, 4037908, 4037909, 4037910, 4037911, 4037912, 4037913, 4037914, 4037915, 4037916, 4037917, 4037918, 4037919, 4037920, 4037921, 4037922, 4037923, 4037924, 4037925, 4037926, 4037927, 4037928, 4037929, 4037930, 4037931, 4037932, 4037933, 4037934, 4037935, 4037936, 4037937, 4037938, 4037939, 4037940, 4037941, 4037942, 4037943, 4037944, 4037945, 4037946, 4037947, 4038703, 4038704, 4038705, 4038706, 4038707, 4038708, 4038709, 4038710, 4038711, 4038712, 4038713, 4038714, 4038715, 4038716, 4038717, 4038718, 4038719, 4038720, 4038721, 4038722, 4038723, 4038724, 4038725, 4038726, 4038727, 4038728, 4038729, 4038730, 4038731, 4038732, 4038733, 4038734, 4038735, 4038736, 4038737, 4038738, 4038739, 4038740, 4038741, 4038742, 4038743, 4038744, 4038745, 4038746, 4038747, 4038748, 4038749, 4038750, 4038751, 4038752, 4038753, 4038754, 4038755, 4038756, 4038757, 4038758, 4038759, 4038760, 4038761, 4038762, 4038763, 4038764, 4038765, 4038766, 4038767, 4038768, 4038769, 4038770, 4038771, 4038772, 4038773, 4038774, 4038775, 4038776, 4038777, 4038778, 4038779, 4038780, 4038781, 4038782, 4038783, 4038784, 4038785, 4038786, 4038787, 4038788, 4038789, 4038790, 4038791, 4038792, 4038793, 4038794, 4038795, 4038796, 4038797, 4038798, 4038799, 4038800, 4038801, 4038802, 4038803, 4038804, 4038805, 4038806, 4038807, 4038808, 4038809, 4038810, 4038811, 4038812, 4038813, 4038814, 4038815, 4038816, 4038817, 4038818, 4038819, 4038820, 4038821, 4038822, 4038823, 4038824, 4038825, 4038826, 4038827, 4038828, 4038829, 4038830, 4038831, 4038832, 4038833, 4038834, 4038835, 4038836, 4038837, 4038838, 4038839, 4038840, 4038841, 4038842, 4038843, 4038844, 4038845, 4038846, 4038847, 4038848, 4038849, 4038850, 4038851, 4038852, 4038853, 4038854, 4038855, 4038856, 4038857, 4038858, 4038859, 4038860, 4038861, 4038862, 4038863, 4038864, 4038865, 4038866, 4038867, 4038868, 4038869, 4038870, 4038871, 4038872, 4038873, 4038874, 4038875, 4038876, 4038877, 4038878, 4038879, 4038880, 4038881, 4038882, 4038883, 4038884, 4038885, 4038886, 4038887, 4038888, 4038889, 4038890, 4038891, 4038892, 4038893, 4038894, 4038895, 4038896, 4038897, 4038898, 4038899, 4038900, 4038901, 4038902, 4038903, 4038904, 4038905, 4038906, 4038907, 4038908, 4038909, 4038910, 4038911, 4038912, 4038913, 4038914, 4038915, 4038916, 4038917, 4038918, 4038919, 4038920, 4038921, 4038922, 4038923, 4038924, 4038925, 4038926, 4038927, 4038928, 4038929, 4038930, 4038931, 4038932, 4038933, 4038934, 4038935, 4038936, 4038937, 4038938, 4038939, 4038940, 4038941, 4038942, 4038943, 4038944, 4038945, 4038946, 4038947, 4038948, 4038949, 4038950, 4038951, 4038952, 4038953, 4038954, 4038955, 4038956, 4038957, 4038958, 4038959, 4038960, 4038961, 4038962, 4038963, 4038964, 4038965, 4038966, 4038967, 4038968, 4038969, 4038970, 4038971, 4038972, 4038973, 4038974, 4038975, 4038976, 4038977, 4038978, 4038979, 4038980, 4038981, 4038982, 4038983, 4038984, 4038985, 4038986, 4038987, 4038988, 4038989, 4038990, 4038991, 4038992, 4038993, 4038994, 4038995, 4038996, 4038997, 4038998, 4038999, 4039000, 4039001, 4039002, 4039003, 4039004, 4039005, 4039006, 4039007, 4039008, 4039009, 4039010, 4039011, 4039012, 4039013, 4039014, 4039015, 4039016, 4039017, 4039018, 4039019, 4039020, 4039021, 4039022, 4039023, 4039024, 4039025, 4039026, 4039027, 4039028, 4039029, 4039030, 4039031, 4039032, 4039033, 4039034, 4039035, 4039036, 4039037, 4039038, 4039039, 4039040, 4039041, 4039042, 4039043, 4039044, 4039045, 4039046, 4039047, 4039048, 4039049, 4039050, 4039051, 4039052, 4039053, 4039054, 4039055, 4039056, 4039057, 4039058, 4039059, 4039060, 4039061, 4039062, 4039063, 4039064, 4039065, 4039066, 4039067, 4039068, 4039069, 4039070, 4039071, 4039072, 4039073, 4039074, 4039075, 4039076, 4039077, 4039078, 4039079, 4039080, 4039081, 4039082, 4039083, 4039084, 4039085, 4039086, 4039087, 4039088, 4039089, 4039090, 4039091, 4039092, 4039093, 4039094, 4039095, 4039096, 4039097, 4039098, 4039099, 4039100, 4039101, 4039102, 4039103, 4039104, 4039105, 4039106, 4039107, 4039108, 4039109, 4039110, 4039111, 4039112, 4039113, 4039114, 4039115, 4039116, 4039117, 4039118, 4039119, 4039120, 4039121, 4039122, 4039123, 4039124, 4039125, 4039126, 4039127, 4039128, 4039129, 4039130, 4039131, 4039132, 4039133, 4039134, 4039135, 4039136, 4039137, 4039138, 4039139, 4039140, 4039141, 4039142, 4039143, 4039144, 4039145, 4039146, 4039147, 4039148, 4039149, 4039150, 4039151, 4039152, 4039153, 4039154, 4039155, 4039156, 4039157, 4039158, 4039159, 4039160, 4039161, 4039162, 4039163, 4039164, 4039165, 4039166, 4039167, 4039168, 4039169, 4039170, 4039171, 4039172, 4039173, 4039174, 4039175, 4039176, 4039177, 4039178, 4039179, 4039180, 4039181, 4039182, 4039183, 4039184, 4039185, 4039186, 4039187, 4039188, 4039189, 4039190, 4039191, 4039192, 4039193, 4039194, 4039195, 4039196, 4039197, 4039198, 4039199, 4039200, 4039201, 4039202, 4040038, 4040039, 4040040, 4040041, 4040042, 4040043, 4040044, 4040045, 4040046, 4040047, 4040048, 4040049, 4040050, 4040051, 4040052, 4040053, 4040054, 4040055, 4040056, 4040057, 4040058, 4040059, 4040060, 4040061, 4040062, 4040063, 4040064, 4040065, 4040066, 4040067, 4040068, 4040069, 4040070, 4040071, 4040072, 4040073, 4040074, 4040075, 4040076, 4040077, 4040078, 4040079, 4040080, 4040081, 4040082, 4040083, 4040084, 4040085, 4040086, 4040087, 4040088, 4040089, 4040090, 4040091, 4040092, 4040093, 4040094, 4040095, 4040096, 4040097, 4040098, 4040099, 4040100, 4040101, 4040102, 4040103, 4040104, 4040105, 4040106, 4040107, 4040108, 4040109, 4040110, 4040111, 4040112, 4040113, 4040114, 4040115, 4040116, 4040117, 4040118, 4040119, 4040120, 4040121, 4040122, 4040123, 4040124, 4040125, 4040126, 4040127, 4040128, 4040129, 4040130, 4040131, 4040132, 4040133, 4040134, 4040135, 4040136, 4040137, 4040138, 4040139, 4040140, 4040141, 4040142, 4040143, 4040144, 4040145, 4040146, 4040147, 4040148, 4040149, 4040150, 4040151, 4040152, 4040153, 4040154, 4040155, 4040156, 4040157, 4040158, 4040159, 4040160, 4040161, 4040162, 4040163, 4040164, 4040165, 4040166, 4040167, 4040168, 4040169, 4040170, 4040171, 4040172, 4040173, 4040174, 4040175, 4040176, 4040177, 4040178, 4040179, 4040180, 4040181, 4040182, 4040183, 4040184, 4040185, 4040186, 4040187, 4040188, 4040189, 4040190, 4040191, 4040192, 4040193, 4040194, 4040195, 4040196, 4040197, 4040198, 4040199, 4040200, 4040201, 4040202, 4040203, 4040204, 4040205, 4040206, 4040207, 4040208, 4040209, 4040210, 4040211, 4040212, 4040213, 4040214, 4040215, 4040216, 4040217, 4040218, 4040219, 4040220, 4040221, 4040222, 4040223, 4040224, 4040225, 4040226, 4040227, 4040228, 4040229, 4040230, 4040231, 4040232, 4040233, 4040234, 4040235, 4040236, 4040237, 4041438, 4041439, 4041440, 4041441, 4041442, 4041443, 4041444, 4041445, 4041446, 4041447, 4041448, 4041449, 4041450, 4041451, 4041452, 4041453, 4041454, 4041455, 4041456, 4041457, 4041458, 4041459, 4041460, 4041461, 4041462, 4041463, 4041464, 4041465, 4041466, 4041467, 4041468, 4041469, 4041470, 4041471, 4041472, 4041473, 4041474, 4041475, 4041476, 4041477, 4041478, 4041479, 4041480, 4041481, 4041482, 4041483, 4041484, 4041485, 4041486, 4041487, 4041488, 4041489, 4041490, 4041491, 4041492, 4041493, 4041494, 4041495, 4041496, 4041497, 4041498, 4041499, 4041500, 4041501, 4041502, 4041503, 4041504, 4041505, 4041506, 4041507, 4041508, 4041509, 4041510, 4041511, 4041512, 4041513, 4041514, 4041515, 4041516, 4041517, 4041518, 4041519, 4041520, 4041521, 4041522, 4041523, 4041524, 4041525, 4041526, 4041527, 4041528, 4041529, 4041530, 4041531, 4041532, 4041533, 4041534, 4041535, 4041536, 4041537, 4041538, 4041539, 4041540, 4041541, 4041542, 4041543, 4041544, 4041545, 4041546, 4041547, 4041548, 4041549, 4041550, 4041551, 4041552, 4041553, 4041554, 4041555, 4041556, 4041557, 4041558, 4041559, 4041560, 4041561, 4041562, 4041563, 4041564, 4041565, 4041566, 4041567, 4041568, 4041569, 4041570, 4041571, 4041572, 4041573, 4041574, 4041575, 4041576, 4041577, 4041578, 4041579, 4041580, 4041581, 4041582, 4041583, 4041584, 4041585, 4041586, 4041587, 4041588, 4041589, 4041590, 4041591, 4041592, 4041593, 4041594, 4041595, 4041596, 4041597, 4041598, 4041599, 4041600, 4041601, 4041602, 4041603, 4041604, 4041605, 4041606, 4041607, 4041608, 4041609, 4041610, 4041611, 4041612, 4041613, 4041614, 4041615, 4041616, 4041617, 4041618, 4041619, 4041620, 4041621, 4041622, 4041623, 4041624, 4041625, 4041626, 4041627, 4041628, 4041629, 4041630, 4041631, 4041632, 4041633, 4041634, 4041635, 4041636, 4041637, 4041638, 4041639, 4041640, 4041641, 4041642, 4041643, 4041644, 4041645, 4041646, 4041647, 4041648, 4041649, 4041650, 4041651, 4041652, 4041653, 4041654, 4041655, 4041656, 4041657, 4041658, 4041659, 4041660, 4041661, 4041662, 4041663, 4041664, 4041665, 4041666, 4041667, 4041668, 4041669, 4041670, 4041671, 4041672, 4041673, 4041674, 4041675, 4041676, 4041677, 4041678, 4041679, 4041680, 4041681, 4041682, 4041683, 4041684, 4041685, 4041686, 4041687, 4041688, 4041689, 4041690, 4041691, 4041692, 4041693, 4041694, 4041695, 4041696, 4041697, 4041698, 4041699, 4041700, 4041701, 4041702, 4041703, 4041704, 4041705, 4041706, 4041707, 4041708, 4041709, 4041710, 4041711, 4041712, 4041713, 4041714, 4041715, 4041716, 4041717, 4041718, 4041719, 4041720, 4041721, 4041722, 4041723, 4041724, 4041725, 4041726, 4041727, 4041728, 4041729, 4041730, 4041731, 4041732, 4041733, 4041734, 4041735, 4041736, 4041737, 4041738, 4041739, 4041740, 4041741, 4041742, 4041743, 4041744, 4041745, 4041746, 4041747, 4041748, 4041749, 4041750, 4041751, 4041752, 4041753, 4041754, 4041755, 4041756, 4041757, 4041758, 4041759, 4041760, 4041761, 4041762, 4041763, 4041764, 4041765, 4041766, 4041767, 4041768, 4041769, 4041770, 4041771, 4041772, 4041773, 4041774, 4041775, 4041776, 4041777, 4041778, 4041779, 4041780, 4041781, 4041782, 4041783, 4041784, 4041785, 4041786, 4041787, 4041788, 4041789, 4041790, 4041791, 4041792, 4041793, 4041794, 4041795, 4041796, 4041797, 4041798, 4041799, 4041800, 4041801, 4041802, 4041803, 4041804, 4041805, 4041806, 4041807, 4041808, 4041809, 4041810, 4041811, 4041812, 4041813, 4041814, 4041815, 4041816, 4041817, 4041818, 4041819, 4041820, 4041821, 4041822, 4041823, 4041824, 4041825, 4041826, 4041827, 4041828, 4041829, 4041830, 4041831, 4041832, 4041833, 4041834, 4041835, 4041836, 4041837, 4041838, 4041839, 4041840, 4041841, 4041842, 4041843, 4041844, 4041845, 4041846, 4041847, 4041848, 4041849, 4041850, 4041851, 4041852, 4041853, 4041854, 4041855, 4041856, 4041857, 4041858, 4041859, 4041860, 4041861, 4041862, 4041863, 4041864, 4041865, 4041866, 4041867, 4041868, 4041869, 4041870, 4041871, 4041872, 4041873, 4041874, 4041875, 4041876, 4041877, 4041878, 4041879, 4041880, 4041881, 4041882, 4041883, 4041884, 4041885, 4041886, 4041887, 4041888, 4041889, 4041890, 4041891, 4041892, 4041893, 4041894, 4041895, 4041896, 4041897, 4041898, 4041899, 4041900, 4041901, 4041902, 4041903, 4041904, 4041905, 4041906, 4041907, 4041908, 4041909, 4041910, 4041911, 4041912, 4041913, 4041914, 4041915, 4041916, 4041917, 4041918, 4041919, 4041920, 4041921, 4041922, 4041923, 4041924, 4041925, 4041926, 4041927, 4041928, 4041929, 4041930, 4041931, 4041932, 4041933, 4041934, 4041935, 4041936, 4041937, 4041938, 4041939, 4041940, 4041941, 4041942, 4041943, 4041944, 4041945, 4041946, 4041947, 4041948, 4041949, 4041950, 4041951, 4041952, 4041953, 4041954, 4041955, 4041956, 4041957, 4041958, 4041959, 4041960, 4041961, 4041962, 4041963, 4041964, 4041965, 4041966, 4041967, 4041968, 4041969, 4041970, 4041971, 4041972, 4041973, 4041974, 4041975, 4041976, 4041977, 4041978, 4041979, 4041980, 4041981, 4041982, 4041983, 4041984, 4041985, 4041986, 4041987, 4041988, 4041989, 4041990, 4041991, 4041992, 4041993, 4041994, 4041995, 4041996, 4041997, 4041998, 4041999, 4042000, 4042001, 4042002, 4042003, 4042004, 4042005, 4042006, 4042007, 4042008, 4042009, 4042010, 4042011, 4042012, 4042013, 4042014, 4042015, 4042016, 4042017, 4042018, 4042019, 4042020, 4042021, 4042022, 4042023, 4042024, 4042025, 4042026, 4042027, 4042028, 4042029, 4042030, 4042031, 4042032, 4042033, 4042034, 4042035, 4042036, 4042037, 4042038, 4042039, 4042040, 4042041, 4042042, 4042043, 4042044, 4042045, 4042046, 4042047, 4042048, 4042049, 4042050, 4042051, 4042052, 4042053, 4042054, 4042055, 4042056, 4042057, 4042058, 4042059, 4042060, 4042061, 4042062, 4042063, 4042064, 4042065, 4042066, 4042067, 4042068, 4042069, 4042070, 4042071, 4042072, 4042073, 4042074, 4042075, 4042076, 4042077, 4042078, 4042079, 4042080, 4042081, 4042082, 4042083, 4042084, 4042085, 4042086, 4042087, 4042088, 4042089, 4042090, 4042091, 4042092, 4042093, 4042094, 4042095, 4042096, 4042097, 4042098, 4042099, 4042100, 4042101, 4042102, 4042103, 4042104, 4042105, 4042106, 4042107, 4042108, 4042109, 4042110, 4042111, 4042112, 4042113, 4042114, 4042115, 4042116, 4042117, 4042118, 4042119, 4042120, 4042121, 4042122, 4042123, 4042124, 4042125, 4042126, 4042127, 4042128, 4042129, 4042130, 4042131, 4042132, 4042133, 4042134, 4042135, 4042136, 4042137, 4042138, 4042139, 4042140, 4042141, 4042142, 4042143, 4042144, 4042145, 4042146, 4042147, 4042148, 4042149, 4042150, 4042151, 4042152, 4042153, 4042154, 4042155, 4042156, 4042157, 4042158, 4042159, 4042160, 4042161, 4042162, 4042163, 4042164, 4042165, 4042166, 4042167, 4042168, 4042169, 4042170, 4042171, 4042172, 4042173, 4042174, 4042175, 4042176, 4042177, 4042178, 4042179, 4042180, 4042181, 4042182, 4042183, 4042184, 4042185, 4042186, 4042187, 4042188, 4042189, 4042190, 4042191, 4042192, 4042193, 4042194, 4042195, 4042196, 4042197, 4042198, 4042199, 4042200, 4042201, 4042202, 4042203, 4042204, 4042205, 4042206, 4042207, 4042208, 4042209, 4042210, 4042211, 4042212, 4042213, 4042214, 4042215, 4042216, 4042217, 4042218, 4042219, 4042220, 4042221, 4042222, 4042223, 4042224, 4042225, 4042226, 4042227, 4042228, 4042229, 4042230, 4042231, 4042232, 4042233, 4042234, 4042235, 4042236, 4042237, 4042238, 4042239, 4042240, 4042241, 4042242, 4042243, 4042244, 4042245, 4042246, 4042247, 4042248, 4042249, 4042250, 4042251, 4042252, 4042253, 4042254, 4042255, 4042256, 4042257, 4042258, 4042259, 4042260, 4042261, 4042262, 4042263, 4042264, 4042265, 4042266, 4042267, 4042268, 4042269, 4042270, 4042271, 4042272, 4042273, 4042274, 4042275, 4042276, 4042277, 4042278, 4042279, 4042280, 4042281, 4042282, 4042283, 4042284, 4042285, 4042286, 4042287, 4042288, 4042289, 4042290, 4042291, 4042292, 4042293, 4042294, 4042295, 4042296, 4042297, 4042298, 4042299, 4042300, 4042301, 4042302, 4042303, 4042304, 4042305, 4042306, 4042307, 4042308, 4042309, 4042310, 4042311, 4042312, 4042313, 4042314, 4042315, 4042316, 4042317, 4042318, 4042319, 4042320, 4042321, 4042322, 4042323, 4042324, 4042325, 4042326, 4042327, 4042328, 4042329, 4042330, 4042331, 4042332, 4042333, 4042334, 4042335, 4042336, 4042337, 4042338, 4042339, 4042340, 4042341, 4042342, 4042343, 4042344, 4042345, 4042346, 4042347, 4042348, 4042349, 4042350, 4042351, 4042352, 4042353, 4042354, 4042355, 4042356, 4042357, 4042358, 4042359, 4042360, 4042361, 4042362, 4042363, 4042364, 4042365, 4042366, 4042367, 4042368, 4042369, 4042370, 4042371, 4042372, 4042373, 4042374, 4042375, 4042376, 4042377, 4042378, 4042379, 4042380, 4042381, 4042382, 4042383, 4042384, 4042385, 4042386, 4042387, 4042388, 4042389, 4042390, 4042391, 4042392, 4042393, 4042394, 4042395, 4042396, 4042397, 4042398, 4042399, 4042400, 4042401, 4042402, 4042403, 4042404, 4042405, 4042406, 4042407, 4042408, 4042409, 4042410, 4042411, 4042412, 4042413, 4042414, 4042415, 4042416, 4042417, 4042418, 4042419, 4042420, 4042421, 4042422, 4042423, 4042424, 4042425, 4042426, 4042427, 4042428, 4042429, 4042430, 4042931, 4042932, 4042933, 4042934, 4042935, 4042936, 4042937, 4043539, 4043540, 4043541, 4043542, 4043543, 4043544, 4043545, 4043546, 4043547, 4043548, 4043549, 4043550, 4043551, 4043552, 4043553, 4043554, 4043555, 4043556, 4043557, 4043558, 4043559, 4043560, 4043561, 4043562, 4043563, 4043564, 4043565, 4043566, 4043567, 4043568, 4043569, 4043570, 4043571, 4043572, 4043573, 4043574, 4043575, 4043576, 4043577, 4043578, 4043579, 4043580, 4043581, 4043582, 4043583, 4043584, 4043585, 4043586, 4043587, 4043588, 4043589, 4043590, 4043591, 4043592, 4043593, 4043594, 4043595, 4043596, 4043597, 4043598, 4043599, 4043600, 4043601, 4043602, 4043603, 4043604, 4043605, 4043606, 4043607, 4043608, 4043609, 4043610, 4043611, 4043612, 4043613, 4043614, 4043615, 4043616, 4043617, 4043618, 4043619, 4043620, 4043621, 4043622, 4043623, 4043624, 4043625, 4043626, 4043627, 4043628, 4043629, 4043630, 4043631, 4043632, 4043633, 4043634, 4043635, 4043636, 4043637, 4043638, 4043639, 4043640, 4043641, 4043642, 4043643, 4043644, 4043645, 4043646, 4043647, 4043648, 4043649, 4043650, 4043651, 4043652, 4043653, 4043654, 4043655, 4043656, 4043657, 4043658, 4043659, 4043660, 4043661, 4043662, 4043663, 4043664, 4043665, 4043666, 4043667, 4043668, 4043669, 4043670, 4043671, 4043672, 4043673, 4043674, 4043675, 4043676, 4043677, 4043678, 4043679, 4043680, 4043681, 4043682, 4043683, 4043684, 4043685, 4043686, 4043687, 4043688, 4043689, 4043690, 4043691, 4043692, 4043693, 4043694, 4043695, 4043696, 4043697, 4043698, 4043699, 4043700, 4043701, 4043702, 4043703, 4043704, 4043705, 4043706, 4043707, 4043708, 4043709, 4043710, 4043711, 4043712, 4043713, 4043714, 4043715, 4043716, 4043717, 4043718, 4043719, 4043720, 4043721, 4043722, 4043723, 4043724, 4043725, 4043726, 4043727, 4043728, 4043729, 4043730, 4043731, 4043732, 4043733, 4043734, 4043735, 4043736, 4043737, 4043738, 4043739, 4043740, 4043741, 4043742, 4043743, 4043744, 4043745, 4043746, 4043747, 4043748, 4043749, 4043750, 4043751, 4043752, 4043753, 4043754, 4043755, 4043756, 4043757, 4043758, 4043759, 4043760, 4043761, 4043762, 4043763, 4043764, 4043765, 4043766, 4043767, 4043768, 4043769, 4043770, 4043771, 4043772, 4043773, 4043774, 4043775, 4043776, 4043777, 4043778, 4043779, 4043780, 4043781, 4043782, 4043783, 4043784, 4043785, 4043786, 4043787, 4043788, 4043789, 4043790, 4043791, 4043792, 4043793, 4043794, 4043795, 4043796, 4043797, 4043798, 4043799, 4043800, 4043801, 4043802, 4043803, 4043804, 4043805, 4043806, 4043807, 4043808, 4043809, 4043810, 4043811, 4043812, 4043813, 4043814, 4043815, 4043816, 4043817, 4043818, 4043819, 4043820, 4043821, 4043822, 4043823, 4043824, 4043825, 4043826, 4043827, 4043828, 4043829, 4043830, 4043831, 4043832, 4043833, 4043834, 4043835, 4043836, 4043837, 4043838, 4043839, 4043840, 4043841, 4043842, 4043843, 4043844, 4043845, 4043846, 4043847, 4043848, 4043849, 4043850, 4043851, 4043852, 4043853, 4043854, 4043855, 4043856, 4043857, 4043858, 4043859, 4043860, 4043861, 4043862, 4043863, 4043864, 4043865, 4043866, 4043867, 4043868, 4043869, 4043870, 4043871, 4043872, 4043873, 4043874, 4043875, 4043876, 4043877, 4043878, 4043879, 4043880, 4043881, 4043882, 4043883, 4043884, 4043885, 4043886, 4043887, 4043888, 4043889, 4043890, 4043891, 4043892, 4043893, 4043894, 4043895, 4043896, 4043897, 4043898, 4043899, 4043900, 4043901, 4043902, 4043903, 4043904, 4043905, 4043906, 4043907, 4043908, 4043909, 4043910, 4043911, 4043912, 4043913, 4043914, 4043915, 4043916, 4043917, 4043918, 4043919, 4043920, 4043921, 4043922, 4043923, 4043924, 4043925, 4043926, 4043927, 4043928, 4043929, 4043930, 4043931, 4043932, 4043933, 4043934, 4043935, 4043936, 4043937, 4043938, 4043939, 4043940, 4043941, 4043942, 4043943, 4043944, 4043945, 4043946, 4043947, 4043948, 4043949, 4043950, 4043951, 4043952, 4043953, 4043954, 4043955, 4043956, 4043957, 4043958, 4043959, 4043960, 4043961, 4043962, 4043963, 4043964, 4043965, 4043966, 4043967, 4043968, 4043969, 4043970, 4043971, 4043972, 4043973, 4043974, 4043975, 4043976, 4043977, 4043978, 4043979, 4043980, 4043981, 4043982, 4043983, 4043984, 4043985, 4043986, 4043987, 4043988, 4043989, 4043990, 4043991, 4043992, 4043993, 4043994, 4043995, 4043996, 4043997, 4043998, 4043999, 4044000, 4044001, 4044002, 4044003, 4044004, 4044005, 4044006, 4044007, 4044008, 4044009, 4044010, 4044011, 4044012, 4044013, 4044014, 4044015, 4044016, 4044017, 4044018, 4044019, 4044020, 4044021, 4044022, 4044023, 4044024, 4044025, 4044026, 4044027, 4044028, 4044029, 4044030, 4044031, 4044032, 4044033, 4044034, 4044035, 4044036, 4044037, 4044038, 4044889, 4044890, 4044891, 4044892, 4044893, 4044894, 4044895, 4044896, 4044897, 4044898, 4044899, 4044900, 4044901, 4044902, 4044903, 4044904, 4044905, 4044906, 4044907, 4044908, 4044909, 4044910, 4044911, 4044912, 4044913, 4044914, 4044915, 4044916, 4044917, 4044918, 4044919, 4044920, 4044921, 4044922, 4044923, 4044924, 4044925, 4044926, 4044927, 4044928, 4044929, 4044930, 4044931, 4044932, 4044933, 4044934, 4044935, 4044936, 4044937, 4044938, 4044939, 4044940, 4044941, 4044942, 4044943, 4044944, 4044945, 4044946, 4044947, 4044948, 4044949, 4044950, 4044951, 4044952, 4044953, 4044954, 4044955, 4044956, 4044957, 4044958, 4044959, 4044960, 4044961, 4044962, 4044963, 4044964, 4044965, 4044966, 4044967, 4044968, 4044969, 4044970, 4044971, 4044972, 4044973, 4044974, 4044975, 4044976, 4044977, 4044978, 4044979, 4044980, 4044981, 4044982, 4044983, 4044984, 4044985, 4044986, 4044987, 4044988, 4044989, 4044990, 4044991, 4044992, 4044993, 4044994, 4044995, 4044996, 4044997, 4044998, 4044999, 4045000, 4045001, 4045002, 4045003, 4045004, 4045005, 4045006, 4045007, 4045008, 4045009, 4045010, 4045011, 4045012, 4045013, 4045014, 4045015, 4045016, 4045017, 4045018, 4045019, 4045020, 4045021, 4045022, 4045023, 4045024, 4045025, 4045026, 4045027, 4045028, 4045029, 4045030, 4045031, 4045032, 4045033, 4045034, 4045035, 4045036, 4045037, 4045038, 4045039, 4045040, 4045041, 4045042, 4045043, 4045044, 4045045, 4045046, 4045047, 4045048, 4045049, 4045050, 4045051, 4045052, 4045053, 4045054, 4045055, 4045056, 4045057, 4045058, 4045059, 4045060, 4045061, 4045062, 4045063, 4045064, 4045065, 4045066, 4045067, 4045068, 4045069, 4045070, 4045071, 4045072, 4045073, 4045074, 4045075, 4045076, 4045077, 4045078, 4045079, 4045080, 4045081, 4045082, 4045083, 4045084, 4045085, 4045086, 4045087, 4045088, 4045089, 4045090, 4045091, 4045092, 4045093, 4045094, 4045095, 4045096, 4045097, 4045098, 4045099, 4045100, 4045101, 4045102, 4045103, 4045104, 4045105, 4045106, 4045107, 4045108, 4045109, 4045110, 4045111, 4045112, 4045113, 4045114, 4045115, 4045116, 4045117, 4045118, 4045119, 4045120, 4045121, 4045122, 4045123, 4045124, 4045125, 4045126, 4045127, 4045128, 4045129, 4045130, 4045131, 4045132, 4045133, 4045134, 4045135, 4045136, 4045137, 4045138, 4045139, 4045140, 4045141, 4045142, 4045143, 4045144, 4045145, 4045146, 4045147, 4045148, 4045149, 4045150, 4045151, 4045152, 4045153, 4045154, 4045155, 4045156, 4045157, 4045158, 4045159, 4045160, 4045161, 4045162, 4045163, 4045164, 4045165, 4045166, 4045167, 4045168, 4045169, 4045170, 4045171, 4045172, 4045173, 4045174, 4045175, 4045176, 4045177, 4045178, 4045179, 4045180, 4045181, 4045182, 4045183, 4045184, 4045185, 4045186, 4045187, 4045188, 4045189, 4045190, 4045191, 4045192, 4045193, 4045194, 4045195, 4045196, 4045197, 4045198, 4045199, 4045200, 4045201, 4045202, 4045203, 4045204, 4045205, 4045206, 4045207, 4045208, 4045209, 4045210, 4045211, 4045212, 4045213, 4045214, 4045215, 4045216, 4045217, 4045218, 4045219, 4045220, 4045221, 4045222, 4045223, 4045224, 4045225, 4045226, 4045227, 4045228, 4045229, 4045230, 4045231, 4045232, 4045233, 4045234, 4045235, 4045236, 4045237, 4045238, 4045239, 4045240, 4045241, 4045242, 4045243, 4045244, 4045245, 4045246, 4045247, 4045248, 4045249, 4045250, 4045251, 4045252, 4045253, 4045254, 4045255, 4045256, 4045257, 4045258, 4045259, 4045260, 4045261, 4045262, 4045263, 4045264, 4045265, 4045266, 4045267, 4045268, 4045269, 4045270, 4045271, 4045272, 4045273, 4045274, 4045275, 4045276, 4045277, 4045278, 4045279, 4045280, 4045281, 4045282, 4045283, 4045284, 4045285, 4045286, 4045287, 4045288, 4045289, 4045290, 4045291, 4045292, 4045293, 4045294, 4045295, 4045296, 4045297, 4045298, 4045299, 4045300, 4045301, 4045302, 4045303, 4045304, 4045305, 4045306, 4045307, 4045308, 4045309, 4045310, 4045311, 4045312, 4045313, 4045314, 4045315, 4045316, 4045317, 4045318, 4045319, 4045320, 4045321, 4045322, 4045323, 4045324, 4045325, 4045326, 4045327, 4045328, 4045329, 4045330, 4045331, 4045332, 4045333, 4045334, 4045335, 4045336, 4045337, 4045338, 4045339, 4045340, 4045341, 4045342, 4045343, 4045344, 4045345, 4045346, 4045347, 4045348, 4045349, 4045350, 4045351, 4045352, 4045353, 4045354, 4045355, 4045356, 4045357, 4045358, 4045359, 4045360, 4045361, 4045362, 4045363, 4045364, 4045365, 4045366, 4045367, 4045368, 4045369, 4045370, 4045371, 4045372, 4045373, 4045374, 4045375, 4045376, 4045377, 4045378, 4045379, 4045380, 4045381, 4045382, 4045383, 4045384, 4045385, 4045386, 4045387, 4045388, 4045889, 4045890, 4045891, 4045892, 4045893, 4045894, 4045895, 4045896, 4045897, 4045898, 4045899, 4045900, 4045901, 4045902, 4045903, 4045904, 4045905, 4045906, 4045907, 4045908, 4045909, 4045910, 4045911, 4045912, 4045913, 4045914, 4045915, 4045916, 4045917, 4045918, 4045919, 4045920, 4045921, 4045922, 4045923, 4045924, 4045925, 4045926, 4045927, 4045928, 4045929, 4045930, 4045931, 4045932, 4045933, 4045934, 4045935, 4045936, 4045937, 4045938, 4045939, 4045940, 4045941, 4045942, 4045943, 4045944, 4045945, 4045946, 4045947, 4045948, 4045949, 4045950, 4045951, 4045952, 4045953, 4045954, 4045955, 4045956, 4045957, 4045958, 4045959, 4045960, 4045961, 4045962, 4045963, 4045964, 4045965, 4045966, 4045967, 4045968, 4045969, 4045970, 4045971, 4045972, 4045973, 4045974, 4045975, 4045976, 4045977, 4045978, 4045979, 4045980, 4045981, 4045982, 4045983, 4045984, 4045985, 4045986, 4045987, 4045988, 4045989, 4045990, 4045991, 4045992, 4045993, 4045994, 4045995, 4045996, 4045997, 4045998, 4045999, 4046000, 4046001, 4046002, 4046003, 4046004, 4046005, 4046006, 4046007, 4046008, 4046009, 4046010, 4046011, 4046012, 4046013, 4046014, 4046015, 4046016, 4046017, 4046018, 4046019, 4046020, 4046021, 4046022, 4046023, 4046024, 4046025, 4046026, 4046027, 4046028, 4046029, 4046030, 4046031, 4046032, 4046033, 4046034, 4046035, 4046036, 4046037, 4046038, 4046039, 4046040, 4046041, 4046042, 4046043, 4046044, 4046045, 4046046, 4046047, 4046048, 4046049, 4046050, 4046051, 4046052, 4046053, 4046054, 4046055, 4046056, 4046057, 4046058, 4046059, 4046060, 4046061, 4046062, 4046063, 4046064, 4046065, 4046066, 4046067, 4046068, 4046069, 4046070, 4046071, 4046072, 4046073, 4046074, 4046075, 4046076, 4046077, 4046078, 4046079, 4046080, 4046081, 4046082, 4046083, 4046084, 4046085, 4046086, 4046087, 4046088, 4046089, 4046090, 4046091, 4046092, 4046093, 4046094, 4046095, 4046096, 4046097, 4046098, 4046099, 4046100, 4046101, 4046102, 4046103, 4046104, 4046105, 4046106, 4046107, 4046108, 4046109, 4046110, 4046111, 4046112, 4046113, 4046114, 4046115, 4046116, 4046117, 4046118, 4046119, 4046120, 4046121, 4046122, 4046123, 4046124, 4046125, 4046126, 4046127, 4046128, 4046129, 4046130, 4046131, 4046132, 4046133, 4046134, 4046135, 4046136, 4046137, 4046138, 4046139, 4046140, 4046141, 4046142, 4046143, 4046144, 4046145, 4046146, 4046147, 4046148, 4046149, 4046150, 4046151, 4046152, 4046153, 4046154, 4046155, 4046156, 4046157, 4046158, 4046159, 4046160, 4046161, 4046162, 4046163, 4046164, 4046165, 4046166, 4046167, 4046168, 4046169, 4046170, 4046171, 4046172, 4046173, 4046174, 4046175, 4046176, 4046177, 4046178, 4046179, 4046180, 4046181, 4046182, 4046183, 4046184, 4046185, 4046186, 4046187, 4046188, 4046189, 4046190, 4046191, 4046192, 4046193, 4046194, 4046195, 4046196, 4046197, 4046198, 4046199, 4046200, 4046201, 4046202, 4046203, 4046204, 4046205, 4046206, 4046207, 4046208, 4046209, 4046210, 4046211, 4046212, 4046213, 4046214, 4046215, 4046216, 4046217, 4046218, 4046219, 4046220, 4046221, 4046222, 4046223, 4046224, 4046225, 4046226, 4046227, 4046228, 4046229, 4046230, 4046231, 4046232, 4046233, 4046234, 4046235, 4046236, 4046237, 4046238, 4046239, 4046240, 4046241, 4046242, 4046243, 4046244, 4046245, 4046246, 4046247, 4046248, 4046249, 4046250, 4046251, 4046252, 4046253, 4046254, 4046255, 4046256, 4046257, 4046258, 4046259, 4046260, 4046261, 4046262, 4046263, 4046264, 4046265, 4046266, 4046267, 4046268, 4046269, 4046270, 4046271, 4046272, 4046273, 4046274, 4046275, 4046276, 4046277, 4046278, 4046279, 4046280, 4046281, 4046282, 4046283, 4046284, 4046285, 4046286, 4046287, 4046288, 4046289, 4046290, 4046291, 4046292, 4046293, 4046294, 4046295, 4046296, 4046297, 4046298, 4046299, 4046300, 4046301, 4046302, 4046303, 4046304, 4046305, 4046306, 4046307, 4046308, 4046309, 4046310, 4046311, 4046312, 4046313, 4046314, 4046315, 4046316, 4046317, 4046318, 4046319, 4046320, 4046321, 4046322, 4046323, 4046324, 4046325, 4046326, 4046327, 4046328, 4046329, 4046330, 4046331, 4046332, 4046333, 4046334, 4046335, 4046336, 4046337, 4046338, 4046339, 4046340, 4046341, 4046342, 4046343, 4046344, 4046345, 4046346, 4046347, 4046348, 4046349, 4046350, 4046351, 4046352, 4046353, 4046354, 4046355, 4046356, 4046357, 4046358, 4046359, 4046360, 4046361, 4046362, 4046363, 4046364, 4046365, 4046366, 4046367, 4046368, 4046369, 4046370, 4046371, 4046372, 4046373, 4046374, 4046375, 4046376, 4046377, 4046378, 4046379, 4046380, 4046381, 4046382, 4046383, 4046384, 4046385, 4046386, 4046387, 4046388, 4047489, 4047490, 4047491, 4047492, 4047493, 4047494, 4047495, 4047496, 4047497, 4047498, 4047499, 4047500, 4047501, 4047502, 4047503, 4047504, 4047505, 4047506, 4047507, 4047508, 4047509, 4047510, 4047511, 4047512, 4047513, 4047514, 4047515, 4047516, 4047517, 4047518, 4047519, 4047520, 4047521, 4047522, 4047523, 4047524, 4047525, 4047526, 4047527, 4047528, 4047529, 4047530, 4047531, 4047532, 4047533, 4047534, 4047535, 4047536, 4047537, 4047538, 4047539, 4047540, 4047541, 4047542, 4047543, 4047544, 4047545, 4047546, 4047547, 4047548, 4047549, 4047550, 4047551, 4047552, 4047553, 4047554, 4047555, 4047556, 4047557, 4047558, 4047559, 4047560, 4047561, 4047562, 4047563, 4047564, 4047565, 4047566, 4047567, 4047568, 4047569, 4047570, 4047571, 4047572, 4047573, 4047574, 4047575, 4047576, 4047577, 4047578, 4047579, 4047580, 4047581, 4047582, 4047583, 4047584, 4047585, 4047586, 4047587, 4047588, 4047589, 4047590, 4047591, 4047592, 4047593, 4047594, 4047595, 4047596, 4047597, 4047598, 4047599, 4047600, 4047601, 4047602, 4047603, 4047604, 4047605, 4047606, 4047607, 4047608, 4047609, 4047610, 4047611, 4047612, 4047613, 4047614, 4047615, 4047616, 4047617, 4047618, 4047619, 4047620, 4047621, 4047622, 4047623, 4047624, 4047625, 4047626, 4047627, 4047628, 4047629, 4047630, 4047631, 4047632, 4047633, 4047634, 4047635, 4047636, 4047637, 4047638, 4047639, 4047640, 4047641, 4047642, 4047643, 4047644, 4047645, 4047646, 4047647, 4047648, 4047649, 4047650, 4047651, 4047652, 4047653, 4047654, 4047655, 4047656, 4047657, 4047658, 4047659, 4047660, 4047661, 4047662, 4047663, 4047664, 4047665, 4047666, 4047667, 4047668, 4047669, 4047670, 4047671, 4047672, 4047673, 4047674, 4047675, 4047676, 4047677, 4047678, 4047679, 4047680, 4047681, 4047682, 4047683, 4047684, 4047685, 4047686, 4047687, 4047688, 4047689, 4047690, 4047691, 4047692, 4047693, 4047694, 4047695, 4047696, 4047697, 4047698, 4047699, 4047700, 4047701, 4047702, 4047703, 4047704, 4047705, 4047706, 4047707, 4047708, 4047709, 4047710, 4047711, 4047712, 4047713, 4047714, 4047715, 4047716, 4047717, 4047718, 4047719, 4047720, 4047721, 4047722, 4047723, 4047724, 4047725, 4047726, 4047727, 4047728, 4047729, 4047730, 4047731, 4047732, 4047733, 4047734, 4047735, 4047736, 4047737, 4047738, 4047739, 4047740, 4047741, 4047742, 4047743, 4047744, 4047745, 4047746, 4047747, 4047748, 4047749, 4047750, 4047751, 4047752, 4047753, 4047754, 4047755, 4047756, 4047757, 4047758, 4047759, 4047760, 4047761, 4047762, 4047763, 4047764, 4047765, 4047766, 4047767, 4047768, 4047769, 4047770, 4047771, 4047772, 4047773, 4047774, 4047775, 4047776, 4047777, 4047778, 4047779, 4047780, 4047781, 4047782, 4047783, 4047784, 4047785, 4047786, 4047787, 4047788, 4100001, 4100002, 4100003, 4100004, 4100005, 4100006, 4100007, 4100008, 4100009, 4100010, 4100011, 4100012, 4100013, 4100014, 4100015, 4100016, 4100017, 4100018, 4100019, 4100020, 4100021, 4100022, 4100023, 4100024, 4100025, 4100026, 4100027, 4100028, 4100029, 4100030, 4103539, 4103540, 4103541, 4103542, 4103543, 4103544, 4103545, 4103546, 4103547, 4103548, 4103549, 4103550, 4103551, 4103552, 4103553, 4103554, 4103555, 4103556, 4103557, 4103558, 4103559, 4103560, 4103561, 4103562, 4103563, 4103564, 4103565, 4103566, 4103567, 4103568, 4103569, 4103570, 4103571, 4103572, 4103573, 4103574, 4103575, 4103576, 4103577, 4103578, 4103579, 4103580, 4103581, 4103582, 4103583, 4103584, 4103585, 4103586, 4103587, 4103588, 4103589, 4103590, 4103591, 4103592, 4103593, 4103594, 4103595, 4103596, 4103597, 4103598, 4103599, 4103600, 4103601, 4103602, 4103603, 4103604, 4103605, 4103606, 4103607, 4103608, 4103609, 4103610, 4103611, 4103612, 4103613, 4103614, 4103615, 4103616, 4103617, 4103618, 4103619, 4103620, 4103621, 4103622, 4103623, 4103624, 4103625, 4103626, 4103627, 4103628, 4103629, 4103630, 4103631, 4103632, 4103633, 4103634, 4103635, 4103636, 4103637, 4103638, 4103639, 4103640, 4103641, 4103642, 4103643, 4103644, 4103645, 4103646, 4103647, 4103648, 4103649, 4103650, 4103651, 4103652, 4103653, 4103654, 4103655, 4103656, 4103657, 4103658, 4103659, 4103660, 4103661, 4103662, 4103663, 4103664, 4103665, 4103666, 4103667, 4103668, 4103669, 4103670, 4103671, 4103672, 4103673, 4103674, 4103675, 4103676, 4103677, 4103678, 4103679, 4103680, 4103681, 4103682, 4103683, 4103684, 4103685, 4103686, 4103687, 4103688, 4103689, 4103690, 4103691, 4103692, 4103693, 4103694, 4103695, 4103696, 4103697, 4103698, 4103699, 4103700, 4103701, 4103702, 4103703, 4103704, 4103705, 4103706, 4103707, 4103708, 4103709, 4103710, 4103711, 4103712, 4103713, 4103714, 4103715, 4103716, 4103717, 4103718, 4103719, 4103720, 4103721, 4103722, 4103723, 4103724, 4103725, 4103726, 4103727, 4103728, 4103729, 4103730, 4103731, 4103732, 4103733, 4103734, 4103735, 4103736, 4103737, 4103738, 4103739, 4103740, 4103741, 4103742, 4103743, 4103744, 4103745, 4103746, 4103747, 4103748, 4103749, 4103750, 4103751, 4103752, 4103753, 4103754, 4103755, 4103756, 4103757, 4103758, 4103759, 4103760, 4103761, 4103762, 4103763, 4103764, 4103765, 4103766, 4103767, 4103768, 4103769, 4103770, 4103771, 4103772, 4103773, 4103774, 4103775, 4103776, 4103777, 4103778, 4103779, 4103780, 4103781, 4103782, 4103783, 4103784, 4103785, 4103786, 4103787, 4103788, 4103789, 4103790, 4103791, 4103792, 4103793, 4103794, 4103795, 4103796, 4103797, 4103798, 4103799, 4103800, 4103801, 4103802, 4103803, 4103804, 4103805, 4103806, 4103807, 4103808, 4103809, 4103810, 4103811, 4103812, 4103813, 4103814, 4103815, 4103816, 4103817, 4103818, 4103819, 4103820, 4103821, 4103822, 4103823, 4103824, 4103825, 4103826, 4103827, 4103828, 4103829, 4103830, 4103831, 4103832, 4103833, 4103834, 4103835, 4103836, 4103837, 4103838, 4103839, 4103840, 4103841, 4103842, 4103843, 4103844, 4103845, 4103846, 4103847, 4103848, 4103849, 4103850, 4103851, 4103852, 4103853, 4103854, 4103855, 4103856, 4103857, 4103858, 4103859, 4103860, 4103861, 4103862, 4103863, 4103864, 4103865, 4103866, 4103867, 4103868, 4103869, 4103870, 4103871, 4103872, 4103873, 4103874, 4103875, 4103876, 4103877, 4103878, 4103879, 4103880, 4103881, 4103882, 4103883, 4103884, 4103885, 4103886, 4103887, 4103888, 4103889, 4103890, 4103891, 4103892, 4103893, 4103894, 4103895, 4103896, 4103897, 4103898, 4103899, 4103900, 4103901, 4103902, 4103903, 4103904, 4103905, 4103906, 4103907, 4103908, 4103909, 4103910, 4103911, 4103912, 4103913, 4103914, 4103915, 4103916, 4103917, 4103918, 4103919, 4103920, 4103921, 4103922, 4103923, 4103924, 4103925, 4103926, 4103927, 4103928, 4103929, 4103930, 4103931, 4103932, 4103933, 4103934, 4103935, 4103936, 4103937, 4103938, 4103939, 4103940, 4103941, 4103942, 4103943, 4103944, 4103945, 4103946, 4103947, 4103948, 4103949, 4103950, 4103951, 4103952, 4103953, 4103954, 4103955.4103956, 4103957, 4103958, 4103959, 4103960, 4103961, 4103962, 4103963, 4103964, 4103965, 4103966, 4103967, 4103968, 4103969, 4103970, 4103971, 4103972, 4103973, 4103974, 4103975, 4103976, 4103977, 4103978, 4103979, 4103980, 4103981, 4103982, 4103983, 4103984, 4103985, 4103986, 4103987, 4103988, 4103989, 4103990, 4103991, 4103992, 4103993, 4103994, 4103995, 4103996, 4103997, 4103998, 4103999, 4104000, 4104001, 4104002, 4104003, 4104004, 4104005, 4104006, 4104007, 4104008, 4104009, 4104010, 4104011, 4104012, 4104013, 4104014, 4104015, 4104016, 4104017, 4104018, 4104019, 4104020, 4104021, 4104022, 4104023, 4104024, 4104025, 4104026, 4104027, 4104028, 4104029, 4104030, 4104031, 4104032, 4104033, 4104034, 4104035, 4104036, 4104037, 4104038, 4105389, 4105390, 4105391, 4105392, 4105393, 4105394, 4105395, 4105396, 4105397, 4105398, 4105399, 4105400, 4105401, 4105402, 4105403, 4105404, 4105405, 4105406, 4105407, 4105408, 4105409, 4105410, 4105411, 4105412, 4105413, 4105414, 4105415, 4105416, 4105417, 4105418, 4105419, 4105420, 4105421, 4105422, 4105423, 4105424, 4105425, 4105426, 4105427, 4105428, 4105429, 4105430, 4105431, 4105432, 4105433, 4105434, 4105435, 4105436, 4105437, 4105438, 4105439, 4105440, 4105441, 4105442, 4105443, 4105444, 4105445, 4105446, 4105447, 4105448, 4105449, 4105450, 4105451, 4105452, 4105453, 4105454, 4105455, 4105456, 4105457, 4105458, 4105459, 4105460, 4105461, 4105462, 4105463, 4105464, 4105465, 4105466, 4105467, 4105468, 4105469, 4105470, 4105471, 4105472, 4105473, 4105474, 4105475, 4105476, 4105477, 4105478, 4105479, 4105480, 4105481, 4105482, 4105483, 4105484, 4105485, 4105486, 4105487, 4105488, 4105489, 4105490, 4105491, 4105492, 4105493, 4105494, 4105495, 4105496, 4105497, 4105498, 4105499, 4105500, 4105501, 4105502, 4105503, 4105504, 4105505, 4105506, 4105507, 4105508, 4105509, 4105510, 4105511, 4105512, 4105513, 4105514, 4105515, 4105516, 4105517, 4105518, 4105519, 4105520, 4105521, 4105522, 4105523, 4105524, 4105525, 4105526, 4105527, 4105528, 4105529, 4105530, 4105531, 4105532, 4105533, 4105534, 4105535, 4105536, 4105537, 4105538, 4105539, 4105540, 4105541, 4105542, 4105543, 4105544, 4105545, 4105546, 4105547, 4105548, 4105549, 4105550, 4105551, 4105552, 4105553, 4105554, 4105555, 4105556, 4105557, 4105558, 4105559, 4105560, 4105561, 4105562, 4105563, 4105564, 4105565, 4105566, 4105567, 4105568, 4105569, 4105570, 4105571, 4105572, 4105573, 4105574, 4105575, 4105576, 4105577, 4105578, 4105579, 4105580, 4105581, 4105582, 4105583, 4105584, 4105585, 4105586, 4105587, 4105588, 4105589, 4105590, 4105591, 4105592, 4105593, 4105594, 4105595, 4105596, 4105597, 4105598, 4105599, 4105600, 4105601, 4105602, 4105603, 4105604, 4105605, 4105606, 4105607, 4105608, 4105609, 4105610, 4105611, 4105612, 4105613, 4105614, 4105615, 4105616, 4105617, 4105618, 4105619, 4105620, 4105621, 4105622, 4105623, 4105624, 4105625, 4105626, 4105627, 4105628, 4105629, 4105630, 4105631, 4105632, 4105633, 4105634, 4105635, 4105636, 4105637, 4105638, 4105639, 4105640, 4105641, 4105642, 4105643, 4105644, 4105645, 4105646, 4105647, 4105648, 4105649, 4105650, 4105651, 4105652, 4105653, 4105654, 4105655, 4105656, 4105657, 4105658, 4105659, 4105660, 4105661, 4105662, 4105663, 4105664, 4105665, 4105666, 4105667, 4105668, 4105669, 4105670, 4105671, 4105672, 4105673, 4105674, 4105675, 4105676, 4105677, 4105678, 4105679, 4105680, 4105681, 4105682, 4105683, 4105684, 4105685, 4105686, 4105687, 4105688, 4105689, 4105690, 4105691, 4105692, 4105693, 4105694, 4105695, 4105696, 4105697, 4105698, 4105699, 4105700, 4105701, 4105702, 4105703, 4105704, 4105705, 4105706, 4105707, 4105708, 4105709, 4105710, 4105711, 4105712, 4105713, 4105714, 4105715, 4105716, 4105717, 4105718, 4105719, 4105720, 4105721, 4105722, 4105723, 4105724, 4105725, 4105726, 4105727, 4105728, 4105729, 4105730, 4105731, 4105732, 4105733, 4105734, 4105735, 4105736, 4105737, 4105738, 4105739, 4105740, 4105741, 4105742, 4105743, 4105744, 4105745, 4105746, 4105747, 4105748, 4105749, 4105750, 4105751, 4105752, 4105753, 4105754, 4105755, 4105756, 4105757, 4105758, 4105759, 4105760, 4105761, 4105762, 4105763, 4105764, 4105765, 4105766, 4105767, 4105768, 4105769, 4105770, 4105771, 4105772, 4105773, 4105774, 4105775, 4105776, 4105777, 4105778, 4105779, 4105780, 4105781, 4105782, 4105783, 4105784, 4105785, 4105786, 4105787, 4105788, 4105789, 4105790, 4105791, 4105792, 4105793, 4105794, 4105795, 4105796, 4105797, 4105798, 4105799, 4105800, 4105801, 4105802, 4105803, 4105804, 4105805, 4105806, 4105807, 4105808, 4105809, 4105810, 4105811, 4105812, 4105813, 4105814, 4105815, 4105816, 4105817, 4105818, 4105819, 4105820, 4105821, 4105822, 4105823, 4105824, 4105825, 4105826, 4105827, 4105828, 4105829, 4105830, 4105831, 4105832, 4105833, 4105834, 4105835, 4105836, 4105837, 4105838, 4105839, 4105840, 4105841, 4105842, 4105843, 4105844, 4105845, 4105846, 4105847, 4105848, 4105849, 4105850, 4105851, 4105852, 4105853, 4105854, 4105855, 4105856, 4105857, 4105858, 4105859, 4105860, 4105861, 4105862, 4105863, 4105864, 4105865, 4105866, 4105867, 4105868, 4105869, 4105870, 4105871, 4105872, 4105873, 4105874, 4105875, 4105876, 4105877, 4105878, 4105879, 4105880, 4105881, 4105882, 4105883, 4105884, 4105885, 4105886, 4105887, 4105888, 4107905, 4107906, 4107907, 4107908, 4107909, 4107910, 4107911, 4107912, 4107913, 4107914, 4107915, 4107916, 4107917, 4107918, 4107919, 4107920, 4107921, 4107922, 4107923, 4107924, 4107925, 4107926, 4107927, 4107928, 4107929, 4107930, 4107931, 4107932, 4107933, 4107934, 4107935, 4107936, 4107937, 4107938, 4107939, 4107940, 4107941, 4107942, 4107943, 4107944, 4107945, 4107946, 4107947, 4107948, 4107949, 4107950, 4107951, 4107952, 4107953, 4107954, 4107955, 4107956, 4107957, 4107958, 4107959, 4107960, 4107961, 4107962, 4107963, 4107964, 4107965, 4107966, 4107967, 4107968, 4107969, 4107970, 4107971, 4107972, 4107973, 4107974, 4107975, 4107976, 4107977, 4107978, 4107979, 4107980, 4107981, 4107982, 4107983, 4107984, 4107985, 4107986, 4107987, 4107988, 4107989, 4107990, 4107991, 4107992, 4107993, 4107994, 4107995, 4107996, 4107997, 4107998, 4107999, 4108000, 4108001, 4108002, 4108003, 4108004, 4108005, 4108006, 4108007, 4108008, 4108009, 4108010, 4108011, 4108012, 4108013, 4108014, 4108015, 4108016, 4108017, 4108018, 4108019, 4108020, 4108021, 4108022, 4108023, 4108024, 4108025, 4108026, 4108027, 4108028, 4108029, 4108030, 4108031, 4108032, 4108033, 4108034, 4108035, 4108036, 4108037, 4108038, 4108039, 4108040, 4108041, 4108042, 4108043, 4108044, 4108045, 4108046, 4108047, 4108048, 4108049, 4108050, 4108051, 4108052, 4108053, 4108054, 4108055, 4108056, 4108057, 4108058, 4108059, 4108060, 4108061, 4108062, 4108063, 4108064, 4108065, 4108066, 4108067, 4108068, 4108069, 4108070, 4108071, 4108072, 4108073, 4108074, 4108075, 4108076, 4108077, 4108078, 4108079, 4108080, 4108081, 4108082, 4108083, 4108084, 4108085, 4108086, 4108087, 4108088, 4108089, 4108090, 4108091, 4108092, 4108093, 4108094, 4108095, 4108096, 4108097, 4108098, 4108099, 4108100, 4108101, 4108102, 4108103, 4108104, 4110105, 4110106, 4110107, 4110108, 4110109, 4110110, 4110111, 4110112, 4110113, 4110114, 4110115, 4110116, 4110117, 4110118, 4110119, 4110120, 4110121, 4110122, 4110123, 4110124, 4110125, 4110126, 4110127, 4110128, 4110129, 4110130, 4110131, 4110132, 4110133, 4110134, 4110135, 4110136, 4110137, 4110138, 4110139, 4110140, 4110141, 4110142, 4110143, 4110144, 4110145, 4110146, 4110147, 4110148, 4110149, 4110150, 4110151, 4110152, 4110153, 4110154, 4110155, 4110156, 4110157, 4110158, 4110159, 4110160, 4110161, 4110162, 4110163, 4110164, 4110165, 4110166, 4110167, 4110168, 4110169, 4110170, 4110171, 4110172, 4110173, 4110174, 4110175, 4110176, 4110177, 4110178, 4110179, 4110180, 4110181, 4110182, 4110183, 4110184, 4110185, 4110186, 4110187, 4110188, 4110189, 4110190, 4110191, 4110192, 4110193, 4110194, 4110195, 4110196, 4110197, 4110198, 4110199, 4110200, 4110201, 4110202, 4110203, 4110204, 4110205, 4110206, 4110207, 4110208, 4110209, 4110210, 4110211, 4110212, 4110213, 4110214, 4110215, 4110216, 4110217, 4110218, 4110219, 4110220, 4110221, 4110222, 4110223, 4110224, 4110225, 4110226, 4110227, 4110228, 4110229, 4110230, 4110231, 4110232, 4110233, 4110234, 4110235, 4110236, 4110237, 4110238, 4110239, 4110240, 4110241, 4110242, 4110243, 4110244, 4110245, 4110246, 4110247, 4110248, 4110249, 4110250, 4110251, 4110252, 4110253, 4110254, 4110255, 4110256, 4110257, 4110258, 4110259, 4110260, 4110261, 4110262, 4110263, 4110264, 4110265, 4110266, 4110267, 4110268, 4110269, 4110270, 4110271, 4110272, 4110273, 4110274, 4110275, 4110276, 4110277, 4110278, 4110279, 4110280, 4110281, 4110282, 4110283, 4110284, 4110285, 4110286, 4110287, 4110288, 4110289, 4110290, 4110291, 4110292, 4110293, 4110294, 4110295, 4110296, 4110297, 4110298, 4110299, 4110300, 4110301, 4110302, 4110303, 4110304, 4110305, 4110306, 4110307, 4110308, 4110309, 4110310, 4110321, 4110322, 4110323, 4110324, 4110325, 4110326, 4110327, 4110328, 4110329, 4110330, 4110331, 4110332, 4110333, 4110334, 4110335, 4110336, 4110337, 4110338, 4110339, 4110340, 4110341, 4110342, 4110343, 4110344, 4110345, 4110346, 4110347, 4110348, 4110349, 4110350, 4110351, 4110352, 4110353, 4110354, 4110355, 4110356, 4110357, 4110358, 4110359, 4110360, 4110361, 4110362, 4110363, 4110364, 4110365, 4110366, 4110367, 4110368, 4110369, 4110370, 4110371, 4110372, 4110373, 4110374, 4110375, 4110376, 4110377, 4110378, 4110379, 4110380, 4110381, 4110382, 4110383, 4110384, 4110385, 4110386, 4110387, 4110388, 4110389, 4110390, 4110391, 4110392, 4110393, 4110394, 4110395, 4110396, 4110397, 4110398, 4110399, 4110400, 4110401, 4110402, 4110403, 4110404, 4110405, 4110406, 4110407, 4110408, 4110409, 4110410, 4110411, 4110412, 4110413, 4110414, 4110415, 4110416, 4110417, 4110418, 4110419, 4110420, 4110421, 4110422, 4110423, 4110424, 4110425, 4110426, 4110427, 4110428, 4110429, 4110430, 4110431, 4110432, 4110433, 4110434, 4110435, 4110436, 4110437, 4110438, 4110439, 4110440, 4110441, 4110442, 4110443, 4110444, 4110445, 4110446, 4110447, 4110448, 4110449, 4110450, 4110451, 4110452, 4110453, 4110454, 4110455, 4110456, 4110457, 4110458, 4110459, 4110460, 4110461, 4110462, 4110463, 4110464, 4110465, 4110466, 4110467, 4110468, 4110469, 4110470, 4110471, 4110472, 4110473, 4110474, 4110475, 4110476, 4110477, 4110478, 4110479, 4110480, 4110481, 4110482, 4110483, 4110484, 4110485, 4110486, 4110487, 4110488, 4110489, 4110490, 4110491, 4110492, 4110493, 4110494, 4110495, 4110610, 4110611, 4110612, 4110613, 4110614, 4110615, 4110616, 4110617, 4110618, 4110619, 4110621, 4110632, 4110633, 4110634, 4110635, 4110637, 4110638, 4110639, 4110640, 4110641, 4110772, 4110814, 4110815, 4110816, 4110817, 4110818, 4110819, 4110820, 4110821, 4110822, 4110823, 4110824, 4110825, 4110826, 4110827, 4110828, 4110829, 4110830, 4110831, 4110832, 4110833, 4110834, 4110835, 4110836, 4110837, 4110838, 4110839, 4110840, 4110841, 4110842, 4110843, 4110844, 4110845, 4110846, 4110847, 4110848, 4110849, 4110850, 4110851, 4110852, 4110853, 4110854, 4110855, 4110856, 4110857, 4110858, 4110859, 4110860, 4110861, 4110862, 4110863, 4110864, 4110865, 4110866, 4110867, 4110868, 4110869, 4110870, 4110871, 4110872, 4110873, 4110874, 4110875, 4110876, 4110877, 4110878, 4110879, 4110880, 4110881, 4110882, 4110883, 4110884, 4110885, 4110886, 4110887, 4110888, 4110889, 4110890, 4110891, 4110892, 4110893, 4110894, 4110895, 4110896, 4110897, 4110898, 4110899, 4110900, 4110901, 4110904, 4110905, 4110906, 4110907, 4110908, 4110909, 4110911, 4110912, 4110913, 4110915, 4110916, 4110917, 4110918, 4110919, 4110920, 4110921, 4110922, 4110923, 4110924, 4110925, 4110926, 4110927, 4110928, 4110929, 4110930, 4110931, 4110932, 4110933, 4110934, 4110935, 4110936, 4110938, 4110939, 4110940, 4110941, 4110942, 4110943, 4110944, 4110945, 4110946, 4110947, 4110948, 4110949, 4110950, 4110951, 4110952, 4110953, 4110954, 4110955, 4110956, 4110957, 4110958, 4110959, 4110960, 4110961, 4110962, 4110963, 4110964, 4110965, 4110966, 4110967, 4110968, 4110969, 4110970, 4110971, 4110972, 4110973, 4110974, 4110975, 4110976, 4110977, 4110978, 4110979, 4110980, 4110981, 4110982, 4110983, 4110984, 4110985, 4110986, 4110987, 4110988, 4110989, 4110990, 4110991, 4110992, 4110993, 4110994, 4110995, 4110996, 4110997, 4110998, 4110999, 4111000, 4111001, 4111002, 4111003, 4111004, 4111005, 4111006, 4111007, 4111008, 4111009, 4111010, 4111011, 4111012, 4111013, 4111014, 4111015, 4111016, 4111017, 4111018, 4111019, 4111020, 4111021, 4111022, 4111023, 4111024, 4111025, 4111026, 4111027, 4111028, 4111029, 4111030, 4111031, 4111032, 4111033, 4111034, 4111035, 4111036, 4111037, 4111038, 4111039, 4111040, 4111041, 4111042, 4111043, 4111044, 4111045, 4111046, 4111047, 4111048, 4111049, 4111050, 4111051, 4111052, 4111053, 4111054, 4111055, 4111056, 4111057, 4111058, 4111059, 4111060, 4111061, 4111062, 4111063, 4111064, 4111065, 4111066, 4111067, 4111068, 4111069, 4111070, 4111071, 4111072, 4111073, 4111074, 4111075, 4111076, 4111077, 4111078, 4111079, 4111080, 4111081, 4111082, 4111083, 4111084, 4111094, 4111103, 4111104, 4111111, 4111112, 4111113, 4111114, 4111115, 4111116, 4111117, 4111118, 4111119, 4111120, 4111121, 4111122, 4111135, 4111136, 4111137, 4111138, 4111139, 4111140, 4111141, 4111142, 4111143, 4111144, 4111145, 4111146, 4111147, 4111148, 4111149, 4111150, 4111151, 4111152, 4111153, 4111154, 4111155, 4111156, 4111157, 4111158, 4111159, 4111160, 4111161, 4111162, 4111163, 4111164, 4111165, 4111166, 4111167, 4111168, 4111169, 4111170, 4111171, 4111172, 4111173, 4111174, 4111175, 4111176, 4111177, 4111178, 4111179, 4111180, 4111181, 4111182, 4111183, 4111184, 4111185, 4111186, 4111187, 4111188, 4111189, 4111190, 4111191, 4111192, 4111193, 4111194, 4111195, 4111196, 4111197, 4111198, 4111199, 4111200, 4111201, 4111202, 4111203, 4111204, 4111205, 4111206, 4111207, 4111208, 4111209, 4111210, 4111211, 4111212, 4111213, 4111214, 4111215, 4111216, 4111217, 4111218, 4111219, 4111220, 4111221, 4111222, 4111223, 4111224, 4111225, 4111226, 4111227, 4111228, 4111229, 4111230, 4111231, 4111232, 4111233, 4111234, 4111235, 4111236, 4111237, 4111238, 4111239, 4111240, 4111241, 4111242, 4111243, 4111244, 4111245, 4111246, 4111247, 4111248, 4111249, 4111250, 4111251, 4111252, 4111253, 4111254, 4111255, 4111256, 4111257, 4111258, 4111259, 4111260, 4111261, 4111262, 4111263, 4111264, 4111265, 4111266, 4111267, 4111268, 4111269, 4111270, 4111271, 4111272, 4111273, 4111274, 4111275, 4111276, 4111277, 4111278, 4111279, 4111280, 4111281, 4111282, 4111283, 4111284, 4111285, 4111286, 4111287, 4111288, 4111289, 4111290, 4111291, 4111292, 4111293, 4111294, 4111295, 4111296, 4111297, 4111298, 4111299, 4111300, 4111301, 4111302, 4111303, 4111304, 4111305, 4111306, 4111307, 4111308, 4111309, 4111310, 4111311, 4111312, 4111313, 4111314, 4111315, 4111316, 4111317, 4111318, 4111319, 4111320, 4111321, 4111322, 4111323, 4111324, 4111325, 4111326, 4111327, 4111328, 4111329, 4111330, 4111331, 4111332, 4111333, 4111334, 4111335, 4111336, 4111337, 4111338, 4111339, 4111340, 4111341, 4111342, 4111343, 4111344, 4111345, 4111346, 4111347, 4111348, 4111349, 4111350, 4111352, 4111353, 4111354, 4111355, 4111356, 4111357, 4111358, 4111359, 4111360, 4111361, 4111362, 4111363, 4111364, 4111365, 4111366, 4111367, 4111368, 4111369, 4111370, 4111371, 4111372, 4111373, 4111374, 4111375, 4111376, 4111377, 4111378, 4111379, 4111380, 4111381, 4111382, 4111383, 4111384, 4111385, 4111386, 4111387, 4111388, 4111389, 4111390, 4111391, 4111392, 4111393, 4111394, 4111395, 4111396, 4111397, 4111398, 4111399, 4111400, 4111401, 4111402, 4111403, 4111404, 4111406, 4111407, 4111408, 4111409, 4111410, 4111411, 4111421, 4111422, 4111423, 4111424, 4111425, 4111426, 4111427, 4111428, 4111682, 4111683, 4111684, 4111685, 4111686, 4111687, 4111688, 4111689, 4111690, 4111691, 4111692, 4111694, 4111695, 4111696, 4111697, 4111698, 4111699, 4111700, 4111701, 4111702, 4111703, 4111704, 4111705, 4111706, 4111707, 4111708, 4111709, 4111710, 4111711, 4111712, 4111713, 4111714, 4111715, 4111716, 4111717, 4111718, 4111719, 4111720, 4111721, 4111722, 4111723, 4111724, 4111725, 4111726, 4111727, 4111728, 4111729, 4111730, 4111731, 4111732, 4111733, 4111734, 4111735, 4111736, 4111737, 4111738, 4111739, 4111740, 4111741, 4111742, 4111743, 4111744, 4111745, 4111768, 4111769, 4111807, 4111808, 4111809, 4111810, 4111811, 4111812, 4111813, 4111814, 4111815, 4111816, 4111817, 4111818, 4111819, 4111820, 4111821, 4111822, 4111823, 4111824, 4111844, 4111845, 4111846, 4111847, 4111848, 4111849, 4111850, 4111851, 4111853, 4111854, 4111855, 4111856, 4111857, 4111858, 4111859, 4111860, 4111861, 4111994, 4111999, 4112000, 4112001, 4112002, 4112003, 4112004, 4112005, 4112006, 4112007, 4112008, 4112009, 4112010, 4112011, 4112013, 4112014, 4112015, 4112016, 4112017, 4112018, 4112019, 4112020, 4112021, 4112022, 4112023, 4112024, 4112025, 4112026, 4112027, 4112028, 4112029, 4112030, 4112031, 4112032, 4112033, 4112034, 4112035, 4112036, 4112037, 4112038, 4112039, 4112040, 4112041, 4112042, 4112043, 4112044, 4112045, 4112046, 4112047, 4112048, 4112049, 4112050, 4112051, 4112052, 4112053, 4112054, 4112055, 4112056, 4112057, 4112058, 4112059, 4112060, 4112061, 4112062, 4112063, 4112064, 4112065, 4112066, 4112067, 4112068, 4112069, 4112070, 4112071, 4112072, 4112073, 4112074, 4112075, 4112076, 4112077, 4112078, 4112079, 4112080, 4112081, 4112082, 4112090, 4112091, 4112092, 4112093, 4112094, 4112095, 4112096, 4112097, 4112098, 4112108, 4112123, 4112124, 4112125, 4112126, 4112127, 4112128, 4112166, 4112172, 4112173, 4112174, 4112175, 4112176, 4112177, 4112178, 4112179, 4112180, 4112181, 4112182, 4112183, 4112184, 4112185, 4112186, 4112187, 4112188, 4112189, 4112190, 4112191, 4112192, 4112193, 4112194, 4112195, 4112196, 4112197, 4112198, 4112199, 4112200, 4112201, 4112202, 4112203, 4112204, 4112205, 4112206, 4112207, 4112208, 4112209, 4112210, 4112211, 4112212, 4112213, 4112214, 4112215, 4112216, 4112217, 4112218, 4112219, 4112220, 4112221, 4112222, 4112223, 4112224, 4112225, 4112226, 4112227, 4112228, 4112229, 4112230, 4112231, 4112232, 4112233, 4112234, 4112235, 4112236, 4112237, 4112238, 4112239, 4112240, 4112241, 4112242, 4112243, 4112244, 4112373, 4112378, 4112379, 4113965, 4113975, 4113987, 4113989, 4114002, 4114005, 4114007, 4114016, 4114017, 4114034, 4114041, 4114043, 4114044, 4114130, 4114134, 4114136, 4114139, 4114144, 4114145, 4114149, 4114150, 4114151, 4114159, 4114160, 4114168, 4114175, 4114192, 4114193, 4114195, 4114198, 4114201, 4114211, 4114212, 4114213, 4114214, 4114216, 4114220, 4114224, 4114232, 4114234, 4114238, 4114240, 4114241, 4114245, 4114248, 4114254, 4114258, 4114262, 4114266, and 4114272. |
Recalling Firm/ Manufacturer |
Cardiac Science Corporation 3303 Monte Villa Pkwy Bothell WA 98021-8969
|
For Additional Information Contact | Bev Magrane 425-402-2000 |
Manufacturer Reason for Recall | Potential for AED to not deliver defibrillation therapy. |
FDA Determined Cause 2 | Software design |
Action | On October 13, 2008 Cardiac Science began sending via Fed-Ex letters and software upgrade kits. Mailing of letters and software upgrade kits is anticipated to be completed by November 21, 2008. The letter instructs consignees to install the software update, assure that the AED information test file generated by the update software is returned to the firm, and complete the postage paid response card and return it to the firm. If the affected devices are no longer in the possession of the consignee, the consignee is to contact the firm and forward the update kit and letter to the person who has the AED in their possession. A separate letter is being sent to distributors. The letter advised to install the software upgrades for any units in stock or to forward the software upgrade along with the end user letter enclosed and that the AED information text file generated by the update software must be returned to the firm. The firm will visit the locations of individual national account customers and install the update. OEM partners will be notified and provided the information and update kits to provide to their customers. |
Quantity in Commerce | 9,633 devices |
Distribution | Worldwide Distribution: USA, Puerto Rico, Virgin Islands, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czechoslovakia, Denmark, Dominican Republic, France, Germany, Greece, Guam, Hong Kong, Hungary, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Kuwait, Libya, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, UK, and Ukraine. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = MKJ
|
|
|
|