• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 046 Date of Entry 12/23/2016 
FR Recognition Number 11-314
Standard
ISO 14242-2 Second edition 2016-09-15
Implants for surgery - Wear of total hip-joint prostheses - Part 2: Methods of measurement
Scope/Abstract
This part of ISO 14242 specifies methods of assessment of wear of the acetabular component of total hip-joint prostheses using gravimetric techniques and changes in dimensional form of components tested in accordance with ISO 14242-1 or ISO 14242-3 as appropriate.

NOTE Some investigators have experienced problems with organic deposits affecting the results of measurements, especially with hard/hard combinations. No specific precautions are included in this part of ISO 14242, but cleaning techniques adopted is intended to be suitable for the soils produced.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§888.3300 Prosthesis, Hip, Constrained, Metal Class 3 KXD
§888.3310 Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer Class 2 KWZ
§888.3310 Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer, + Additive Class 2 PBI
§888.3320 Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) Class 3 JDL
§888.3330 Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component) Class 3 KWA
§888.3340 Prosthesis, Hip, Semi-Constrained, Composite/Metal Class 2 KMC
§888.3350 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented Class 2 JDI
§888.3350 Hip, Semi-Constrained, Cemented, Metal/Polymer + Additive, Cemented Class 2 OQH
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented Class 2 LZO
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish Class 2 MAY
§888.3353 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate Class 2 MEH
§888.3353 Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented Class 2 OQI
§888.3358 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented Class 2 LPH
§888.3358 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous Class 2 MBL
§888.3358 Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented Class 2 OQG
§888.3360 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented Class 2 LWJ
§888.3390 Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented Class 2 KWY
§888.3410 Prosthesis, Hip, Pelvifemoral Resurfacing, Metal/Polymer Class 3 KXB
§888.3410 Prosthesis, Hip, Pelvifemoral Resurfacing, Metal/Polymer, Uncemented Class 3 OCG
N/A Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic, Cemented Class 3 LPF
N/A Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented Class 3 MRA
N/A Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing Class 3 NXT
N/A Prosthesis, Hip, Semi-Constrained, Ceramic-On-Metal Articulation Class 3 OVO
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Lindsay Jeffries
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIA
  301-796-3740
  Lindsay.Jeffries@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Orthopedic
*These are provided as examples and others may be applicable.
-
-