• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS LUCEA; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS LUCEA; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARD368614998
Device Problems Crack (1135); Device Fell (4014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/14/2018
Event Type  malfunction  
Manufacturer Narrative
The issue is being investigated by manufacturing site.(b)(4).
 
Event Description
On 14th september, 2018 maquet (b)(4) became aware of an incident with one of surgical lights- lucea.As it was stated, there were a cracks near switch and there was a possibility that some particles fell down.We decided to report the issue in abundance of caution as any particles falling might be a source of contamination.Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
(b)(4).Exemption # e2018005.(b)(4).The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
Getinge became aware of an issue with a lucea device.As it was stated, cracks on the cover of the light have been found during daily checkup of the device.There was no injury reported however we decided to report the issue based on the potential related to parts falling into the sterile field, which may be a source of contamination.It was established that when the event occurred, the surgical light did not meet its specification and it contributed to event.None of the provided information indicate that upon the event occurrence the device was being used for patient treatment.During the investigation it was found that in the past the reported scenario has never lead to serious injury or worse, to death.Cover cracking is indicated by our product experts to likely be caused by combination of different factors such as: mechanical stress, environmental conditions (such a high temperature and humidity variations during transport and storage), use of inappropriate cleaning/disinfections products, or inappropriate cleaning and disinfection protocols.We believe that all remaining devices are performing correctly in the market.We also believe that if the preventive maintenance would have been followed the incident could have been avoided.Given the circumstances we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Event Description
Manufacturer's reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
orléans cedex 2
FR 
MDR Report Key7958379
MDR Text Key123673862
Report Number9710055-2018-00098
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
PMA/PMN Number
K113679
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 04/14/2020
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received10/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberARD368614998
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-