• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Stryker Renaissance Series Stretchers

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Stryker Renaissance Series Stretcherssee related information
Date Initiated by FirmMarch 19, 2010
Date PostedAugust 09, 2010
Recall Status1 Terminated 3 on May 02, 2011
Recall NumberZ-2185-2010
Recall Event ID 55139
510(K)NumberK022309 
Product Classification Powered Wheeled Stretcher - Product Code INK
ProductStryker Renaissance Series Stretchers, Model 1231Z Emergency Care with Zoom, Stryker Medical, Portage, MI.
Code Information Model 1231: 39 Units 0310049178, 0310049179, 0311058120, 0312050809, 0407059532, 0407059533, 0407059656, 0407059657, 0407059851, 0407059852, 0407059853, 0407059854, 0407059855, 0407059856, 0407059857, 0407059858, 0407059859, 0407059860, 0407059861, 0407059862, 0407059863, 0407059864, 0407059865, 0407059866, 0407059867, 0407059868, 0407059869, 0407059870, 0408059073, 0408059123, 0408059124, 0408059871, 0408059872, 0408059873, 0408059874, 0408059875, 0408059876, 0408059877, and 0409062555. 
FEI Number 1831750
Recalling Firm/
Manufacturer
Stryker Medical Div. of Stryker Corporation
3800 E. Centre Ave.
Portage MI 49002
For Additional Information Contact
800-787-9537 Ext. 3
Manufacturer Reason
for Recall
The retractable cord stretchers were fitted with power cords, manufactured by Electri-Cord, whose prongs have the potential to fracture inside the molded section of the plug, causing melting and charring.
FDA Determined
Cause 2
Component design/selection
ActionThe firm has sent Urgent - Medical Device Recall letters, dated March 19, 2010, to all direct accounts notifying them of the recall. The letter identified the affected product and described the reason for recall. Customers were asked to locate the stretchers identified in the letter and visually inspect each cord. If the plug shows signs of excessive wear, etc. the stretcher is to be taken out of service immediately and the plug removed. Customers are to call the number provided in order to be placed on the priority replacement list. If the cord does not show signs of damage then the stretcher can remain in service until a replacement cord is available. The enclosed post card should be completed and returned to the firm. A copy of the notice should be forwarded to any users that stretchers have been sold or loaned to. The firm should be provided with the serial numbers of any stretchers that have been disposed of. Questions should be directed towards Stryker at 800-STRYKER.
Quantity in Commerce39 Stretchers
DistributionWorldwide Distribution -- USA, Canada, Brazil, Japan, Singapore, Spain, and the Far East.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = INK
-
-