Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 056 Date of Entry 06/07/2021 
FR Recognition Number 11-382
Standard
ASTM F3090-20
Standard Test Method for Fatigue Testing of Acetabular Devices for Total Hip Replacement
Scope/Abstract
1.1 This test method is intended to evaluate the fatigue strength of metallic acetabular shells with hemispheric outer surfaces.

1.2 This test method, as described, is not intended to evaluate the following: the strength of components that may be mated with the acetabular shells (for example, augments, acetabular liners); attributes of the shells not related to strength (for example, fixation, coating adhesion); the strength of acetabular shell features away from, or loaded differently than, the primary load bearing region of the shell (for example, screws, spikes, flanges); non-hemispherical shells (for example, patient-matched geometries); or corrosion between modular components.

1.3 Modifications to this test method (for example, different support medium, different /position of unsupported region, different testing environment) may result in a method appropriate to evaluate the characteristics listed in 1.2. Such modification must have adequate justification.

1.4 Although the methodology described does not replicate all physiological force conditions, it is a means of in vitro comparison of acetabular device designs under repeated forces.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§888.3353 Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented Class 2 OQI
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish Class 2 MAY
§888.3353 Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented Class 2 LZO
§888.3353 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate Class 2 MEH
Unclassified Prosthesis, Hip, Semi-Constrained, Ceramic-On-Metal Articulation Class 3 OVO
Unclassified Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic, Cemented Class 3 LPF
Unclassified Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented Class 3 MRA
Unclassified Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing Class 3 NXT
§888.3300 Prosthesis, Hip, Constrained, Metal Class 3 KXD
§888.3330 Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component) Class 3 KWA
§888.3350 Hip, Semi-Constrained, Cemented, Metal/Polymer + Additive, Cemented Class 2 OQH
§888.3350 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented Class 2 JDI
§888.3358 Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented Class 2 OQG
§888.3358 Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented Class 2 LPH
§888.3358 Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous Class 2 MBL
§888.3310 Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer Class 2 KWZ
§888.3310 Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer, + Additive Class 2 PBI
§888.3320 Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) Class 3 JDL
§888.3340 Prosthesis, Hip, Semi-Constrained, Composite/Metal Class 2 KMC
§888.3410 Prosthesis, Hip, Pelvifemoral Resurfacing, Metal/Polymer Class 3 KXB
§888.3410 Prosthesis, Hip, Pelvifemoral Resurfacing, Metal/Polymer, Uncemented Class 3 OCG
FDA Technical Contact
 John Goode
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIA/
  301-796-6407
  john.goode@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Orthopedic
*These are provided as examples and others may be applicable.
-
-