• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT / ABBOTT DIABETES CARE INC. FREESTYLE LIBRE 14-DAY CGM; SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT / ABBOTT DIABETES CARE INC. FREESTYLE LIBRE 14-DAY CGM; SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER Back to Search Results
Model Number JNGY296-T3155
Device Problems Break (1069); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Insufficient Information (4580)
Event Date 02/02/2021
Event Type  malfunction  
Event Description
I use an abbott freestyle libre 14 day cgm and have been successfully using this product for over a year.The sensor i started on (b)(6) 2021 worked ok for three days then starting (b)(6) 2021 began giving intermittent readings that seemed intermittent and inaccurate.I checked it against a finger stick and found it to be off by 44%.The cgm read 90 and finger stick was 163.A replacement sensor worked ok.However, upon examination when the cgm was removed i have found that the mechanism that attached the sensor probe to the electronics module appears to be breaking.The product is a two part unit which is latched together when applied to the skin.There are three small plastic barbed prongs which appear to latch the probe to the electronics module.These hold the probe conductive contacts against the circuit board pads for enabling the measurements.I found that all three were broken so that the only thing holding the probe to the module was the fact that the adhesive between the module and my skin was captivating the probe.My background is in medical electronics and i took pictures with my microscope to see this.Additionally, i have carefully performed the same examination on a subsequent cgm module and found that the prongs on this one were also all broken.I assume that the purpose of these prongs is to firmly hold the probe to the module so the electrical contact can be established and maintained.The cgm sensor was removed without mechanical stress so as not to cause any damage to the assembly as i used an adhesive solvent (d-limonene / orange peel oil) for easy removal.I thought i would notify you and if i am wrong with my assumptions, you can disregard, but you may want to investigate, especially if the same mechanism is used in a feedback loop with an insulin pump.This could be a patient safety issue.Fda safety report id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14-DAY CGM
Type of Device
SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER
Manufacturer (Section D)
ABBOTT / ABBOTT DIABETES CARE INC.
MDR Report Key11602790
MDR Text Key243643038
Report NumberMW5100485
Device Sequence Number2
Product Code NBW
Combination Product (y/n)Y
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Patient
Type of Report Initial
Report Date 03/30/2021
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received03/31/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberJNGY296-T3155
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N
Patient Sequence Number1
Patient Age69 YR
Patient Weight69
-
-