|
Class 2 Device Recall NeuroMuscular Transmission Module, ENMT |
 |
Date Initiated by Firm |
April 09, 2014 |
Date Posted |
May 21, 2014 |
Recall Status1 |
Terminated 3 on September 28, 2015 |
Recall Number |
Z-1640-2014 |
Recall Event ID |
67620 |
510(K)Number |
K051635 K955026
|
Product Classification |
Stimulator, nerve, peripheral, electric - Product Code KOI
|
Product |
GE Healthcare, NeuroMuscular Transmission Module, E-NMT-01
Product Usage: The Datex-Ohmeda S/5TM E-NMT module is indicated for monitoring the relaxation of the patient and regional block stimulation for nerve location. |
Code Information |
Mfg Lot or Serial # SJC13057999HA SJC13058001HA SJC13058005HA SJC13058006HA SJC13058007HA SJC13058008HA SJC13058009HA SJC13058010HA SJC13058011HA SJC13058012HA SJC13058013HA SJC13058014HA SJC13058015HA SJC13058017HA SJC13058018HA SJC13058020HA SJC13058022HA SJC13058024HA SJC13058025HA SJC13058028HA SJC13172551HA SJC13172553HA SJC13172554HA SJC13172555HA SJC13172556HA SJC13172558HA SJC13172559HA SJC13172562HA SJC13172564HA SJC13172565HA SJC13172568HA SJC13172620HA SJC13172630HA SJC13172766HA SJC13172767HA SJC13172769HA SJC13172770HA SJC13172772HA SJC13172774HA SJC13172775HA SJC13172776HA SJC13193229HA SJC13193233HA SJC13193234HA SJC13203267HA SJC13203268HA SJC13203269HA SJC13234522HA SJC13234523HA SJC13234524HA SJC13234528HA SJC13234529HA SJC13234530HA SJC13234531HA SJC13234532HA SJC13234533HA SJC13234534HA SJC13234535HA SJC13234538HA SJC13234540HA SJC13234541HA SJC13234542HA SJC13234543HA SJC13244575HA SJC13244581HA SJC13244583HA SJC13244584HA SJC13244588HA SJC13244589HA SJC13244590HA SJC13244592HA SJC13244593HA SJC13244594HA SJC13244665HA SJC13244666HA SJC13244668HA SJC13244669HA SJC13244670HA SJC13244672HA SJC13244674HA SJC13244675HA SJC13244677HA SJC13244678HA SJC13244679HA SJC13244680HA SJC13244681HA SJC13244682HA SJC13244929HA SJC13244930HA SJC13244931HA SJC13244932HA SJC13244938HA SJC13245120HA SJC13245121HA SJC13245122HA SJC13245123HA SJC13245124HA SJC13255139HA SJC13255140HA SJC13255141HA SJC13255142HA SJC13255143HA SJC13255144HA SJC13255145HA SJC13255146HA SJC13255147HA SJC13255148HA SJC13255320HA SJC13255322HA SJC13255324HA SJC13255328HA SJC13255329HA SJC13255350HA SJC13255351HA SJC13255353HA SJC13255355HA SJC13255358HA SJC13255359HA SJC13265506HA SJC13265508HA SJC13265509HA SJC13265555HA SJC13265556HA SJC13265557HA SJC13265558HA SJC13265562HA SJC13265570HA SJC13265571HA SJC13265572HA SJC13265573HA SJC13265574HA SJC13265575HA SJC13265576HA SJC13265577HA SJC13265578HA SJC13265580HA SJC13265582HA SJC13275790HA SJC13275791HA SJC13275792HA SJC13275793HA SJC13275794HA SJC13275795HA SJC13275796HA SJC13275797HA SJC13275804HA SJC13275805HA SJC13276097HA SJC13276098HA SJC13276099HA SJC13276101HA SJC13276102HA SJC13276104HA SJC13276106HA SJC13276109HA SJC13276110HA SJC13276111HA SJC13276112HA SJC13276113HA SJC13286329HA SJC13286333HA SJC13286334HA SJC13286335HA SJC13286340HA SJC13286341HA SJC13286401HA SJC13286407HA SJC13286410HA SJC13286609HA SJC13286615HA SJC13286616HA SJC13296973HA SJC13296974HA SJC13296978HA SJC13327704HA SJC13327705HA SJC13327706HA SJC13327710HA SJC13327711HA SJC13327712HA SJC13327941HA SJC13327942HA SJC13327943HA SJC13327944HA SJC13327945HA SJC13327946HA SJC13327947HA SJC13327948HA SJC13327949HA SJC13327950HA SJC13327951HA SJC13327952HA SJC13327953HA SJC13327954HA SJC13327955HA SJC13327956HA SJC13327957HA SJC13327958HA SJC13327959HA SJC13327960HA SJC13338115HA SJC13338116HA SJC13338117HA SJC13338118HA SJC13338119HA SJC13338120HA SJC13338121HA SJC13338122HA SJC13338123HA SJC13338124HA SJC13338125HA SJC13338126HA SJC13338127HA SJC13338128HA SJC13338129HA SJC13338130HA SJC13338131HA SJC13338132HA SJC13338133HA SJC13338137HA SJC13338138HA SJC13338139HA SJC13338140HA SJC13338141HA SJC13338142HA SJC13338143HA SJC13338144HA SJC13338145HA SJC13338146HA SJC13338148HA SJC13338149HA SJC13338150HA SJC13338151HA SJC13338152HA SJC13338154HA SJC13338155HA SJC13338156HA SJC13338158HA SJC13338159HA SJC13369172HA SJC13369180HA SJC13369288HA SJC13369289HA SJC13369302HA SJC13369304HA SJC13369305HA SJC13369308HA SJC13369309HA SJC13369311HA SJC13369312HA SJC13379525HA SJC13379526HA SJC13379527HA SJC13379528HA SJC13380324HA SJC13380335HA SJC13380336HA SJC13380340HA SJC13390823HA SJC13390824HA SJC13390825HA SJC13390826HA SJC13390827HA SJC13390828HA SJC13390829HA SJC13390830HA SJC13390831HA SJC13390832HA SJC13390833HA SJC13390834HA SJC13390835HA SJC13390836HA SJC13390837HA SJC13390838HA SJC13390839HA SJC13390840HA SJC13390841HA SJC13390842HA SJC13401196HA SJC13401197HA SJC13401198HA SJC13401199HA SJC13401200HA SJC13401202HA SJC13401330HA SJC13401332HA SJC13401335HA SJC13401356HA SJC13422120HA SJC13422125HA SJC13422127HA SJC13422130HA SJC13422131HA SJC13422132HA SJC13422133HA SJC13422135HA SJC13422136HA SJC13443320HA SJC13443330HA SJC13443581HA SJC13453996HA SJC13453997HA SJC13453998HA SJC13453999HA SJC13454000HA SJC13454001HA SJC13454002HA SJC13454003HA SJC13454004HA SJC13454005HA SJC13454006HA SJC13454008HA SJC13476331HA SJC13476390HA SJC13476391HA SJC13476392HA SJC13476393HA SJC13476396HA SJC13476397HA SJC13476398HA SJC13486747HA SJC13486748HA SJC13487117HA SJC13487118HA SJC13487119HA SJC13487120HA SJC13487121HA SJC13487122HA SJC13487125HA SJC13487126HA SJC13487127HA SJC13487128HA SJC13497710HA SJC13508540HA SJC13508543HA SJC13508545HA SJC13508546HA SJC13508547HA SJC13508548HA SJC13508549HA SJC13508550HA SJC13508551HA SJC13508552HA SJC13508553HA SJC13508554HA SJC13508555HA SJC13508556HA SJC13508557HA SJC13508686HA SJC13508687HA SJC13508689HA SJC13508690HA SJC13508691HA SJC13508692HA SJC13508693HA SJC13508694HA SJC13508695HA SJC13508696HA SJC13508697HA SJC13508698HA SJC13508699HA SJC13508700HA SJC13508701HA SJC13508702HA SJC13508703HA SJC13508704HA SJC13508821HA SJC13508822HA SJC13508823HA SJC13508824HA SJC13508825HA SJC13508826HA SJC13508827HA SJC13508828HA SJC13508829HA SJC13508838HA SJC13508855HA SJC13508860HA SJC13508871HA SJC13508884HA SJC13508890HA SJC13518937HA SJC13518938HA SJC13518942HA SJC13518949HA SJC13518950HA SJC13518952HA SJC13518953HA SJC13519179HA SJC13519180HA SJC13519181HA SJC13519182HA SJC13519183HA SJC13529499HA SJC14019738HA SJC14019739HA SJC14019740HA SJC14019741HA SJC14019742HA SJC14019743HA SJC14019744HA SJC14019745HA SJC14019747HA SJC14019748HA SJC14019749HA SJC14019750HA SJC14019751HA SJC14019752HA SJC14019753HA SJC14019754HA SJC14019755HA SJC14019756HA SJC14019757HA SJC14019758HA SJC14019759HA SJC14019760HA SJC14019761HA SJC14019762HA SJC14019763HA SJC14019764HA SJC14019765HA SJC14019766HA SJC14019767HA SJC14020207HA SJC14020208HA SJC14020209HA SJC14020210HA SJC14020211HA SJC14020212HA SJC14020213HA SJC14020214HA SJC14020215HA SJC14020219HA SJC14020220HA SJC14020222HA SJC14020227HA SJC14020232HA SJC14020234HA SJC14030671HA SJC14030672HA SJC14030673HA SJC14030677HA SJC14030678HA SJC14030679HA SJC14030680HA SJC14030682HA SJC14030683HA SJC14030684HA SJC14030685HA SJC14030763HA SJC14030766HA SJC14030767HA SJC14030768HA SJC14030769HA SJC14030771HA SJC14030772HA SJC14030773HA SJC14030774HA SJC14030780HA SJC14030781HA SJC14030782HA SJC14041563HA SJC14041565HA SJC14041789HA SJC14041804HA SJC14041806HA SJC14041809HA SJC14041810HA SJC14041811HA SJC14041812HA SJC14041813HA SJC13172552HA SJC13172567HA SJC13172601HA SJC13172765HA SJC13172768HA SJC13265751HA SJC13286400HA SJC13338369HA SJC13348830HA SJC13379521HA SJC13379523HA SJC13379524HA SJC13379530HA SJC13379531HA SJC13379532HA SJC13379535HA SJC13497673HA SJC13497679HA SJC13213581HA SJC13380222HA SJC13476343HA SJC13476347HA SJC13497695HA SJC13275803HA SJC13286409HA SJC13286411HA SJC13286612HA SJC13487260HA SJC13487262HA SJC13487263HA SJC13487264HA SJC13487266HA SJC13487267HA SJC13338463HA SJC13401370HA SJC13172607HA SJC13172611HA SJC13172617HA SJC13172627HA SJC13172759HA SJC13193222HA SJC13203253HA SJC13203264HA SJC13203275HA SJC13265708HA SJC13275799HA SJC13275800HA SJC13275806HA SJC13296971HA SJC13380001HA SJC13380002HA SJC13380010HA SJC13401201HA SJC13401203HA SJC13401204HA SJC13401205HA SJC13401206HA SJC13401208HA SJC13401209HA SJC13401210HA SJC13401211HA
SJC13401212HA SJC13401213HA SJC13401216HA SJC13401217HA SJC13401218HA SJC13401219HA SJC13401220HA SJC13401221HA SJC13401222HA SJC13401223HA SJC13401224HA SJC13401317HA SJC13401319HA SJC13401324HA SJC13464868HA SJC13464877HA SJC13464881HA SJC13487140HA SJC13487274HA SJC13487275HA SJC13487276HA SJC13487279HA SJC13497669HA SJC13497670HA SJC13497671HA SJC13497680HA SJC13497681HA SJC13497683HA SJC13497685HA SJC13172625HA SJC13172757HA SJC13172758HA SJC13172761HA SJC13213499HA SJC13213500HA SJC13244585HA SJC13296983HA SJC13380227HA SJC13380228HA SJC13380230HA SJC13380323HA SJC13380326HA SJC13380330HA SJC13380339HA SJC13380342HA SJC13464876HA SJC13476328HA SJC13476388HA SJC13476389HA SJC13508879HA SJC13508880HA SJC13519273HA SJC13519276HA SJC13519277HA SJC13519390HA SJC13519395HA SJC14020231HA SJC14030681HA SJC14030761HA SJC14030762HA SJC14030776HA SJC14030783HA SJC14030784HA SJC14030821HA SJC14030822HA 8225152 8249709 8294161 8311854 8311856 8311857 SJC13213517HA SJC13223969HA SJC13327707HA SJC13327708HA SJC13327713HA SJC13369310HA SJC13379976HA SJC13380011HA SJC13380023HA SJC13380220HA SJC13380223HA SJC13380224HA SJC13380225HA SJC13380226HA SJC13380229HA SJC13476373HA SJC13487143HA SJC13497677HA SJC13497678HA SJC13497701HA SJC13497732HA SJC13508852HA SJC13508854HA SJC13508857HA SJC13508859HA SJC13508872HA SJC13508873HA SJC13508875HA SJC13508877HA SJC13508883HA SJC13529498HA SJC14020229HA SJC14031146HA SJC13476368HA SJC13476369HA SJC13476370HA SJC14041790HA SJC14041791HA SJC14041792HA SJC14041793HA SJC14041794HA SJC14041795HA SJC14041796HA SJC14041797HA SJC14041798HA SJC14041799HA SJC14041800HA SJC14041801HA SJC14041802HA SJC14041803HA SJC14041817HA SJC14041822HA SJC14041824HA SJC14041825HA SJC14041826HA SJC14041827HA SJC14042243HA SJC14042244HA SJC14042245HA SJC13172557HA SJC13172569HA SJC13172570HA SJC13172604HA SJC13213571HA SJC13276095HA SJC13454019HA SJC13487138HA SJC13487144HA SJC14030819HA SJC14030830HA SJC14041818HA SJC14041819HA SJC14041820HA SJC13193231HA SJC13213506HA SJC13348832HA SJC13369139HA SJC13369140HA SJC13369157HA SJC13369161HA SJC13369169HA SJC13369173HA SJC13369295HA SJC13369297HA SJC13369299HA SJC13422139HA SJC13422207HA SJC13433030HA SJC13433031HA SJC13433037HA SJC13433038HA SJC13433039HA SJC13433040HA SJC13433041HA SJC13433042HA SJC13433043HA SJC13433044HA SJC13433045HA SJC13433047HA SJC13433051HA SJC13433052HA SJC13433053HA SJC13433054HA SJC13433055HA SJC13433056HA SJC13433057HA SJC13433058HA SJC13433059HA SJC13433071HA SJC13433075HA SJC13433076HA SJC13433078HA SJC13433079HA SJC13433092HA SJC13443325HA SJC13476344HA SJC13476375HA SJC13476399HA SJC13476400HA SJC13476401HA SJC13476402HA SJC13476403HA SJC13476404HA SJC13487154HA SJC13497687HA SJC13497704HA SJC13497708HA SJC13497709HA SJC13497711HA SJC13497728HA SJC13497729HA SJC13497739HA SJC13497740HA SJC13508466HA SJC13508467HA SJC13508472HA SJC13508473HA SJC13508474HA SJC13508475HA SJC13508477HA SJC13508480HA SJC13508481HA SJC13508538HA SJC13508542HA SJC13508688HA SJC13508881HA SJC13508882HA SJC13508885HA SJC13508888HA SJC13508889HA SJC13518934HA SJC13518935HA SJC13518936HA SJC13518947HA 8249700 8249739 8277062 8277149 8283742 8283775 8286326 SJC13213580HA SJC13203250HA SJC13203260HA SJC13203261HA SJC13203248HA SJC13203249HA SJC13379529HA SJC13379534HA SJC13379540HA SJC13379603HA SJC13379604HA SJC13380003HA SJC13380004HA SJC13380006HA SJC13380007HA SJC13380008HA SJC13380009HA SJC13380012HA SJC13380013HA SJC13380014HA SJC13380015HA SJC13380016HA SJC13380017HA SJC13380018HA SJC13380019HA SJC13380020HA SJC13380021HA SJC13380022HA SJC13380024HA SJC13380026HA SJC13380027HA SJC13487277HA SJC13487278HA SJC13497668HA SJC13497682HA SJC13497693HA SJC13497696HA SJC13497697HA SJC13497700HA SJC13497703HA SJC13497705HA SJC13497706HA SJC13497707HA SJC13497730HA SJC13497731HA SJC13508830HA SJC13508831HA SJC13508832HA SJC13508834HA SJC13508836HA SJC13508837HA SJC13508839HA SJC13508840HA SJC13508843HA SJC13508844HA SJC13497690HA SJC13497691HA SJC13497741HA SJC13497742HA SJC13497743HA SJC13508464HA SJC13508465HA SJC13508468HA SJC13508469HA SJC13487273HA SJC13172773HA SJC13422123HA SJC14030754HA SJC14030756HA SJC14030760HA SJC14030764HA SJC13058016HA SJC13172563HA SJC13172771HA SJC13213579HA SJC13213582HA SJC13223962HA SJC13234214HA SJC13234537HA SJC13244582HA SJC13244587HA SJC13244591HA SJC13244667HA SJC13244671HA SJC13244683HA SJC13255323HA SJC13255325HA SJC13255352HA SJC13255356HA SJC13255357HA SJC13265507HA SJC13265552HA SJC13265553HA SJC13265559HA SJC13265560HA SJC13265561HA SJC13265579HA SJC13265709HA SJC13265710HA SJC13265713HA SJC13265715HA SJC13265747HA SJC13275807HA SJC13286405HA SJC13286406HA SJC13296972HA 8277138 6638349 SJC13223960HA SJC13234229HA SJC13234230HA SJC13234236HA SJC13234238HA SJC13234240HA SJC13234536HA SJC13244933HA SJC13244935HA SJC13244936HA SJC13244937HA SJC13245119HA SJC13422208HA SJC13422210HA SJC13433033HA SJC13443113HA SJC13464866HA SJC13476333HA SJC13476334HA SJC13476335HA SJC13476339HA SJC13476340HA SJC13476341HA SJC13476376HA SJC13476378HA SJC13476379HA SJC13476380HA SJC13476381HA SJC13476382HA SJC13476383HA SJC13476387HA SJC13497674HA SJC13497689HA SJC14020216HA SJC14020217HA SJC13203246HA SJC13203257HA SJC13213498HA SJC13213501HA SJC13213502HA SJC13213504HA SJC13213505HA SJC13213507HA SJC13213515HA SJC13213516HA SJC13213568HA SJC13213569HA SJC13213792HA SJC13213793HA SJC13265703HA SJC13265707HA SJC13265714HA SJC13265749HA SJC13265750HA SJC13276096HA SJC13276103HA SJC13276107HA SJC13276108HA SJC13286403HA SJC13286404HA SJC13286618HA SJC13296975HA SJC13296979HA SJC13296980HA SJC13296984HA SJC13296985HA SJC13338147HA SJC13338153HA SJC13338160HA SJC13338161HA SJC13338162HA SJC13338163HA SJC13338164HA SJC13338355HA SJC13338356HA SJC13338357HA SJC13338358HA SJC13338359HA SJC13338361HA SJC13338362HA SJC13338363HA SJC13338364HA SJC13338365HA SJC13338366HA SJC13338367HA SJC13338368HA SJC13338370HA SJC13338371HA SJC13338372HA SJC13338373HA SJC13338374HA SJC13338437HA SJC13338438HA SJC13338439HA SJC13338440HA SJC13338441HA SJC13338442HA SJC13338443HA SJC13338444HA SJC13338445HA SJC13338446HA SJC13338447HA SJC13338448HA SJC13338465HA SJC13338467HA SJC13338469HA SJC13338474HA SJC13401354HA SJC13401357HA SJC13401361HA SJC13401362HA SJC13401364HA SJC13433048HA SJC13433073HA SJC13433074HA SJC13443321HA SJC13443326HA SJC13443474HA SJC13443476HA SJC13443477HA SJC13443479HA SJC13443480HA SJC13443481HA SJC13443482HA SJC13443483HA SJC13443484HA SJC13443485HA SJC13443576HA SJC13443578HA SJC13443579HA SJC13476326HA SJC13487155HA SJC13487156HA SJC13487158HA SJC13487159HA SJC13487261HA SJC13213513HA SJC13348503HA SJC13348514HA SJC13348828HA SJC13369136HA SJC13369141HA SJC13379536HA SJC13379537HA SJC13433064HA SJC13433066HA SJC13433072HA SJC13433090HA SJC13433091HA SJC13443112HA SJC13443322HA SJC13443327HA SJC13443328HA SJC13443329HA SJC13476315HA SJC13476316HA SJC13476320HA SJC13476324HA SJC13476329HA SJC13476330HA SJC13476337HA SJC13476346HA SJC13476374HA SJC13380328HA SJC13380331HA SJC13380332HA SJC13380333HA SJC13380341HA SJC13172762HA SJC13213584HA SJC13508845HA SJC13508846HA SJC13508847HA SJC13508858HA SJC13275789HA SJC13275801HA SJC13275802HA SJC13286330HA SJC13286331HA SJC13286332HA SJC13296988HA SJC13296989HA SJC13348509HA SJC13172560HA SJC13172577HA SJC13172634HA SJC13172760HA SJC13193217HA SJC13193219HA SJC13203258HA SJC13203259HA SJC13203271HA SJC13203272HA SJC13203273HA SJC13213488HA SJC13213489HA SJC13213493HA SJC13213496HA SJC13213510HA SJC13213511HA SJC13213512HA SJC13213787HA SJC13234237HA SJC13265711HA SJC13296970HA SJC13348497HA SJC13348498HA SJC13348499HA SJC13348504HA SJC13348505HA SJC13380327HA SJC13401326HA SJC13401327HA SJC13401329HA SJC13401337HA SJC13401355HA SJC13401359HA SJC13401363HA SJC13433065HA SJC13433077HA SJC13443473HA SJC13454016HA SJC13476377HA SJC13487124HA SJC13487150HA SJC13487151HA SJC13487271HA SJC13519274HA SJC13519275HA SJC13519278HA SJC13519388HA SJC13519389HA SJC13519391HA SJC13519392HA SJC13519393HA SJC13529492HA SJC13529494HA SJC14030812HA SJC14030823HA SJC14031147HA SJC14041562HA SJC14041808HA SJC14052422HA SJC14052423HA 8220841 SJC13172576HA SJC13172579HA SJC13172621HA SJC13172632HA SJC13172633HA SJC13172763HA SJC13193220HA SJC13203266HA SJC13213508HA SJC13213790HA SJC13213791HA SJC13213795HA SJC13213798HA SJC13213799HA SJC13244934HA SJC13255321HA SJC13255326HA SJC13255327HA SJC13265554HA SJC13275808HA SJC13286605HA SJC13286608HA SJC13286610HA SJC13348826HA SJC13379696HA SJC13380025HA SJC13380029HA SJC13401320HA SJC13422121HA SJC13422138HA SJC13476348HA SJC13476349HA SJC13476384HA SJC13476385HA SJC13476386HA SJC13508685HA SJC14020218HA SJC13203251HA SJC13203274HA SJC13338475HA SJC13379539HA SJC13379601HA SJC13379602HA SJC13379687HA SJC13379688HA SJC13379689HA SJC13379691HA SJC13379977HA SJC13379978HA SJC13401367HA SJC13476317HA SJC13476318HA SJC13508887HA SJC13518948HA SJC14030747HA SJC14030752HA SJC14030753HA SJC14030755HA SJC13348824HA SJC13348825HA SJC13519272HA SJC13296976HA SJC13296982HA SJC13296986HA SJC13296987HA SJC13172764HA SJC13203270HA SJC13213490HA SJC13213491HA SJC13213583HA SJC13487123HA SJC13487129HA SJC13487130HA SJC13487131HA SJC13487133HA SJC13487134HA SJC13487135HA SJC13487136HA SJC13487137HA SJC13487149HA SJC13433069HA SJC13476367HA SJC14030778HA SJC14030779HA SJC13172622HA SJC13203255HA SJC13203256HA SJC13422203HA SJC13465031HA SJC13476345HA SJC13203252HA SJC13443114HA SJC13487152HA SJC13487153HA SJC13497675HA SJC13497694HA SJC13519185HA SJC13519187HA SJC13519188HA SJC13519269HA SJC13519271HA SJC13519397HA 8225151 8249678 8249680 8249778 8249780 8277008 8277103 8277104 8286324 8286325 8294158 8294162 8294163 SJC13172578HA SJC13172605HA SJC13172612HA SJC13172613HA SJC13172614HA SJC13172615HA SJC13172618HA SJC13172624HA SJC13172626HA SJC13172628HA SJC13172629HA SJC13172631HA SJC13193215HA SJC13193218HA SJC13193221HA SJC13193223HA SJC13193226HA SJC13193228HA SJC13203247HA SJC13203254HA SJC13203263HA SJC13203265HA SJC13213572HA SJC13213575HA SJC13213576HA SJC13213585HA SJC13213587HA SJC13213780HA SJC13213781HA SJC13213796HA SJC13223955HA SJC13223963HA SJC13223965HA SJC13223966HA SJC13223970HA SJC13234233HA SJC13234234HA SJC13234235HA SJC13234520HA SJC13234544HA SJC13244684HA SJC13245125HA SJC13255354HA SJC13265716HA SJC13276094HA SJC13276105HA SJC13286336HA SJC13286337HA SJC13286338HA SJC13286339HA SJC13286408HA SJC13286606HA SJC13286607HA SJC13286611HA SJC13286613HA SJC13286617HA SJC13296977HA SJC13296981HA SJC13327709HA SJC13338134HA SJC13338135HA SJC13338136HA SJC13338360HA SJC13338462HA SJC13338464HA SJC13338466HA SJC13338468HA SJC13338470HA SJC13348831HA SJC13348833HA SJC13369171HA SJC13369174HA SJC13369292HA SJC13369306HA SJC13380028HA SJC13422134HA SJC13422137HA SJC13422202HA SJC13422204HA SJC13422205HA SJC13422209HA SJC13422211HA SJC13433032HA SJC13433034HA SJC13433035HA SJC13433046HA SJC13433049HA SJC13433050HA SJC13433068HA SJC13433080HA SJC13433081HA SJC13433082HA SJC13433083HA SJC13433084HA SJC13433085HA SJC13433086HA SJC13433087HA SJC13433088HA SJC13433089HA SJC13443323HA SJC13443577HA SJC13443580HA SJC13443582HA SJC13487270HA SJC13487272HA SJC13497676HA SJC13508835HA SJC13508851HA SJC13508853HA SJC13508856HA SJC13508874HA SJC13508876HA SJC13508878HA SJC13518939HA SJC14030757HA SJC14030816HA SJC14030820HA SJC14030872HA SJC14041555HA SJC14041556HA SJC14041557HA SJC14041558HA SJC14041560HA SJC14041805HA SJC14041807HA SJC14041814HA SJC14041815HA SJC14041816HA SJC14041821HA SJC14041823HA 8184638 8225147 8225149 8225150 8249781 8277009 8277102 8277140 8277145 8277146 8286323 8294164 SJC13172619HA SJC13172623HA SJC13193224HA SJC13193225HA SJC13193227HA SJC13193230HA SJC13193232HA SJC13203262HA SJC13213492HA SJC13213494HA SJC13213586HA SJC13213778HA SJC13213779HA SJC13223954HA SJC13244586HA SJC13244673HA SJC13244676HA SJC13245127HA SJC13245128HA SJC13265583HA SJC13265584HA SJC13265705HA SJC13265753HA SJC13276100HA SJC13369156HA SJC13369296HA SJC13369303HA SJC13401369HA SJC13422122HA SJC13422124HA SJC13422206HA SJC13433036HA SJC13443319HA SJC13443324HA SJC13454013HA SJC13464864HA SJC13497686HA SJC13497688HA SJC13497692HA SJC13497698HA SJC13497699HA SJC13497702HA SJC13497744HA SJC13497745HA SJC13508539HA SJC13508541HA SJC13518941HA SJC13518943HA SJC13518944HA SJC13518945HA SJC13518946HA SJC13519184HA SJC13519396HA SJC14020230HA SJC14020233HA SJC14030758HA SJC14030759HA SJC14030814HA 8220843 6147434 SJC13213570HA SJC13058019HA SJC13213497HA SJC13213776HA SJC13234213HA SJC13369175HA SJC13369293HA SJC13369298HA SJC13369300HA SJC13454011HA SJC13454018HA SJC13454021HA SJC13454024HA SJC13454025HA SJC13286402HA SJC13401333HA SJC13508886HA SJC13213509HA SJC13401358HA SJC13401365HA SJC13422128HA SJC13265746HA SJC13265748HA SJC13265752HA SJC13265754HA SJC13369301HA SJC13348500HA SJC13348501HA SJC13348513HA SJC13476362HA SJC13476363HA SJC13487268HA SJC13487269HA SJC14041561HA SJC14052424HA SJC14052425HA SJC13348827HA SJC13379608HA SJC13380221HA SJC13380234HA SJC13401195HA SJC13454015HA SJC13454022HA SJC13454023HA SJC13476332HA SJC13476342HA SJC13476352HA SJC13476394HA SJC13476395HA SJC13486749HA SJC13486750HA SJC13213495HA SJC13213574HA SJC13213577HA SJC13213775HA SJC13213777HA SJC13223953HA SJC13223956HA SJC13223957HA SJC13223959HA SJC13234215HA SJC13234216HA SJC13234217HA SJC13234218HA SJC13234221HA SJC13234222HA SJC13234223HA SJC13234224HA SJC13234225HA SJC13234226HA SJC13234227HA SJC13234228HA SJC13234231HA SJC13234232HA SJC13234239HA SJC13234521HA SJC13234526HA SJC13234539HA SJC13234545HA SJC13234546HA SJC13234547HA SJC13234548HA SJC13244576HA SJC13244577HA SJC13244578HA SJC13244579HA SJC13244580HA SJC13338157HA SJC13379522HA SJC13379533HA SJC13379538HA SJC13379605HA SJC13379606HA SJC13380231HA SJC13380232HA SJC13380233HA SJC13380325HA SJC13380329HA SJC13380334HA SJC13380337HA SJC13380338HA SJC13401316HA SJC13401336HA SJC13422126HA SJC13443332HA SJC13476325HA SJC13476336HA SJC13476338HA SJC13497735HA SJC13508470HA SJC13508471HA SJC13508476HA SJC13508478HA SJC13508479HA SJC13508482HA SJC13508483HA SJC13518951HA SJC14019746HA SJC14030874HA SJC14030875HA SJC14030885HA SJC14030886HA SJC14030887HA SJC14041550HA 8220847 SJC13213783HA SJC13213785HA SJC13213786HA SJC13234549HA 8220807 SJC13213784HA SJC13223958HA SJC13234527HA SJC13172574HA SJC13172575HA SJC13213503HA SJC13213514HA SJC13213788HA SJC13213789HA SJC13213794HA SJC13223951HA SJC13223952HA SJC13286614HA SJC13519270HA SJC13193216HA SJC13265704HA SJC13265712HA SJC13265755HA SJC13348508HA SJC13348510HA SJC13348515HA SJC13369170HA SJC13369179HA SJC13369183HA SJC13369290HA SJC13369291HA SJC13369294HA SJC13369307HA SJC13379607HA SJC13379609HA SJC13379610HA SJC13379686HA SJC13379690HA SJC13379692HA SJC13379693HA SJC13379694HA SJC13379695HA SJC13379697HA SJC13379698HA SJC13379969HA SJC13379970HA SJC13379971HA SJC13379972HA SJC13379973HA SJC13379974HA SJC13379975HA SJC13380005HA SJC13401214HA SJC13401315HA SJC13401318HA SJC13401321HA SJC13401322HA SJC13401323HA SJC13401331HA SJC13401334HA SJC13401338HA SJC13433067HA SJC13433093HA SJC13443107HA SJC13443116HA SJC13443478HA SJC13464860HA SJC13464861HA SJC13464862HA SJC13464865HA SJC13464869HA SJC13464872HA SJC13464873HA SJC13464875HA SJC13487157HA SJC13487265HA SJC13497672HA SJC13497684HA SJC13497733HA SJC13497734HA SJC13497736HA SJC13497737HA SJC13497738HA SJC14020221HA SJC14020223HA SJC14020225HA SJC14020226HA SJC14020228HA SJC14020235HA 8184633 8220842 8220844 8220846 8220851 8225148 8277100 8277101 8283770 8283771 8283772 6343819 6665782 6843533 6820071 6820072 6843531 6843534 6572663 6892566 SJC13213573HA SJC13213578HA SJC13213782HA SJC13213797HA SJC13223961HA SJC13223964HA SJC13223967HA SJC13223968HA SJC13234211HA SJC13234212HA SJC13234219HA SJC13234220HA SJC13265581HA SJC13265706HA SJC13275798HA SJC13286412HA SJC13338471HA SJC13338472HA SJC13338473HA SJC13348496HA SJC13348502HA SJC13348506HA SJC13348507HA SJC13348511HA SJC13348512HA SJC13348829HA SJC13369137HA SJC13369138HA SJC13369142HA SJC13369143HA SJC13369145HA SJC13369146HA SJC13369147HA SJC13369148HA SJC13369149HA SJC13369150HA SJC13369151HA SJC13369152HA SJC13369153HA SJC13369154HA SJC13369155HA SJC13369158HA SJC13369159HA SJC13369160HA SJC13369162HA SJC13369163HA SJC13369164HA SJC13369165HA SJC13369166HA SJC13369167HA SJC13369168HA SJC13369176HA SJC13369177HA SJC13369178HA SJC13369181HA SJC13369182HA SJC13369184HA SJC13401366HA SJC13401368HA SJC13422129HA SJC13454007HA SJC13454009HA SJC13454010HA SJC13464878HA SJC13465034HA SJC13465035HA SJC13465036HA SJC13465141HA SJC13465142HA SJC13465144HA SJC13465145HA SJC13476319HA SJC13476321HA SJC13476322HA SJC13476323HA SJC13476327HA SJC13476350HA SJC13476351HA SJC13476353HA SJC13476354HA SJC13476355HA SJC13476356HA SJC13476357HA SJC13476358HA SJC13476359HA SJC13476360HA SJC13476361HA SJC13476364HA SJC13476365HA SJC13476366HA SJC13476371HA SJC13476372HA SJC13487132HA SJC13487139HA SJC13487141HA SJC13487142HA SJC13487145HA SJC13487146HA SJC13487147HA SJC13487148HA SJC13529493HA SJC13529495HA SJC13529496HA SJC13529497HA SJC13529500HA SJC13529501HA SJC14030674HA SJC14030675HA SJC14030676HA SJC14030748HA SJC14030749HA SJC14030750HA SJC14030751HA SJC14030777HA SJC14030813HA SJC14030815HA SJC14030818HA SJC14030824HA 8286322 6395876 SJC13245126HA SJC13519186HA SJC13519394HA SJC14041559HA SJC13234525HA SJC13369144HA |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
TECHNICAL SUPPORT 262-513-4122
|
Manufacturer Reason for Recall |
When the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation visual movements of the hand are seen after TOF (Train of Four) stimulation, but the patient monitor shows no counts, or counts are not corresponding to the actual a
|
FDA Determined Cause 2 |
Software design |
Action |
GE Healthcare sent an "Urgent Medical Device Correction" letter (Ref 36099) dated April 9, 2014 to affected customers. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information (contact Technical Support at 1-800-558-7044 or your local Service Representative).
UPDATE:
On May 11, 2015 a second GE Healthcare letter FMI 36099-2 was sent to OUS consignees. The updated report was for additional affected modules due to FRU installation. |
Quantity in Commerce |
1,794 (471 units US, 1,323 units OUS) |
Distribution |
US Nationwide Distribution in the states of: AR, CA, CO,DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MI, MN, MO, NC, NV, NY, OH, OK, TN, TX, UT, WA, WI, WY.
OUS: ARGENTINA. AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, CANADA, CHILE, COLOMBIA, COSTA RICA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGRY, INDIA, IRA, ISRAEL, ITALY, JAPAN, KAZAKTSTAN, KUWAIT, LATVIA, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PANAMA, POLAND, QATAR, ROMANIA, RUSSIAN, F¿D¿RATION, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, TURKEY, TAIWAN, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KOI and Original Applicant = DATEX MEDICAL INSTRUMENTATION, INC. 510(K)s with Product Code = KOI and Original Applicant = GE HEALTHCARE
|
|
|
|